Comparison of the Baska® mask with the single-use laryngeal mask airway in low-risk female patients undergoing ambulatory surgery

被引:24
|
作者
Alexiev, V. [1 ,2 ]
Ochana, A. [1 ,2 ]
Abdelrahman, D. [1 ,2 ]
Coyne, J. [1 ,2 ]
McDonnell, J. G. [1 ,2 ]
O'Toole, D. P. [1 ,2 ]
Neligan, P. [1 ,2 ]
Laffey, J. G. [3 ,4 ]
机构
[1] Galway Univ Hosp, Dept Anaesthesia, Galway, Ireland
[2] Natl Univ Ireland, Galway, Ireland
[3] St Michaels Hosp, Dept Anesthesia, Toronto, ON M5B 1W8, Canada
[4] Univ Toronto, Toronto, ON, Canada
关键词
CERVICAL-SPINE IMMOBILIZATION; ANESTHETIZED ADULT PATIENTS; SUPRAGLOTTIC AIRWAY; TRACHEAL INTUBATION; I-GEL; MACINTOSH LARYNGOSCOPES; TRIAL; AIRWAYSCOPE(R); AIRTRAQ(R); MANAGEMENT;
D O I
10.1111/anae.12356
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We compared the Baska((R)) mask with the single-use classic laryngeal mask airway (cLMA) in 150 females at low risk for difficult tracheal intubation in a randomised, controlled clinical trial. We found that median (IQR [range]) seal pressure was significantly higher with the Baska mask compared with the cLMA (40 (34-40 [16-40]) vs 22 (18-25 [14-40])cmH(2)O, respectively, p<0.001), indicating a better seal. In contrast, the first time success rate for insertion of the Baska mask was lower than that seen with the cLMA (52/71 (73%) vs 77/99 (98%), respectively, p<0.001). There were no differences in overall device insertion success rates (78/79 (99%) vs 68/71 (96%), respectively, p=0.54). The Baska mask proved more difficult to insert, requiring more insertion attempts, taking longer to insert and had higher median (IQR [range]) insertion difficulty scores (1.6 (0.8-2.2 [0.1-5.6]) vs 0.5 (0.3-1.4 [0.1-4.0]), respectively, p<0.001). There was also an increased rate of minor blood staining of the Baska mask after removal, but there were no differences in other complication rates, such as laryngospasm, or in the severity of throat discomfort. In conclusion, in clinical situations where the seal with the glottic aperture takes priority over ease of insertion, the Baska mask may provide a useful alternative to the cLMA.
引用
收藏
页码:1026 / 1032
页数:7
相关论文
共 50 条
  • [1] Clinical evaluation of the Baska Mask (R) laryngeal mask in adult patients in ambulatory surgery
    Lopez, A. M.
    Munoz-Rojas, G.
    Fontanals, M.
    de San Jose, I.
    Hermoso, A.
    Valero, R.
    [J]. REVISTA ESPANOLA DE ANESTESIOLOGIA Y REANIMACION, 2015, 62 (10): : 551 - 556
  • [2] A randomised crossover trial comparing a single-use polyvinyl chloride laryngeal mask airway with a single-use silicone laryngeal mask airway
    Bell, S. F.
    Morris, N. G.
    Rao, A.
    Wilkes, A. R.
    Goodwin, N.
    [J]. ANAESTHESIA, 2012, 67 (12) : 1337 - 1342
  • [3] A comparison of endotracheal intubation and use of the laryngeal mask airway for ambulatory oral surgery patients
    Todd, DW
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2002, 60 (01) : 2 - 4
  • [4] A comparison of the oropharyngeal leak pressure between the reusable Classic™ Laryngeal Mask Airway and the single-use Soft Seal™ Laryngeal Mask Airway
    Hanning, SJ
    McCulloch, TJ
    Orr, B
    Anderson, SP
    [J]. ANAESTHESIA AND INTENSIVE CARE, 2006, 34 (02) : 237 - 239
  • [5] A comparison of endotracheal intubation and use of the laryngeal mask airway for ambulatory oral surgery patients - Discussion
    Bennett, J
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2002, 60 (01) : 4 - 5
  • [6] COMPARISON OF THREE METHODS FOR THE CONFIRMATION OF LARYNGEAL MASK AIRWAY PLACEMENT IN FEMALE PATIENTS UNDERGOING GYNECOLOGIC SURGERY
    Zhou, Zhen-feng
    Xia, Chen-zhong
    Wu, Meng
    Yu, Li-na
    Yan, Guo-zhang
    Ren, Qiu-sheng
    Hu, Ci-xian
    Yan, Min
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (05): : 1212 - 1220
  • [7] Use of the laryngeal mask airway in the ambulatory setting
    Wat, LI
    Brimacombe, JR
    White, PF
    Templin, PA
    Lynch, ME
    Hammamura, RK
    [J]. JOURNAL OF CLINICAL ANESTHESIA, 1998, 10 (05) : 386 - 388
  • [8] Comparison of the Baska mask and the i-gel supraglottic airway devices in patients undergoing elective surgery
    Sinasamy, Thanesh Kumar
    Hassan, Wan Mohd Nazaruddin Wan
    Zaini, Rhendra Hardy Mohamad
    Seevaunnamtum, Praveena
    Ab Mukmin, Laila
    [J]. ANAESTHESIOLOGY INTENSIVE THERAPY, 2020, 52 (05) : 383 - 388
  • [9] Comparison of the standard laryngeal mask airway and the ProSeal laryngeal mask airway in obese patients
    Natalini, G
    Franceschetti, ME
    Pantelidi, MT
    Rosano, A
    Lanza, G
    Bernardini, A
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2003, 90 (03) : 323 - 326
  • [10] Comparison of the single use and reusable intubating laryngeal mask airway
    Teoh, W. H. L.
    Lim, Y.
    [J]. ANAESTHESIA, 2007, 62 (04) : 381 - 384