Partial least squares regression calibration for determining wax content in processed flax fiber by near-infrared spectroscopy

被引:5
|
作者
Sohn, M [1 ]
Himmelsbach, DS [1 ]
Morrison, WH [1 ]
Akin, DE [1 ]
Barton, FE [1 ]
机构
[1] USDA ARS, Richard B Russell Agr Res Ctr, Athens, GA 30605 USA
关键词
near-infrared spectroscopy; NIRS; partial least squares; PLS; principal component analysis; PCA; flax fiber; wax;
D O I
10.1366/000370206776593663
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The quality of flax fiber in the textile industry is closely related to the wax content remaining on the fiber after the cleaning process. Extraction by organic solvents, which is currently used for determining wax content, is very time consuming and produces chemical waste. In this study, near-infrared (NIR) spectroscopy was used as a rapid analytical technique to develop models for wax content associated with flax fiber. Calibration samples (n = 11) were prepared by manually mixing dewaxed fiber and isolated wax to provide a range of wax content front 0 to 5%. A total of fourteen flax fiber samples obtained after a cleaning process were used for prediction. Principal component analysis demonstrated that one principal component is enough to separate the flax fibers by their wax content. The most highly correlated wavelengths were 2312, 2352, 1732, and 1766 nm, in order of significance. Partial least squares models were developed with various chemometric preprocessing approaches to obtain the best model performance. Two models, one using the entire region (1100-2498 nm) and the other using the selected wavelengths, were developed and the accuracies compared. For the model using the entire region, the correlation coefficient (RI) between actual and predicted values was 0.996 and the standard error of prediction (RMSEP) was 0.289%. For the selected-wavelengths model, the RI was 0.997 and RMSEP was 0.272%. The results suggested that NIR spectroscopy can be used to determine wax content in very clean flax fiber and that development of a low-cost device, using few wavelengths, should be possible.
引用
收藏
页码:437 / 440
页数:4
相关论文
共 50 条
  • [1] Application of partial least-squares regression to near-infrared reflectance spectroscopic determination of shive content in flax
    Sohn, M
    Barton, FE
    Morrison, WH
    Archibald, DD
    [J]. APPLIED SPECTROSCOPY, 2003, 57 (05) : 551 - 556
  • [2] Optimisation of partial least squares regression calibration models in near-infrared spectroscopy: a novel algorithm for wavelength selection
    Smith, MR
    Jee, RD
    Moffat, AC
    Rees, DR
    Broad, NW
    [J]. ANALYST, 2003, 128 (11) : 1312 - 1319
  • [3] Application of latent root regression for calibration in near-infrared spectroscopy. Comparison with principal component regression and partial least squares
    Vigneau, E
    Bertrand, D
    Qannari, EM
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1996, 35 (02) : 231 - 238
  • [4] Variable space boosting partial least squares for multivariate calibration of near-infrared spectroscopy
    Bian, Xihui
    Li, Shujuan
    Shao, Xueguang
    Liu, Peng
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2016, 158 : 174 - 179
  • [5] Near-Infrared Spectroscopy Analytical Model Using Ensemble Partial Least Squares Regression
    Luo, Na
    Han, Ping
    Wang, Shifang
    Wang, Dong
    Zhao, Chunjiang
    [J]. ANALYTICAL LETTERS, 2019, 52 (11) : 1732 - 1756
  • [6] Variable selection in random calibration of near-infrared instruments: ridge regression and partial least squares regression settings
    Gusnanto, A
    Pawitan, Y
    Huang, J
    Lane, B
    [J]. JOURNAL OF CHEMOMETRICS, 2003, 17 (03) : 174 - 185
  • [7] Analysis of fiber content in flax stems by near-infrared spectroscopy
    Barton, FE
    Akin, DE
    Morrison, WH
    Ulrich, A
    Archibald, DD
    [J]. JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2002, 50 (26) : 7576 - 7580
  • [8] Hemodialysis monitoring using mid- and near-infrared spectroscopy with partial least squares regression
    Henn, Raphael
    Kirchler, Christian G.
    Schirmeister, Zora L.
    Roth, Andreas
    Maentele, Werner
    Huck, Christian W.
    [J]. JOURNAL OF BIOPHOTONICS, 2018, 11 (07)
  • [9] Determination of organic matter in soil using near-infrared spectroscopy and partial least squares regression
    Fidêncio, PH
    Poppi, RJ
    de Andrade, JC
    Cantarella, H
    [J]. COMMUNICATIONS IN SOIL SCIENCE AND PLANT ANALYSIS, 2002, 33 (9-10) : 1607 - 1615
  • [10] Calibration standardization algorithm for partial least-squares regression: Application to the determination of physiological levels of glucose by near-infrared spectroscopy
    Zhang, L
    Small, GW
    Arnold, MA
    [J]. ANALYTICAL CHEMISTRY, 2002, 74 (16) : 4097 - 4108