Validation of the continuous glucose monitoring sensor in preterm infants

被引:52
|
作者
Beardsall, K. [1 ,2 ]
Vanhaesebrouck, S. [3 ]
Ogilvy-Stuart, A. L. [2 ]
Vanhole, C. [3 ]
vanWeissenbruch, M. [4 ]
Midgley, P. [5 ]
Thio, M. [6 ]
Cornette, L. [7 ]
Ossuetta, I. [8 ]
Palmer, C. R. [9 ]
Iglesias, I. [6 ]
de Jong, M. [4 ]
Gill, B. [7 ]
de Zegher, F. [3 ]
Dunger, D. B. [1 ]
机构
[1] Univ Cambridge, Cambridge Univ Hosp NHS Fdn Trust, Dept Paediat, Cambridge CB2 1TN, England
[2] Cambridge Univ Hosp NHS Fdn Trust, Neonatal Unit, Cambridge, England
[3] Univ Louvain, Dept Woman & Child, Louvain, Belgium
[4] Vrije Univ Amsterdam Med Ctr, Amsterdam, Netherlands
[5] Royal Infirm Edinburgh NHS Trust, Simpson Ctr Reprod Hlth, Edinburgh, Midlothian, Scotland
[6] Hosp Univ St Joan de Deu, Barcelona, Spain
[7] Leeds Gen Infirm, Leeds, W Yorkshire, England
[8] Luton & Dunstable Hosp, Luton, Beds, England
[9] Univ Cambridge, Inst Publ Hlth, Ctr Appl Med Stat, Dept Publ Hlth & Primary Care, Cambridge CB2 1TN, England
关键词
INSULIN THERAPY; RISK-FACTOR; SYSTEM; HYPERGLYCEMIA; PERFORMANCE; ACCURACY; DEATH; CGMS;
D O I
10.1136/archdischild-2012-301661
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective Recent studies have highlighted the need for improved methods of monitoring glucose control in intensive care to reduce hyperglycaemia, without increasing the risk of hypoglycaemia. Continuous glucose monitoring is increasingly used in children with diabetes, but there are little data regarding its use in the preterm infant, particularly at extremes of glucose levels and over prolonged periods. This study aimed to assess the accuracy of the continuous glucose monitoring sensor (CGMS) across the glucose profile, and to determine whether there was any deterioration over a 7 day period. Design Prospectively collected CGMS data from the NIRTURE Trial was compared with the data obtained simultaneously using point of care glucose monitors. Setting An international multicentre randomised controlled trial. Patients One hundred and eighty-eight very low birth weight control infants. Outcome measures Optimal accuracy, performance goals (American Diabetes Association consensus), Bland Altman, Error Grid analyses and accuracy. Results The mean (SD) duration of CGMS recordings was 156.18 (29) h (6.5 days), with a total of 5207 paired glucose levels. CGMS data correlated well with point of care devices (r=0.94), with minimal bias. It met the Clarke Error Grid and Consensus Grid criteria for clinical significance. Accuracy of single readings to detect set thresholds of hypoglycaemia, or hyperglycaemia was poor. There was no deterioration over time from insertion. Conclusions CGMS can provide information on trends in glucose control, and guidance on the need for blood glucose assessment. This highlights the potential use of CGMS in optimising glucose control in preterm infants.
引用
收藏
页码:F136 / F140
页数:5
相关论文
共 50 条
  • [1] Real-Time Continuous Glucose Monitoring in Preterm Infants
    Joseph, Rachel
    Rosa, Jenna Hope-Altomari
    [J]. NEONATAL NETWORK, 2022, 41 (06): : 376 - 376
  • [2] Continuous glucose monitoring for the prevention of morbidity and mortality in preterm infants
    Galderisi, Alfonso
    Bruschettini, Matteo
    Russo, Chiara
    Hall, Rebecka
    Trevisanuto, Daniele
    [J]. COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2020, (12):
  • [3] Continuous glucose monitoring for the prevention of morbidity and mortality in preterm infants
    Galderisi, Alfonso
    Trevisanuto, Daniele
    Russo, Chiara
    Hall, Rebecka
    Bruschettini, Matteo
    [J]. COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2021, (12):
  • [4] Targeting glucose control in preterm infants: pilot studies of continuous glucose monitoring
    Thomson, Lynn
    Elleri, Daniela
    Bond, Simon
    Howlett, James
    Dunger, David B.
    Beardsall, Kathryn
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2019, 104 (04): : F353 - F359
  • [5] Continuous Glucose Monitoring in Very Preterm Infants: A Randomized Controlled Trial
    Galderisi, Alfonso
    Facchinetti, Andrea
    Steil, Garry M.
    Ortiz-Rubio, Paulina
    Cavallin, Francesco
    Tamborlane, William V.
    Baraldi, Eugenio
    Cobelli, Claudio
    Trevisanuto, Daniele
    [J]. PEDIATRICS, 2017, 140 (04)
  • [6] Feasibility of automated insulin delivery guided by continuous glucose monitoring in preterm infants
    Beardsall, Kathryn
    Thomson, Lynn
    Elleri, Daniela
    Dunger, David B.
    Hovorka, Roman
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2020, 105 (03): : F279 - F284
  • [7] Continuous glucose monitoring in preterm infants: evaluation by a modified Clarke error grid
    Eloisa Tiberi
    Francesco Cota
    Giovanni Barone
    Alessandro Perri
    Valerio Romano
    Rossella Iannotta
    Costantino Romagnoli
    Enrico Zecca
    [J]. Italian Journal of Pediatrics, 42
  • [8] Procedural Pain during Insertion of a Continuous Glucose Monitoring Device in Preterm Infants
    Galderisi, Alfonso
    Lago, Paola
    Steil, Garry M.
    Ghirardo, Martina
    Cobelli, Claudio
    Baraldi, Eugenio
    Trevisanuto, Daniele
    [J]. JOURNAL OF PEDIATRICS, 2018, 200 : 261 - +
  • [9] NEONATAL HYPOGLYCEMIA CONTINUOUS GLUCOSE MONITORING: A RANDOMIZED CONTROLLED TRIAL IN PRETERM INFANTS
    Galderisi, A.
    Facchinetti, A.
    Steil, G. M.
    Ortiz-Rubio, P.
    Cobelli, C.
    Trevisanuto, D.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 : A57 - A57
  • [10] Continuous glucose monitoring in preterm infants: evaluation by a modified Clarke error grid
    Tiberi, Eloisa
    Cota, Francesco
    Barone, Giovanni
    Perri, Alessandro
    Romano, Valerio
    Iannotta, Rossella
    Romagnoli, Costantino
    Zecca, Enrico
    [J]. ITALIAN JOURNAL OF PEDIATRICS, 2016, 42