Remote control of implanted devices through Home Monitoring™ technology improves detection and clinical management of atrial fibrillation

被引:114
|
作者
Ricci, Renato Pietro [1 ]
Morichelli, Loredana [1 ]
Santini, Massimo [1 ]
机构
[1] San Filippo Neri Hosp, Dept Cardiol, I-00135 Rome, Italy
来源
EUROPACE | 2009年 / 11卷 / 01期
关键词
CARDIOVERTER-DEFIBRILLATORS; PACEMAKER; BRADYCARDIA; STROKE; CARE; MORTALITY; WIRELESS; IMPACT; RISK;
D O I
10.1093/europace/eun303
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To evaluate the impact of Home Monitoring (TM) (HM) technology on detection and treatment of atrial fibrillation (AF). One hundred and sixty-six patients (mean age 73 +/- 10 years, 112 males) received HM devices [121 pacemakers, 22 implantable cardioverter defibrillators (ICDs), and 23 ICDs combined with cardiac resynchronization therapy]. A nurse consulted the cardioreports on the website and submitted the alerts for AF (new onset AF, persistent AF, and AF burden of > 10% for > 5 consecutive days) to the physician. During a mean follow-up of 488 +/- 203 days, 42 patients (26%) had alerts for AF. Twenty-two patients had no history of AF before implant. In nine patients, no further action was taken. For the remaining 33 patients, an unscheduled follow-up was performed: in 16, antiarrhythmic drug therapy was introduced/modified; in 15, anticoagulation was started; in 2, antiplatelet drugs were introduced; in 7, an external cardioversion was performed; in 2 patients, the device was reprogrammed to avoid intermittent atrial undersensing during AF; and in 4 patients, no further action was taken. In four cases the arrhythmia was not confirmed (false positive). The median time to the first intervention for AF was 50 days (148 days before the scheduled follow-up). The HM technology allowed early detection of AF in paced patients and early reaction to optimize medical treatment.
引用
收藏
页码:54 / 61
页数:8
相关论文
共 50 条
  • [1] Remote control of implanted devices through home monitoring technology improves detection and clinical management of atrial fibrillation
    Ricci, R. P.
    Morichelli, L.
    Santini, M.
    EUROPEAN HEART JOURNAL, 2008, 29 : 115 - 116
  • [2] Cost-effectiveness of early detection of atrial fibrillation via remote control of implanted devices
    Lorenzoni, Giulia
    Folino, Franco
    Soriani, Nicola
    Iliceto, Sabino
    Gregori, Dario
    JOURNAL OF EVALUATION IN CLINICAL PRACTICE, 2014, 20 (05) : 570 - 577
  • [3] REMOTE CONTROL OF IMPLANTED ELECTRONIC DEVICES THROUGH HOME MONITORING TECHNOLOGY IN THE ERA OF COVID-19: OUR EXPERIENCE IN REMOTE ENROLLMENT
    Fais, L.
    Floris, R.
    Frau, B.
    Farris, F.
    Angei, A.
    Racis, M.
    Giau, G.
    Delogu, G.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2021, 23 (0C) : C111 - C111
  • [4] Detection and management of atrial fibrillation using remote monitoring
    Hickey, Kathleen T.
    Riga, Teresa C.
    Mitha, Shazia A.
    Reading, Meghan J.
    NURSE PRACTITIONER, 2018, 43 (03): : 24 - 30
  • [5] Improving Atrial Fibrillation Detection in Patients with Implantable Cardiac Devices by Means of a Remote Monitoring and Management Application
    Zoppo, Franco
    Facchin, Domenico
    Molon, Giulio
    Zanotto, Gabriele
    Catanzariti, Domenico
    Rossillo, Antonio
    Baccillieri, Maria Stella
    Menard, Cecile
    Comisso, Jennifer
    Gentili, Alessandra
    Grammatico, Andrea
    Bertaglia, Emanuele
    Proclemer, Alessandro
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (12): : 1610 - 1618
  • [6] Management of atrial fibrillation—What are the possibilities of early detection with Home Monitoring?
    R. P. Ricci
    M. Russo
    M. Santini
    Clinical Research in Cardiology, 2006, 95 : iii10 - iii16
  • [7] Management of atrial fibrillation - What are the possibilities of early detection with Home Monitoring?
    Ricci, RP
    Russo, M
    Santini, M
    CLINICAL RESEARCH IN CARDIOLOGY, 2006, 95 (Suppl 3) : 10 - 16
  • [8] Disease management: atrial fibrillation and Home Monitoring
    Ricci, Renato Pietro
    EUROPACE, 2013, 15 : 35 - 39
  • [9] Home monitoring improves detection of atrial tachyarrhythmia
    Russo, M
    Zabel, M
    Morais, C
    Wunderlich, E
    Hartmann, A
    Winter, S
    Nürnberg, M
    Santini, M
    EUROPEAN HEART JOURNAL, 2004, 25 : 166 - 166
  • [10] PAROXYSMAL ATRIAL FIBRILLATION PICTURED BY IMPLANTABLE DEVICES WITH REMOTE MONITORING
    Varma, N.
    JOURNAL OF EPIDEMIOLOGY AND COMMUNITY HEALTH, 2011, 65 : A415 - A415