A sealless centrifugal blood pump with passive magnetic and hydrodynamic bearings

被引:23
|
作者
Wampler, R [1 ]
Lancisi, D [1 ]
Indravudh, V [1 ]
Gauthier, R [1 ]
Fine, R [1 ]
机构
[1] Kriton Med Inc, Citrus Hts, CA 95610 USA
关键词
magnetically suspended centrifugal pump; ventricular assist device; axial flux gap motor; radial magnetic bearing; hydrodynamic bearings;
D O I
10.1046/j.1525-1594.1999.06422.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We are developing a permanently implantable ventricular assist system based on a sealless centrifugal blood pump. The impeller of the pump is supported by a passive radial magnetic bearing acting in synergy with hydrodynamic bearings. Torque is transmitted to the impeller by electromagnetic coupling via an integrated axial flux gap motor. Computer modeling has been used extensively to guide the hydraulic and electromagnetic design of the pump. As part of the development effort, a prototype system was built, which consisted of a radial magnetic bearing, an axial air gap motor, and a pivot bearing to constrain the axial motion. The following testing has been completed to validate the design. First, hydraulic tests have demonstrated sufficient hydraulic performance. Second, preliminary in vitro evaluation of hemolysis was low compared to that of a BioPump control. Third, a 6 h in vivo experiment was successfully completed.
引用
收藏
页码:780 / 784
页数:5
相关论文
共 50 条
  • [1] A sealless centrifugal blood pump with passive magnetic and hydrodynamic bearings
    Kriton Medical, Inc., Citrus Heights, CA, United States
    不详
    [J]. Artif. Organs, 8 (780-784):
  • [2] Hybrid magnetic bearings for a centrifugal blood pump
    Ren, Zhaohui
    Jahanmir, Said
    Heshmat, Hooshang
    Walton, James
    [J]. PROCEEDINGS OF THE ASME/STLE INTERNATIONAL JOINT TRIBOLOGY CONFERENCE, PTS A AND B, 2008, : 997 - 999
  • [3] Design of a Small Centrifugal Blood Pump With Magnetic Bearings
    Jahanmir, Said
    Hunsberger, Andrew Z.
    Ren, Zhaohui
    Heshmat, Hooshang
    Heshmat, Crystal
    Tomaszewski, Michael J.
    Walton, James F.
    [J]. ARTIFICIAL ORGANS, 2009, 33 (09) : 714 - 726
  • [4] A self-bearing centrifugal blood pump based on induction motor with active and passive magnetic bearings
    Wu, P
    Tseng, KJ
    [J]. PEDS 2003 : FIFTH INTERNATIONAL CONFERENCE ON POWER ELECTRONICS AND DRIVE SYSTEMS, VOLS 1 AND 2, PROCEEDINGS, 2003, : 1642 - 1646
  • [5] Improvement of hemolysis in a centrifugal blood pump with hydrodynamic bearings and semi-open impeller
    Kosaka, Ryo
    Yamane, Takashi
    Maruyama, Osamu
    Nishida, Masahiro
    Yada, Toru
    Saito, Sakae
    Hirai, Shusaku
    [J]. 2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 3982 - 3985
  • [6] Implantable centrifugal pump with hybrid magnetic bearings
    Bearnson, GB
    Olsen, DB
    Khanwilkar, PS
    Long, JW
    Sinnott, M
    Kumar, A
    Allaire, PE
    Baloh, M
    Decker, J
    [J]. ASAIO JOURNAL, 1998, 44 (05) : M733 - M736
  • [7] Implantable centrifugal pump with hybrid magnetic bearings
    Bearnson, Gill B.
    Olsen, Don B.
    Khanwilkar, Pratap S.
    Long, James W.
    Sinnott, Mary
    Kumar, Ajit
    Allaire, Paul E.
    Baloh, Michael
    Decker, Jeff
    [J]. ASAIO Journal, 44 (05):
  • [8] A New Magnetic Sealless Coupling Axial Flow Blood Pump
    LIN Chang-yan
    [J]. Chinese Journal of Biomedical Engineering, 2003, (03) : 133 - 138
  • [9] Development of a compact, sealless, tripod supported, magnetically driven centrifugal blood pump
    Yuhki, A
    Nogawa, M
    Takatani, S
    [J]. ARTIFICIAL ORGANS, 2000, 24 (06) : 501 - 505
  • [10] BLDC-control of forces in passive magnetic bearings in blood pump
    Stepien, M.
    Grzesik, B.
    [J]. 2012 15TH INTERNATIONAL POWER ELECTRONICS AND MOTION CONTROL CONFERENCE (EPE/PEMC), 2012,