Assessing the Firing Properties of the Electrically Stimulated Auditory Nerve Using a Convolution Model

被引:12
|
作者
Strahl, Stefan B. [1 ]
Ramekers, Dyan [3 ]
Nagelkerke, Marjolijn M. B. [2 ]
Schwarz, Konrad E. [1 ]
Spitzer, Philipp [1 ]
Klis, Sjaak F. L. [3 ]
Grolman, Wilko [3 ]
Versnel, Huib [3 ]
机构
[1] R&D MED EL GmbH, Innsbruck, Austria
[2] Univ Med Ctr Utrecht, Dept Otorhinolaryngol & Head & Neck Surg, Utrecht, Netherlands
[3] Univ Med Ctr Utrecht, Brain Ctr Rudolf Magnus, Dept Otorhinolaryngol & Head & Neck Surg, Utrecht, Netherlands
关键词
Auditory nerve; eCAP; Cochlear implant; Deconvolution; Firing probability; Neural health; COMPOUND ACTION-POTENTIALS; UNIT RESPONSE; SINGLE-FIBER; PHYSIOLOGICAL-PROPERTIES; ROUND-WINDOW; GUINEA-PIG; RECORDINGS; SYSTEM;
D O I
10.1007/978-3-319-25474-6_16
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
The electrically evoked compound action potential (eCAP) is a routinely performed measure of the auditory nerve in cochlear implant users. Using a convolution model of the eCAP, additional information about the neural firing properties can be obtained, which may provide relevant information about the health of the auditory nerve. In this study, guinea pigs with various degrees of nerve degeneration were used to directly relate firing properties to nerve histology. The same convolution model was applied on human eCAPs to examine similarities and ultimately to examine its clinical applicability. For most eCAPs, the estimated nerve firing probability was bimodal and could be parameterised by two Gaussian distributions with an average latency difference of 0.4 ms. The ratio of the scaling factors of the late and early component increased with neural degeneration in the guinea pig. This ratio decreased with stimulation intensity in humans. The latency of the early component decreased with neural degeneration in the guinea pig. Indirectly, this was observed in humans as well, assuming that the cochlear base exhibits more neural degeneration than the apex. Differences between guinea pigs and humans were observed, among other parameters, in the width of the early component: very robust in guinea pig, and dependent on stimulation intensity and cochlear region in humans. We conclude that the deconvolution of the eCAP is a valuable addition to existing analyses, in particular as it reveals two separate firing components in the auditory nerve.
引用
收藏
页码:143 / 153
页数:11
相关论文
共 50 条
  • [1] Development of a Model of the Electrically Stimulated Auditory Nerve
    Nogueira, W.
    Wuerfel, W.
    Buechner, A.
    [J]. BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 : S757 - +
  • [2] A Phenomenological Model of the Electrically Stimulated Auditory Nerve Fiber: Temporal and Biphasic Response Properties
    Horne, Colin D. F.
    Sumner, Christian J.
    Seeber, Bernhard U.
    [J]. FRONTIERS IN COMPUTATIONAL NEUROSCIENCE, 2016, 10
  • [3] Simulation of the electrically stimulated auditory nerve
    Rattay, F
    [J]. ARTIFICIAL ORGANS, 1997, 21 (03) : 213 - 215
  • [4] Amplitude modulation discrimination in a model of the electrically stimulated auditory nerve
    Joshua H Goldwyn
    Eric Shea-Brown
    [J]. BMC Neuroscience, 9 (Suppl 1)
  • [5] Measuring the refractoriness of the electrically stimulated auditory nerve
    Morsnowski, Andre
    Charasse, Basile
    Collet, Lionel
    Killian, Matthijs
    Mueller-Deile, Joachim
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2006, 11 (06) : 389 - 402
  • [6] Refractory behaviour of the electrically stimulated auditory nerve
    Morsnowski, A.
    Charasse, B.
    Collet, L.
    Killian, M.
    Mueller-Deile, J.
    [J]. HNO, 2008, 56 (02) : 131 - 138
  • [7] Stochastic resonance in the electrically stimulated auditory nerve: Predictions using a stochastic model of neural responsiveness
    Xu, YF
    Collins, L
    [J]. FLUCTUATIONS AND NOISE IN BIOLOGICAL, BIOPHYSICAL, AND BIOMEDICAL SYSTEMS, 2003, 5110 : 77 - 91
  • [8] A stochastic model of the electrically stimulated auditory nerve: Pulse-train response
    Bruce, IC
    Irlicht, LS
    White, MW
    O'Leary, SJ
    Dynes, S
    Javel, E
    Clark, GM
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (06) : 630 - 637
  • [9] A stochastic model of the electrically stimulated auditory nerve: Single-pulse response
    Bruce, IC
    White, MW
    Irlicht, LS
    O'Leary, SJ
    Dynes, S
    Javel, E
    Clark, GM
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (06) : 617 - 629
  • [10] A Phenomenological Model Reproducing Temporal Response Characteristics of an Electrically Stimulated Auditory Nerve Fiber
    Takanen, Marko