Temporal Trends in Strut-Level Optical Coherence Tomography Evaluation of Coronary Stent Coverage: A Systematic Review and Meta-Analysis

被引:12
|
作者
Lee, Kwan S. [1 ]
Lee, Justin Z. [1 ]
Hsu, Chiu-Hsieh [1 ]
Husnain, Muhammad [1 ]
Riaz, Haris [2 ]
Bin Riaz, Irbaz [1 ]
Thai, Hoang [1 ]
Cassese, Salvatore [3 ]
Finn, Aloke [4 ]
Samady, Habib [4 ]
Byrne, Robert A. [3 ]
机构
[1] Univ Arizona, Tucson, AZ USA
[2] Cleveland Clin, Cleveland, OH 44106 USA
[3] Tech Univ Munich, Deutsch Herzzentrum Munchen, Munich, Germany
[4] Emory Univ, Atlanta, GA 30322 USA
关键词
stents; coronary disease; meta-analysis; DRUG-ELUTING STENTS; THROMBOSIS; RESTENOSIS;
D O I
10.1002/ccd.26374
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectivesWe sought to pool data from all studies with reported strut-level data in human subjects evaluated by optical coherence tomography (OCT) surveillance and to compare the aggregate data of stent strut coverage on a longitudinal temporal timeline from initial implantation for different coronary stent subtypes. BackgroundDelayed strut coverage following drug-eluting stent (DES) implantation is an important contributor to late stent thrombosis (LST). OCT can detect stent strut coverage. MethodsWe conducted a systematic search of published or presented studies reporting OCT stent strut coverage of bare-metal stents (BMS) and DES in PubMed, Embase, Web of Science, and the Cochrane Central Register of Controlled Trials to June 2014. Data from 3,515 patients with strut-level data on 731,761 struts were analyzed. ResultsThe temporal kinetics of strut coverage differed according to stent type. The rate of strut coverage, reflected by the calculated regression coefficient, was found to be the highest in BMS, followed by early generation zotarolimus-eluting stent, everolimus-eluting stent, newer-generation zotarolimus-eluting stent, paclitaxel-eluting stent, and sirolimus-eluting stent (p<0.0001). ConclusionsAggregate rates of stent strut coverage assessed by OCT surveillance differed according to stent type. The clinical implications of these differences require further study but may underlie the differences in rates of stent thrombosis observed in clinical trials with different stent types. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:1083 / 1093
页数:11
相关论文
共 50 条
  • [1] Strut Level Optical Coherence Tomography Evaluation of Coronary Stent Strut Coverage Temporal Trends: A Systematic Review
    Lee, Kwan S.
    Lee, Justin Z.
    Riaz, Irbaz B.
    Husnain, Muhammad
    Riaz, Haris
    Lotun, Kapil
    Kern, Karl B.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) : B255 - B256
  • [2] TEMPORAL TRENDS IN STENT COVERAGE OF BIORESORBABLE POLYMER AND DURABLE POLYMER DRUG-ELUTING STENTS: A META-ANALYSIS OF STRUT-LEVEL OPTICAL COHERENCE TOMOGRAPHY DATA
    Koester, Stefan
    Lee, Justin
    Lee, Chun-Yu
    Hsu, Chiu-Hsieh
    Dhakal, Bishnu
    Dinh, Huy
    Ranga, Asri
    Lee, Kwan
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (09) : 1306 - 1306
  • [3] Assessment of Heterogeneity of Strut-Level Neointima Hyperplasia After Stent Implantation by Optical Coherence Tomography
    Kyono, Hiroyuki
    Guagliumi, Giulio
    Sirbu, Vasile
    Rosenthal, Noah
    Tahara, Satoko
    Musumeci, Giuseppe
    Fiocca, Luigi
    Valsecchi, Orazio
    Simon, Daniel
    Bezerra, Hiram G.
    Costa, Marco A.
    [J]. CIRCULATION, 2009, 120 (18) : S922 - S922
  • [4] Optical coherence tomography (OCT) strut-level analysis of drug-eluting stents (DES) in human coronary bifurcations
    Kyono, Hiroyuki
    Guagliumi, Giulio
    Sirbu, Vasile
    Rosenthal, Noah
    Tahara, Satoko
    Musumeci, Giuseppe
    Trivisonno, Antonio
    Bezerra, Hiram G.
    Costa, Marco A.
    [J]. EUROINTERVENTION, 2010, 6 (01) : 69 - 77
  • [5] Optical coherence tomography versus angiography and intravascular ultrasound to guide coronary stent implantation: A systematic review and meta-analysis
    Siddiqi, Tariq Jamal
    Khan, Muhammad Shahzeb
    Galougahi, Keyvan Karimi
    Shlofmitz, Evan
    Moses, Jeffrey W.
    Rao, Sunil
    West, Nick E. J.
    Wolff, Eric
    Hochler, Jason
    Chau, Karen
    Khalique, Omar
    Shlofmitz, Richard A.
    Jeremias, Allen
    Ali, Ziad A.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2022, 100 : S44 - S56
  • [6] Automated stent strut coverage and apposition analysis of in-vivo intra coronary Optical Coherence Tomography images
    Ughi, Giovanni J.
    Adriaenssens, Tom
    Onsea, Kevin
    Kayaert, Peter
    Dubois, Christophe
    Coosemans, Mark
    Sinnaeve, Peter
    Desmet, Walter
    D'hooge, Jan
    [J]. OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XV, 2011, 7889
  • [7] Automated measurement of stent strut coverage in intravascular optical coherence tomography
    Chi Young Ahn
    Byeong-Keuk Kim
    Myeong-Ki Hong
    Yangsoo Jang
    Jung Heo
    Chulmin Joo
    Jin Keun Seo
    [J]. Journal of the Korean Physical Society, 2015, 66 : 558 - 570
  • [8] Optical coherence tomography endpoints in stent clinical investigations: strut coverage
    Tahara, Satoko
    Chamie, Daniel
    Baibars, Motaz
    Alraies, Chadi
    Costa, Marco
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (02): : 271 - 287
  • [9] Optical coherence tomography endpoints in stent clinical investigations: strut coverage
    Satoko Tahara
    Daniel Chamié
    Motaz Baibars
    Chadi Alraies
    Marco Costa
    [J]. The International Journal of Cardiovascular Imaging, 2011, 27 : 271 - 287
  • [10] Automated Measurement of Stent Strut Coverage in Intravascular Optical Coherence Tomography
    Ahn, Chi Young
    Kim, Byeong-Keuk
    Hong, Myeong-Ki
    Jang, Yangsoo
    Heo, Jung
    Joo, Chulmin
    Seo, Jin Keun
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (04) : 558 - 570