Stereotactic CT-Guided Percutaneous Microwave Ablation of Liver Tumors With the Use of High-Frequency Jet Ventilation: An Accuracy and Procedural Safety Study

被引:71
|
作者
Engstrand, Jennie [1 ]
Toporek, Grzegorz [2 ]
Harbut, Piotr [1 ]
Jonas, Eduard [1 ]
Nilsson, Henrik [1 ]
Freedman, Jacob [1 ]
机构
[1] Danderyd Hosp AB, Danderyd Hosp, Dept Clin Sci, Karolinska Inst, SE-18288 Stockholm, Sweden
[2] Univ Bern, ARTORG Ctr Biomed Engn Res, Bern, Switzerland
关键词
accuracy; interventional radiology; liver; microwave ablation; stereotaxy; RADIOFREQUENCY ABLATION; HEPATIC-TUMORS; HEPATOCELLULAR-CARCINOMA; NAVIGATION SYSTEM; THERMAL ABLATION; GUIDANCE; PHANTOM; BIOPSY; THERMOABLATION; FEASIBILITY;
D O I
10.2214/AJR.15.15803
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of the present study is to evaluate the accuracy and safety of antenna placement performed with the use of a CT-guided stereotactic navigation system for percutaneous ablation of liver tumors and to assess the safety of high-frequency jet ventilation for target motion control. MATERIALS AND METHODS. Twenty consecutive patients with malignant liver lesions for which surgical resection was contraindicated or that were not readily visible on ultrasound or not accessible by ultrasound guidance were included in the study. Patients were treated with percutaneous microwave ablation performed using a CT-guided stereotactic navigation system. High-frequency jet ventilation was used to reduce liver motion during all interventions. The accuracy of antenna placement, the number of needle readjustments required, overall safety, and the radiation doses were assessed. RESULTS. Microwave ablation was completed for 20 patients (28 lesions). Performance data could be evaluated for 17 patients with 25 lesions (mean [+/- SD] lesion diameter, 14.9 +/- 5.9 mm; mean lesion location depth, 87.5 +/- 27.3 mm). The antennae were placed with a mean lateral error of 4.0 +/- 2.5 mm, a depth error of 3.4 +/- 3.2 mm, and a total error of 5.8 +/- 3.2 mm in relation to the intended target. The median number of antenna readjustments required was zero (range, 0-1 adjustment). No major complications were related to either the procedure or the use of high-frequency jet ventilation. The mean total patient radiation dose was 957.5 +/- 556.5 mGy x cm, but medical personnel were not exposed to irradiation. CONCLUSION. Percutaneous microwave ablation performed with CT-guided stereotactic navigation provides sufficient accuracy and requires almost no repositioning of the -needle. Therefore, it is technically feasible and applicable for safe treatments.
引用
收藏
页码:193 / 200
页数:8
相关论文
共 50 条
  • [11] CT-Guided Microwave Ablation of Liver Tumors in Anatomically Challenging Locations
    Van Dai Vo Chieu
    Thomas Werncke
    Bennet Hensen
    Frank Wacker
    Kristina I. Ringe
    CardioVascular and Interventional Radiology, 2018, 41 : 1520 - 1529
  • [12] CT-guided Percutaneous Microwave Ablation of Liver Metastases from Nasopharyngeal Carcinoma
    Li, Xin
    Fan, Wei-jun
    Zhang, Liang
    Zhang, Xiao-ping
    Jiang, Hua
    Zhang, Jian-lei
    Zhang, Hao
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (05) : 680 - 684
  • [13] Safety of High-Frequency Jet Ventilation During Image-Guided Thermal Ablation Procedures
    Théophane Trochu
    Nathalie Desfriches-Doria
    Nicolas Grillot
    Fanny Feuillet
    David Lair
    Renan Liberge
    Frédéric Douane
    Romain Dumont
    Arthur David
    CardioVascular and Interventional Radiology, 2023, 46 : 360 - 368
  • [14] Safety of High-Frequency Jet Ventilation During Image-Guided Thermal Ablation Procedures
    Trochu, Theophane
    Desfriches-Doria, Nathalie
    Grillot, Nicolas
    Feuillet, Fanny
    Lair, David
    Liberge, Renan
    Douane, Frederic
    Dumont, Romain
    David, Arthur
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2023, 46 (03) : 360 - 368
  • [15] Accuracy and diagnostic yield of CT-guided stereotactic liver biopsy of primary and secondary liver tumors
    Schullian, Peter
    Widmann, Gerlig
    Lang, Thomas Bob
    Knoflach, Martin
    Bale, Reto
    COMPUTER AIDED SURGERY, 2011, 16 (04) : 181 - 187
  • [16] CT-guided microwave ablation through the lungs for treating liver tumors near the diaphragm
    Qi, Han
    Zhang, Hao
    Wan, Chao
    Xie, Lin
    Song, Ze
    Fan, Weijun
    ONCOTARGET, 2017, 8 (45) : 79270 - 79278
  • [17] CT-guided Percutaneous Microwave Ablation Combined with Local Radiotherapy or Chemotherapy of Malignant Pulmonary Tumors
    Xu, Rongde
    Chen, Jingjing
    Chen, Daohua
    Zhang, Xiaobo
    Cui, Wei
    Deng, Yi
    Sun, Danxiong
    Yuan, Bing
    Li, Jing
    CURRENT RADIOPHARMACEUTICALS, 2024, 17 (02) : 184 - 199
  • [18] Efficacy and safety of artificial pneumothorax with position adjustment for CT-guided percutaneous transthoracic microwave ablation of small subpleural lung tumors
    Jia, Haipeng
    Tian, Jie
    Liu, Bo
    Meng, Hong
    Pan, Fengmin
    Li, Chunhai
    THORACIC CANCER, 2019, 10 (08) : 1710 - 1716
  • [19] High-Frequency Jet Ventilation Versus Spontaneous Respiration for Percutaneous Cryoablation of Lung Tumors: Comparison of Adverse Events and Procedural Efficiency
    Graur, Alexander
    Mercaldo, Nathaniel D.
    Simon, Judit
    Alici, Cagatay
    Saenger, Jonathan A.
    Cahalane, Alexis M.
    Vazquez, Rafael
    Fintelmann, Florian J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (04)
  • [20] Percutaneous lung and liver CT-guided ablation on swine model using microwave ablation to determine ablation size for clinical practice
    Habert, Paul
    Di Bisceglie, Mathieu
    Hak, Jean-Francois
    Brige, Pauline
    Chopinet, Sophie
    Mancini, Julien
    Bartoli, Axel
    Vidal, Vincent
    Roux, Charles
    Tselikas, Lambros
    De Baere, Thierry
    Gaubert, Jean-Yves