Development, Test and Validation of a Mechatronic Device for Spasticity Quantification

被引:4
|
作者
Ferreira, Joao F. [1 ]
Moreira, Vitor [2 ]
Lima, Carlos S. [1 ]
Soares, Filomena [1 ]
Machado, Jose [3 ]
Leao, Celina P. [4 ]
机构
[1] Univ Minho, Ind Elect Dept, ALGORITMI Res Ctr, Guimaraes, Portugal
[2] Hosp Braga, Braga, Portugal
[3] Univ Minho, Dept Mech Engn, Res Ctr CT2M, Guimaraes, Portugal
[4] Univ Minho, Prod & Syst Dept, ALGORITMI Res Ctr, Guimaraes, Portugal
关键词
Mechatronic Device; Spasticity Quantification; Improvement of Quality of Life; Human Safety; STRETCH REFLEX THRESHOLD; PATHOPHYSIOLOGY; ELBOW;
D O I
10.5772/56334
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
Spasticity is a common pathological phenomenon in clinical practices, frequently expressed as a stroke, multiple sclerosis or cerebral palsy. The accurate quantification of spasticity allows early action to prevent the potentially irreversible consequences. The aim of this work is to develop and implement a mechatronic device for the accurate quantification of spasticity, which is able to differentiate spasticity from other motor disorders. The proposed method is based on the quantification of the tonic stretch reflex threshold (TSRT) for the assessment of the range of motion of the limb affected by spasticity. In order to validate the developed device, experimental trials have been conducted, considering the flexor muscle of the elbow joint. The developed device was tested, first in a laboratory environment with healthy subjects and secondly, in a clinical environment with the collaboration of patients with spasticity. The evaluations in th clinical environment were performed on three different days in order to evaluate the reproducibility of the obtained results. The experimental trials confirm the sensibility, reproducibility and reliability of spasticity quantification based on the TSRT method. This project has been developed in cooperation with the Hospital of Braga and Fisimaia rehabilitation clinic, both in Portugal.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A Mechatronic Device for Spasticity Quantification
    Ferreira, Joao
    Moreira, Vitor
    Machado, Jose
    Soares, Filomena
    [J]. 2011 9TH IEEE INTERNATIONAL CONFERENCE ON INDUSTRIAL INFORMATICS (INDIN), 2011,
  • [2] Improved Biomedical Device for Spasticity Quantification
    Ferreira, Joao
    Moreira, Vitor
    Machado, Jose
    Soares, Filomena
    [J]. 2013 IEEE 3RD PORTUGUESE MEETING IN BIOENGINEERING (ENBENG), 2013,
  • [3] Small-Size Robot Platform as Test and Validation Tool for the Development of Mechatronic Systems
    Kemper, Philipp
    Tetzlaff, Thomas
    Witkowski, Ulf
    Zandian, Reza
    Mamrot, Michel
    Marchlewitz, Stefan
    Nicklas, Jan-Peter
    Winzer, Petra
    [J]. ROBOT INTELLIGENCE TECHNOLOGY AND APPLICATIONS 4, 2017, 447 : 383 - 396
  • [4] Development and validation of Spasticity IndexAmyotrophic Lateral Sclerosis
    Milinis, K.
    Tennant, A.
    Mills, R. J.
    Al-Chalabi, A.
    Burke, G.
    Dick, D. J.
    Ealing, J.
    Hanemann, C. O.
    Harrower, T.
    McDermott, C. J.
    Majeed, T.
    Pinto, A.
    Talbot, K.
    Williams, T.
    Young, C. A.
    [J]. ACTA NEUROLOGICA SCANDINAVICA, 2018, 138 (01): : 47 - 54
  • [5] QUANTIFICATION OF SPASTICITY
    KNUTSSON, E
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1983, 56 (03): : S17 - S17
  • [6] Development and Earliest Validation of a Portable Device for Quantification of Hallux Extension Strength (QuHalEx)
    Hile, Elizabeth S.
    Ghazi, Mustafa
    Chandrashekhar, Raghuveer
    Rippetoe, Josiah
    Fox, Ashley
    Wang, Hongwu
    [J]. SENSORS, 2023, 23 (10)
  • [7] Development and validation of ultraviolet spectrophotometric methods for lamivudine and zidovudine quantification in dissolution test
    Perez-Chauca, Enma
    Alva-Plasencia, Pedro M.
    Ferraz, Humberto Gomes
    [J]. JOURNAL OF PHARMACY & PHARMACOGNOSY RESEARCH, 2022, 10 (05): : 847 - 856
  • [8] Validation for tremor quantification of an electromagnetic tracking device
    O'Suilleabhain, PE
    Dewey, RB
    [J]. MOVEMENT DISORDERS, 2001, 16 (02) : 265 - 271
  • [9] Biomedical Device for Spasticity Quantification Based on the Velocity Dependence of the Stretch Reflex Threshold
    Ferreira, Joao
    Machado, Jose
    Moreira, Vitor
    Soares, Filomena
    [J]. 2011 IEEE 16TH CONFERENCE ON EMERGING TECHNOLOGIES AND FACTORY AUTOMATION (ETFA), 2011,
  • [10] Model-based Test and Validation Support for Autonomous Mechatronic Systems
    Mamrot, Michel
    Marchlewitz, Stefan
    Nicklas, Jan-Peter
    Winzer, Petra
    Tetzlaff, Thomas
    Kemper, Philipp
    Witkowski, Ulf
    [J]. 2015 IEEE INTERNATIONAL CONFERENCE ON SYSTEMS, MAN, AND CYBERNETICS (SMC 2015): BIG DATA ANALYTICS FOR HUMAN-CENTRIC SYSTEMS, 2015, : 701 - 706