The effect of gradient sampling schemes on measures derived from diffusion tensor MRI: A Monte Carlo study

被引:610
|
作者
Jones, DK [1 ]
机构
[1] NICHHD, Sect Tissue Biophys & Biomimet, Lab Integrat Med & Biophys, NIH, Bethesda, MD 20892 USA
关键词
sampling schemes; electrostatic respulsion; cone of uncertainty; trace; anisotropy; rotational invariance;
D O I
10.1002/mrm.20033
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
There are conflicting opinions in the literature as to whether it is more beneficial to use a large number of gradient sampling orientations in diffusion tensor MRI (DT-MRI) experiments than to use a smaller number of carefully chosen orientations. In this study, Monte Carlo simulations were used to study the effect of using different gradient sampling schemes on estimates of tensor-derived quantities assuming a b-value of 1000 smm(-2). The study focused in particular on the effect that the number of unique gradient orientations has on uncertainty in estimates of tensor-orientation, and on estimates of the trace and anisotropy of the diffusion tensor. The results challenge the recently proposed notion that a set of six icosahedrally-arranged orientations is optimal for DT-MRI. It is shown that at least 20 unique sampling orientations are necessary for a robust estimation of anisotropy, whereas at least 30 unique sampling orientations are required for a robust estimation of tensor-orientation and mean diffusivity. Finally, the performance of sampling schemes that use low numbers of sampling orientations, but make efficient use of available gradient power, are compared to less efficient schemes with larger numbers of sampling orientations, and the relevant scenarios in which each type of scheme should be used are discussed. Published 2004 Wiley-Liss, Inc.(1)
引用
收藏
页码:807 / 815
页数:9
相关论文
共 50 条
  • [1] Effects of random subject rotation on optimised diffusion gradient sampling schemes in diffusion tensor MRI
    Maniega, Susana Munoz
    Bastin, Mark E.
    Armitage, Paul A.
    MAGNETIC RESONANCE IMAGING, 2008, 26 (04) : 451 - 460
  • [2] Anisotropic noise propagation in diffusion tensor MRI sampling schemes
    Batchelor, PG
    Atkinson, D
    Hill, DLG
    Calamante, F
    Connelly, A
    MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (06) : 1143 - 1151
  • [3] Comparison of gradient encoding schemes for diffusion-tensor MRI
    Hasan, KM
    Parker, DL
    Alexander, AL
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (05) : 769 - 780
  • [4] The effect of gradient sampling schemes on diffusion metrics derived from probabilistic analysis and tract-based spatial statistics
    Hope, Tuva
    Westlye, Lars Tjelta
    Bjornerud, Atle
    MAGNETIC RESONANCE IMAGING, 2012, 20 (03) : 402 - 412
  • [5] A Monte Carlo-based fiber tracking algorithm using diffusion tensor MRI
    Prados, F.
    Bardera, A.
    Sbert, M.
    Bouda, I.
    Feixas, M.
    19TH IEEE INTERNATIONAL SYMPOSIUM ON COMPUTER-BASED MEDICAL SYSTEMS, PROCEEDINGS, 2006, : 353 - +
  • [6] The effect of Gibbs ringing artifacts on measures derived from diffusion MRI
    Perrone, Daniele
    Aelterman, Jan
    Pizurica, Aleksandra
    Jeurissen, Ben
    Philips, Wilfried
    Leemans, Alexander
    NEUROIMAGE, 2015, 120 : 441 - 455
  • [7] A Monte Carlo Study of Restricted Diffusion: Implications for Diffusion MRI of Prostate Cancer
    Gilani, Nima
    Malcolm, Paul
    Johnson, Glyn
    MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (04) : 1671 - 1677
  • [8] Robust and Fast Markov Chain Monte Carlo Sampling of Diffusion MRI Microstructure Models
    Harms, Robbert L.
    Roebroeck, Alard
    FRONTIERS IN NEUROINFORMATICS, 2018, 12
  • [9] A comparative study of acquisition schemes for diffusion tensor imaging using MRI
    Papadakis, NG
    Xing, D
    Huang, CLH
    Hall, LD
    Carpenter, TA
    JOURNAL OF MAGNETIC RESONANCE, 1999, 137 (01) : 67 - 82