Modeling of bone around hydroxyapatite-coated femoral stems

被引:105
|
作者
DAntonio, JA [1 ]
Township, M [1 ]
Capello, WN [1 ]
Manley, MT [1 ]
机构
[1] INDIANA UNIV, SCH MED, DEPT ORTHOPAED SURG, INDIANAPOLIS, IN 46202 USA
来源
关键词
D O I
10.2106/00004623-199608000-00013
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Two hundred and twenty-four total hip arthroplasties were performed in 201 patients with use of a femoral component with hydroxyapatite coating of the proximal portion of the stem, The mean duration of follow-up was seventy-one months (range, fifty-eight to eighty-seven months). Of the 224 arthroplasties, 208 (93 per cent; 190 patients) yielded a good or excellent clinical result, Four patients (2 per cent) reported mild-to-moderate activity-related pain in the thigh, and two (1 per cent) had aseptic loosening, The radiographic findings of progressive new-bone formation (cancellous condensation and cortical hypertrophy) throughout the zones adjacent to the middle and distal portions of the stein Were evidence of early, extensive proximal fixation of the implant, with distal stress transfer through the implant, which is stiffer than the surrounding bone, Remodeling of the femur began early was:predictable, and progressed throughout the follow-up period, Cortical hypertrophy about the middle and distal portions of the stem occurred predominantly in the mediolateral plane (in 105 hips [47 per cent], compared with thirteen hips [6 per cent] in the anteroposterior plane), and it was more common in patients who had had poorer bone quality preoperatively, Intramedullary osteolysis was present in one femur (0.4 per cent) at five years; the osteolytic area was less than Five millimeters in ifs greatest dimension and had not progressed at the time of the six-year follow-up evaluation, This low rate of osteolysis suggests-that a circumferential coating of hydroxyapatite may effectively minimize migration of wear debris along the femoral stem. The progressive remodeling of the femur about the middle and distal portions of the stem, as evidenced by cancellous condensation and cortical hypertrophy, has not, to our knowledge, been described previously to this magnitude in association with proximally coated (porous or hydroxyapatite-coated) femoral implants.
引用
收藏
页码:1226 / 1234
页数:9
相关论文
共 50 条
  • [1] Periprosthetic fractures around cementless hydroxyapatite-coated femoral stems
    van der Wal, BCH
    Vischjager, M
    Grimm, B
    Heyligers, IC
    Tonino, AJ
    [J]. INTERNATIONAL ORTHOPAEDICS, 2005, 29 (04) : 235 - 240
  • [2] Periprosthetic fractures around cementless hydroxyapatite-coated femoral stems
    B. C. H. van der Wal
    M. Vischjager
    B. Grimm
    I. C. Heyligers
    A. J. Tonino
    [J]. International Orthopaedics, 2005, 29 : 235 - 240
  • [3] Osteointegration of hydroxyapatite-coated stems of femoral prostheses
    D. C. R. Hardy
    P. Frayssinet
    P. E. Delince
    [J]. European Journal of Orthopaedic Surgery & Traumatology, 1999, 9 (2) : 75 - 81
  • [4] Radiographic Analysis of Bone Remodeling around Hydroxyapatite Coated Femoral Stems
    Marinescu, Rodica
    Laptoiu, Dan
    Botezatu, Iozefina
    [J]. BIOCERAMICS, VOL 25, 2014, 587 : 436 - 441
  • [5] Hydroxyapatite-coated versus grit-blasted femoral stems
    Amstutz, HC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (08): : 1206 - 1207
  • [6] Hydroxyapatite-coated femoral stems -: Histology and histomorphometry around five components retrieved at post mortem
    Tonino, AJ
    Thèrin, M
    Doyle, C
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01): : 148 - 154
  • [7] Proximal radiolucent lines around fully hydroxyapatite-coated tapered femoral stems: should we be concerned?
    Berliner, Zachary P.
    Jawad, Muhammad Umar
    Matzko, Chelsea
    Cooper, H. John
    Rodriguez, Jose A.
    Hepinstall, Matthew S.
    [J]. HIP INTERNATIONAL, 2024, 34 (01) : 49 - 56
  • [8] HYDROXYAPATITE-COATED FEMORAL STEMS - HISTOLOGICAL ANALYSIS OF COMPONENTS RETRIEVED AT AUTOPSY
    BAUER, TW
    GEESINK, RCT
    ZIMMERMAN, R
    MCMAHON, JT
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (10): : 1439 - 1452
  • [9] Hydroxyapatite-coated versus grit-blasted femoral stems - Reply
    Nizard, R
    Hamadouche, M
    Sedel, L
    Witvoet, J
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (08): : 1207 - 1207
  • [10] Hydroxyapatite-coated femoral stems - A matched-pair analysis of coated and uncoated implants
    Rothman, RH
    Hozack, WJ
    Ranawat, A
    Moriarty, L
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (03): : 319 - 324