Stability of hydroxyapatite-coatings on titanium oral implants (IMZ) - 2 retrieved cases

被引:31
|
作者
Liao, HH [1 ]
Fartash, B [1 ]
Li, JG [1 ]
机构
[1] KAROLINSKA INST, CTR ORAL BIOL NOVUM, S-14104 HUDDINGE, SWEDEN
关键词
histology; hydroxyapatite; retrieved oral implant;
D O I
10.1111/j.1600-0501.1997.tb00009.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
2 hydroxyapatite(HA)-coated titanium oral implants (IMZ) were retrieved from 2 patients after more than three years in clinical function. One implant (case I) was removed because of extensive bone resorption and loss of integration, and the other (case II) was retrieved due to fracture of the abutment in sprit of clinical rigidity. The implants and the surrounding soft (case I) and hard (case II) tissues were examined histologically and by scanning electron microscopy (SEM). In case I, the HA-coating was almost degraded, as SEM revealed a resulted rough titanium surface. Histology showed HA particles and inflammation were seen near the implant. In Case II, the HA-coating seemed to be intact and bonded to bone, but scattered HA particles, embedded in bone at a distance from the implant were also observed. The results indicate that HA-coatings may come loose during clinical function. The long-term stability of HA-coatings and the clinical indications for HA-coated implants warrant further study.
引用
收藏
页码:68 / 72
页数:5
相关论文
共 50 条
  • [1] Hydroxyapatite coatings on titanium implants
    Ciobanu, Margareta Gabriela
    Pop, Gheorghe
    Ciobanu, Octavian
    REVISTA DE CHIMIE, 2007, 58 (12): : 1313 - 1315
  • [2] Antibacterial hydroxyapatite coatings on titanium dental implants
    Ziming Liao
    Jingxuan Li
    Yimeng Su
    Fenyan Miao
    Xiumei Zhang
    Yu Gu
    Jingjing Du
    Ruiqiang Hang
    Yan Wei
    Weiyi Chen
    Di Huang
    Frontiers of Materials Science, 2023, 17
  • [3] Mechanical integrity of hydroxyapatite coatings on titanium implants
    Leng, Y
    Zhang, C
    STRUCTURAL BIOMATERIALS FOR THE 21ST CENTURY, 2001, : 221 - 228
  • [4] Antibacterial hydroxyapatite coatings on titanium dental implants
    Liao, Ziming
    Li, Jingxuan
    Su, Yimeng
    Miao, Fenyan
    Zhang, Xiumei
    Gu, Yu
    Du, Jingjing
    Hang, Ruiqiang
    Wei, Yan
    Chen, Weiyi
    Huang, Di
    FRONTIERS OF MATERIALS SCIENCE, 2023, 17 (01)
  • [5] CLINICAL CONDITION OF ORAL-MUCOSA AND BONE AT OSSEOINTEGRATED TITANIUM IMPLANTS (IMZ)
    ISIDOR, F
    KAABER, S
    JOURNAL OF DENTAL RESEARCH, 1988, 67 (04) : 769 - 769
  • [6] Bone-hydroxyapatite interface in retrieved hydroxyapatite-coated titanium implants: A clinical and histologic report
    Piattelli, A
    Scarano, A
    Di Alberti, L
    Piattelli, M
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1999, 14 (02) : 233 - 238
  • [7] SEM and EDX studies of bioactive hydroxyapatite coatings on titanium implants
    Ciobanu, Gabriela
    Carja, Gabriela
    Ciobanu, Octavian
    Sandu, Ion
    Sandu, Andrei
    MICRON, 2009, 40 (01) : 143 - 146
  • [8] A review on plasma sprayed titanium and hydroxyapatite coatings on polyetheretherketone implants
    Guner, Ali Tekin
    Meran, Cemal
    INTERNATIONAL JOURNAL OF SURFACE SCIENCE AND ENGINEERING, 2019, 13 (04) : 237 - 262
  • [9] Structure of the bone-titanium interface in retrieved clinical oral implants
    Sennerby, L.
    Ericson, L. E.
    Thomsen, P.
    Lekholm, U.
    Astrand, P.
    CLINICAL ORAL IMPLANTS RESEARCH, 1991, 2 (03) : 103 - 111
  • [10] Sol-Gel Derived Hydroxyapatite Coatings for Titanium Implants: A Review
    Jaafar, Alaa
    Hecker, Christine
    Arki, Pal
    Joseph, Yvonne
    BIOENGINEERING-BASEL, 2020, 7 (04): : 1 - 23