Retrograde approach for device closure of muscular ventricular septal defects in children and adolescents, using the amplatzer muscular ventricular septal defect occluder

被引:16
|
作者
Jameel, Al-Ata
Arfi, Amin Muhammed
Arif, Hussain
Amjad, Kouatli
Omar, Galal Mohammed
机构
[1] King Faisal Specialist Hosp & Res Ctr, Pediat Cardiol Sect, Dept Cardiovasc Dis, Jeddah 21499, Saudi Arabia
[2] Univ Pediat Hosp Essen, Essen, Germany
关键词
transcatheter closure; percutaneous device occlusion; cardiac catheterization;
D O I
10.1007/s00246-006-1365-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study presents technique and initial experience of retrograde deployment of the Amplatzer muscular ventricular septal defect occluder (AmVSDo) for closure of muscular ventricular septal defects (VSDs). The conventional technique for closing muscular VSDs involves the creation of an arteriovenous guidewire circuit and use of a transvenous approach for device deployment. Seven patients aged 2.2-15 years underwent transcatheter closure of a muscular VSD using the retrograde approach without making the arteriovenous wire circuit. Mean fluoroscopy and procedural times were compared to those previously reported in publications describing the use of the antegrade approach. Unpaired Student's t-test was used to compare the two parameters in two groups. Our technique was successful in all patients reported. The mean fluoroscopy time in the retrograde versus the antegrade group was 33.8 +/- 20.9 and 41.9 +/- 6.2 minutes, respectively (not significant), and the mean procedural time in the two groups was 91.1 +/- 22.1 and 114 +/- 33.9 minutes respectively (p = 0.025). No complications were noted. We suggest that some muscular VSDs can be safely closed retrogradely without the use of an arteriovenous loop, thus reducing the radiation exposure and also the cost of the procedure. Further studies are needed to confirm this initial experience.
引用
收藏
页码:720 / 728
页数:9
相关论文
共 50 条
  • [1] Retrograde Approach for Device Closure of Muscular Ventricular Septal Defects in Children and Adolescents, Using the Amplatzer Muscular Ventricular Septal Defect Occluder
    Al-Ata Jameel
    Amin Muhammed Arfi
    Hussain Arif
    Kouatli Amjad
    Galal Mohammed Omar
    Pediatric Cardiology, 2006, 27 : 720 - 728
  • [2] Closure of a muscular ventricular septal defect using the Amplatzer ventricular occluder
    Dehghani, Majid
    Sharifkazemi, Mohammad-Bagher
    Aslani, Amir
    Hoseini, Ehsan
    Samin, Mohammad-Mehdi
    ARCHIVES OF IRANIAN MEDICINE, 2007, 10 (04) : 543 - 545
  • [3] Outcome of transcatheter closure of muscular ventricular septal defects using the amplatzer ventricular septal defect occluder
    Thanopoulos, BD
    Rigby, ML
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (06) : 473A - 473A
  • [4] Outcome of transcatheter closure of muscular ventricular septal defects with the Amplatzer ventricular septal defect occluder
    Thanopoulos, BD
    Rigby, ML
    HEART, 2005, 91 (04) : 513 - 516
  • [5] Use of the Amplatzer muscular ventricular septal defect occluder for closure of perimembranous ventricular septal defects
    Szkutnik, M.
    Qureshi, S. A.
    Kusa, J.
    Rosenthal, E.
    Bialkowski, J.
    HEART, 2007, 93 (03) : 355 - 358
  • [6] Transcatheter closure of muscular ventricular septal defects with the Amplatzer ventricular septal defect occluder: Initial clinical applications in children
    Thanopoulos, BD
    Tsaousis, GS
    Konstadopoulou, GN
    Zarayelyan, AG
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 33 (05) : 1395 - 1399
  • [7] Device closure of muscular ventricular septal defects using the amplatzer muscular ventricular septal defect occluder - Immediate and mid-term results of a US registry
    Holzer, R
    Balzer, D
    Cao, QL
    Lock, K
    Hijazi, ZM
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (07) : 1257 - 1263
  • [8] Transcatheter closure of muscular ventricular septal defect using Amplatzer Ductal Occluder device in low weight children
    Vaujois, L.
    Tissot, C.
    Beghetti, M.
    Aggoun, Y.
    Tissot, Cecile
    SWISS MEDICAL WEEKLY, 2010, 140 (21-22) : 12S - 12S
  • [9] Transcatheter closure of a mid-muscular ventricular septal defect with an Amplatzer VSD occluder device
    Tofeig, M
    Patel, RG
    Walsh, KP
    HEART, 1999, 81 (04) : 438 - 440
  • [10] Perventricular closure of a muscular ventricular septal defect with an Amplatzer occlusive device
    Eguia, Ibon
    Crespo, Alejandro
    Hamzeh, Gadah
    Alcibar, Juan
    Ignacio Aramendi, Jose
    Miguel Galdeano, Jose
    CIRUGIA CARDIOVASCULAR, 2007, 14 (03): : 261 - 263