摘要
This paper reports on the development of an autonomous centrifugal microfluidic device, which automatically executes an enzyme-linked immunosorbent assay (ELISA) on a steady rotational frequency. The device was designed on the principles of CLOCK (Control of Liquid Operation on Centrifugal hydroKinetics) to achieve autonomous centrifugal microfluidics. These demonstrations confirmed that the device achieves accurate flow control of an automated ELISA with fluctuations of the interval time for implementation of each unit (antigen-antibody reaction, wash and injection of detection substrate) being less than 5%. The most optimal method for the immobilization of primary antibodies was investigated for the developed platform. We compared three methods of antibody immobilization on the developed platform, by using polystyrene microbeads, a polyurethane foam structure, and direct immobilization to the reaction chamber. We chose the use of polyurethane foam as the best immobilization method based on reaction efficiency and the convenient handling of the immobilized antibody on the solid phase. Measurement of human albumin was demonstrated using the developed device, and a calibration curve of human albumin detection was successfully prepared with an overall assay operation time of 18 min. The limit of detection (LOD) was estimated to be 0.516 ng/ml, which is comparable with the LOD of 0.707 ng/ml obtained by the assay on a conventional titer plate assay platform. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:264 / 270
页数:7
相关论文
共 50 条
  • [1] A lab in a bento box: an autonomous centrifugal microfluidic system for an enzyme-linked immunosorbent assay
    Abe, Takaaki
    Okamoto, Shunya
    Taniguchi, Akinobu
    Fukui, Michiyasu
    Yamaguchi, Akinobu
    Utsumi, Yuichi
    Ukita, Yoshiaki
    ANALYTICAL METHODS, 2020, 12 (40) : 4858 - 4866
  • [2] Microfluidic enzyme-linked immunosorbent assay technology
    Lee, L. James
    Yang, Shang-Tiang
    Lai, Siyi
    Bai, Yunling
    Huang, Wei-Cho
    Juang, Yi-Je
    ADVANCES IN CLINICAL CHEMISTRY, VOL 42, 2006, 42 : 255 - 295
  • [3] Automatic microdispenser-integrated multiplex enzyme-linked immunosorbent assay device with autonomously driven centrifugal microfluidic system
    Okamoto, Shunya
    Nagai, Moeto
    Shibata, Takayuki
    Ukita, Yoshiaki
    RSC ADVANCES, 2024, 14 (20) : 13827 - 13836
  • [4] Microfluidic enzyme-linked immunosorbent assay in a region of finite length
    Yanagisawa, Naoki
    Dutta, Debashis
    ANALYTICA CHIMICA ACTA, 2014, 817 : 28 - 32
  • [5] Enzyme-linked immunosorbent assay utilizing thin-layered microfluidics
    Nakao, Tatsuro
    Mawatari, Kazuma
    Kazoe, Yutaka
    Mori, Emi
    Shimizu, Hisashi
    Kitamori, Takehiko
    ANALYST, 2019, 144 (22) : 6625 - 6634
  • [6] Automated sample preparation for electrospray ionization mass spectrometry based on CLOCK-controlled autonomous centrifugal microfluidics
    Futami, Masahiro
    Naito, Hiroki
    Ninomiya, Satoshi
    Chen, Lee Chuin
    Iwano, Tomohiko
    Yoshimura, Kentaro
    Ukita, Yoshiaki
    BIOMEDICAL MICRODEVICES, 2024, 26 (02)
  • [7] Electrowetting-based Digital Microfluidics Platform for Automated Enzyme-linked Immunosorbent Assay
    Dimov, Nikolay
    McDonnell, Martin B.
    Munro, Ian
    Mccluskey, Daniel K.
    Johnston, Ian D.
    Tan, Christabel K. L.
    Coudron, Loic
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2020, (156):
  • [8] An enzyme-linked immunosorbent assay on a centrifugal platform using magnetic beads
    Shih, Chih-Hsin
    Wu, Ho-Chin
    Chang, Chong-Yi
    Huang, Wen-Hong
    Yang, Yi-Feng
    BIOMICROFLUIDICS, 2014, 8 (05):
  • [9] An optical pickup enzyme-linked immunosorbent assay (ELISA) with a microfluidic disk
    Yoshikawa, H.
    Yoshinaga, M.
    Tamiya, E.
    RSC ADVANCES, 2018, 8 (26): : 14510 - 14514
  • [10] Design and testing of a microfluidic biochip for cytokine enzyme-linked immunosorbent assay
    He, Hongyan
    Yuan, Yuan
    Wang, Weixiong
    Chiou, Nan-Rong
    Epstein, Arthur J.
    Lee, L. James
    BIOMICROFLUIDICS, 2009, 3 (02)