Experimental Platform for Ultra-high Dose Rate FLASH Irradiation of Small Animals Using a Clinical Linear Accelerator

被引:179
|
作者
Schuler, Emil [1 ]
Trovati, Stefania [1 ]
King, Gregory [1 ]
Lartey, Frederick [1 ]
Rafat, Marjan [1 ]
Villegas, Manuel [1 ]
Praxel, A. Joe [1 ]
Loo, Billy W., Jr. [1 ,2 ]
Maxim, Peter G. [1 ,2 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Sch Med, Stanford, CA 94305 USA
[2] Stanford Univ, Stanford Canc Inst, Sch Med, Stanford, CA 94305 USA
关键词
MONTE-CARLO SIMULATIONS; ENERGY ELECTRON-BEAMS; PROTON IRRADIATION; RADIATION; SURVIVAL; PHOTON; DISTRIBUTIONS; RADIOTHERAPY; DEPENDENCE; MICRORT;
D O I
10.1016/j.ijrobp.2016.09.018
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A key factor limiting the effectiveness of radiation therapy is normal tissue toxicity, and recent preclinical data have shown that ultra-high dose rate irradiation (>50 Gy/s, "FLASH") potentially mitigates this effect. However, research in this field has been strongly limited by the availability of FLASH irradiators suitable for small animal experiments. We present a simple methodologic approach for FLASH electron small animal irradiation with a clinically available linear accelerator (LINAC). Methods and Materials: We investigated the FLASH irradiation potential of a Varian Clinac 21EX in both clinical mode and after tuning of the LINAC. We performed detailed FLUKA Monte Carlo and experimental dosimetric characterization at multiple experimental locations within the LINAC head. Results: Average dose rates of <= 74 Gy/s were achieved in clinical mode, and the dose rate after tuning exceeded 900 Gy/s. We obtained 220 Gy/s at 1-cm depth for a >4-cm field size with 90% homogeneity throughout a 2-cm-thick volume. Conclusions: We present an approach for using a clinical LINAC for FLASH irradiation. We obtained dose rates exceeding 200 Gy/s after simple tuning of the LINAC, with excellent dosimetric properties for small animal experiments. This will allow for increased availability of FLASH irradiation to the general research community. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:195 / 203
页数:9
相关论文
共 50 条
  • [1] Modifying a clinical linear accelerator for delivery of ultra-high dose rate irradiation
    Lempart, Michael
    Blad, Borje
    Adrian, Gabriel
    Back, Sven
    Knoos, Tommy
    Ceberg, Crister
    Petersson, Kristoffer
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 139 : 40 - 45
  • [2] Design of an ultra-high dose rate cell irradiation experimental platform using a medical proton linear injector
    Liu, Junya
    Yu, Xiao
    Wu, Chao
    Sun, Xiangshang
    Shang, Haijiao
    Liao, Wentao
    Yan, Nan
    Zhou, Yun
    Pu, Yuehu
    [J]. He Jishu/Nuclear Techniques, 2023, 46 (11):
  • [3] Characterization of a Modified Clinical Linear Accelerator for Ultra-High Dose Rate Beam Delivery
    Deut, Umberto
    Camperi, Aurora
    Cavicchi, Cristiano
    Cirio, Roberto
    Data, Emanuele Maria
    Durisi, Elisabetta Alessandra
    Ferrero, Veronica
    Ferro, Arianna
    Giordanengo, Simona
    Villarreal, Oscar Marti
    Mas Milian, Felix
    Medina, Elisabetta
    Olivares, Diango M. Montalvan
    Mostardi, Franco
    Monti, Valeria
    Sacchi, Roberto
    Salmeri, Edoardo
    Vignati, Anna
    [J]. APPLIED SCIENCES-BASEL, 2024, 14 (17):
  • [4] Ultra-high dose rate FLASH irradiator at the radiological research accelerator facility
    Garty, Guy
    Obaid, Razib
    Deoli, Naresh
    Royba, Ekaterina
    Tan, Yuewen
    Harken, Andrew D.
    Brenner, David J.
    [J]. SCIENTIFIC REPORTS, 2022, 12 (01)
  • [5] Ultra-high dose rate FLASH irradiator at the radiological research accelerator facility
    Guy Garty
    Razib Obaid
    Naresh Deoli
    Ekaterina Royba
    Yuewen Tan
    Andrew D. Harken
    David J. Brenner
    [J]. Scientific Reports, 12
  • [6] Enabling ultra-high dose rate electron beams at a clinical linear accelerator for isocentric treatments
    Dal Bello, Riccardo
    von der Grun, Jens
    Fabiano, Silvia
    Rudolf, Thomas
    Saltybaeva, Natalia
    Stark, Luisa S.
    Ahmed, Md
    Bathula, Manohar
    Dogan, Serpil Kucuker
    McNeur, Joshua
    Guckenberger, Matthias
    Tanadini-Lang, Stephanie
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 187
  • [7] Understanding the FLASH effect to unravel the potential of ultra-high dose rate irradiation
    Kacem, Houda
    Almeida, Aymeric
    Cherbuin, Nicolas
    Vozenin, Marie-Catherine
    [J]. INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2022, 98 (03) : 506 - 516
  • [8] Neuroprotection of Radiosensitive Juvenile Mice by Ultra-High Dose Rate FLASH Irradiation
    Alaghband, Yasaman
    Cheeks, Samantha N.
    Allen, Barrett D.
    Montay-Gruel, Pierre
    Doan, Ngoc-Lien
    Petit, Benoit
    Jorge, Patrik Goncalves
    Giedzinski, Erich
    Acharya, Munjal M.
    Vozenin, Marie-Catherine
    Limoli, Charles L.
    [J]. CANCERS, 2020, 12 (06) : 1 - 21
  • [9] Feasibility of Clinically Practical Ultra-High Dose Rate (FLASH) Radiation Delivery by a Reversible Configuration of a Standard Clinical-Use Linear Accelerator
    No, H. J.
    Wu, Y. F.
    Manjappa, R.
    Skinner, L.
    Lau, B.
    Melemenidis, S.
    Yu, S. J.
    Surucu, M.
    Schueler, E.
    Bush, K.
    Graves, E. E.
    Maxim, P. G.
    Loo, B. W., Jr.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : S32 - S32
  • [10] Electron ultra-high dose rate FLASH irradiation study using a clinical linac: Linac modification, dosimetry, and radiobiological outcome
    Xie, De-Huan
    Li, Yi-Chuan
    Ma, Sai
    Yang, Xin
    Lan, Ruo-Ming
    Chen, Ao-Qiang
    Zhu, Hong-Yu
    Mei, Yan
    Peng, Li-Xia
    Li, Zuo-Feng
    Huang, Bi-Jun
    Chen, Yan
    Huang, Xiao-Yan
    Qian, Chao-Nan
    [J]. MEDICAL PHYSICS, 2022, 49 (10) : 6728 - 6738