Performance of quasi-exact cone-beam filtered backprojection algorithm for axially truncated helical data

被引:15
|
作者
Kudo, H [1 ]
Park, S
Noo, F
Defrise, M
机构
[1] Univ Tsukuba, Inst Informat Sci & Elect, Tsukuba, Ibaraki 3058573, Japan
[2] Univ Tsukuba, Masters Program Sci & Engn, Tsukuba, Ibaraki 3058573, Japan
[3] Univ Liege, Inst Elect Montefiore, B-4000 Cointe Ougree, Belgium
[4] Free Univ Brussels, AZ VUB, Div Nucl Med, B-1090 Brussels, Belgium
基金
日本学术振兴会;
关键词
D O I
10.1109/23.775587
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper concerns the problem of tomographic reconstruction from axially truncated cone-beam projections acquired with a helical vertex path. Very recently, Kudo et al. [1] developed a quasi-exact filtered backprojection algorithm for this problem. The purpose of this paper is to evaluate the performances of the quasi-exact algorithm by comparing it with variants of the approximate Feldkamp algorithm. The results show that the quasi-exact algorithm possesses the following two advantages over the Feldkamp algorithm. The first advantage is the improved image quality when the pitch of the helix is large. The second advantage is the use of a detector area close to the minimum detector area compatible with exact reconstruction. This area is shown to be much. smaller than that required by the Feldkamp algorithm for typical geometries. One disadvantage of the quasi-exact algorithm is the computational time which is much longer than for the Feldkamp algorithm.
引用
收藏
页码:608 / 617
页数:10
相关论文
共 50 条
  • [1] Quasi-exact filtered backprojection algorithm for long-object problem in helical cone-beam tomography
    Kudo, H
    Noo, F
    Defrise, M
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) : 902 - 921
  • [2] Cone-beam filtered-backprojection algorithm for truncated helical data
    Kudo, H
    Noo, F
    Defrise, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10): : 2885 - 2909
  • [3] New approximate filtered backprojection algorithm for helical cone-beam CT with redundant data
    Kudo, H
    Noo, F
    Defrise, M
    Rodet, T
    [J]. 2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 3211 - 3215
  • [4] Quasi-exact region-of-interest reconstruction from helical cone-beam data
    Defrise, M
    Noo, F
    Kudo, H
    [J]. 1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1101 - 1103
  • [5] A Filtered Backprojection Algorithm for Triple-Source Helical Cone-Beam CT
    Zhao, Jun
    Jin, Yannan
    Lu, Yang
    Wang, Ge
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (03) : 384 - 393
  • [6] Generalized quasi-exact algorithms for image reconstruction in helical cone-beam CT
    Zou, Y
    Sidky, EY
    Pan, XC
    [J]. MEDICAL IMAGING 2003: IMAGE PROCESSING, PTS 1-3, 2003, 5032 : 696 - 702
  • [7] Fast Exact/Quasi-Exact FBP Algorithms for Triple-Source Helical Cone-Beam CT
    Lu, Yang
    Katsevich, Alexander
    Zhao, Jun
    Yu, Hengyong
    Wang, Ge
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (03) : 756 - 770
  • [8] Regularized iterative weighted filtered backprojection for helical cone-beam CT
    Sunnegardh, Johan
    Danielsson, Per-Erik
    [J]. MEDICAL PHYSICS, 2008, 35 (09) : 4173 - 4185
  • [9] Quasi-exact helical cone beam reconstruction for micro CT
    Hu, JC
    Johnson, R
    Molthen, R
    Haworth, S
    Dawson, C
    [J]. 2002 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING, PROCEEDINGS, 2002, : 681 - 684
  • [10] A hierarchical algorithm for fast backprojection in helical cone-beam tomography
    Bresler, Y
    Brokish, J
    [J]. 2004 2ND IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1 AND 2, 2004, : 1420 - 1423