Automated on-line noninvasive optical glucose monitoring in a cell culture system

被引:17
|
作者
Jung, BJ
Lee, S
Yang, IH
Good, T
Coté, GL
机构
[1] Texas A&M Univ, Biomed Engn Program, College Stn, TX 77843 USA
[2] Texas A&M Univ, Dept Chem Engn, College Stn, TX 77843 USA
关键词
spectroscopy; on-line; glucose; fiber-optic sensors; partial least-squares; bioprocess monitoring; rotary cell culture system; RCCS;
D O I
10.1366/0003702021954421
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Current techniques for monitoring glucose concentration during cell culture are invasively performed using an off-line measurement system. Even though in situ or ex situ analyte sensors have been tried for on-line measurement, they suffer from difficulties including the inability to keep them sterile, their limited lifetime, and their lack of stability. In this research an alternative optical noninvasive on-line monitoring system based on near-infrared absorbance spectroscopy was developed to measure glucose concentration of cell culture media in a rotary cell culture system. The system included an automatic sampling component controlled by a LabVIEW program and a high-throughput fiber coupling component connected to a Fourier transform near-infrared spectrometer. The effectiveness of the system was investigated via noninvasive on-line measurement of glucose absorbance spectra in the cell culture media during T-cell culture. Glucose absorbance spectra were collected in the spectral range of 2.0 to 2.5 mum. Partial least-squares regression was employed to build a successful multivariate calibration model. The standard error of prediction and mean percent error for glucose were 7.7 mg/dL and 1.0%, respectively. The successful results represent an important step in the development of a noninvasive, closed-loop, system for monitoring analytes in cell culture.
引用
收藏
页码:51 / 57
页数:7
相关论文
共 50 条
  • [1] BIOT 229-Evaluation of an automated bioreactor sampling system for on-line monitoring a perfusion cell culture process
    Tengtrakool, Jennifer
    Johnson, Timothy
    Zhou, Weichang
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2009, 238
  • [2] On-line optical monitoring system for arc welding
    Sforza, P
    de Blasiis, D
    NDT & E INTERNATIONAL, 2002, 35 (01) : 37 - 43
  • [3] Optical biosensor optimized for continuous in-line glucose monitoring in animal cell culture
    Mircea Tric
    Mario Lederle
    Lisa Neuner
    Igor Dolgowjasow
    Philipp Wiedemann
    Stefan Wölfl
    Tobias Werner
    Analytical and Bioanalytical Chemistry, 2017, 409 : 5711 - 5721
  • [4] Optical biosensor optimized for continuous in-line glucose monitoring in animal cell culture
    Tric, Mircea
    Lederle, Mario
    Neuner, Lisa
    Dolgowjasow, Igor
    Wiedemann, Philipp
    Woelfl, Stefan
    Werner, Tobias
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2017, 409 (24) : 5711 - 5721
  • [5] SENSOR SYSTEM FOR ON-LINE MONITORING OF CELL CULTURES
    Moldovan, Carmen
    Iosub, Rodica
    Cornel, Radu
    Moore, Eric
    Paschero, Anna
    Messina, Walter
    Demarchi, Danilo
    Codreanu, Cecilia
    Necula, Daniel
    Codreanu, Nita
    Dinescu, Adrian
    Firtat, Bogdan
    CAS: 2009 INTERNATIONAL SEMICONDUCTOR CONFERENCE, VOLS 1 AND 2, PROCEEDINGS, 2009, : 263 - +
  • [6] An Automated On-line Monitoring and Fault Diagnosis System for Power Transformers
    Liang, Yishan
    Wang, Zhenyuan
    Liu, Yilu
    2006 IEEE/PES POWER SYSTEMS CONFERENCE AND EXPOSITION. VOLS 1-5, 2006, : 1105 - +
  • [7] Design of on-line monitoring system for optical transport network
    Xi'an Communication Institute of PLA, Xi'an 710065, China
    Guangdianzi Jiguang, 2007, 2 (213-215):
  • [8] In vivo monitoring of cerebral glucose with an updated on-line electroanalytical system
    Gu, Hui
    Xiong, Ping
    Tang, Huiling
    Chen, Shu
    Long, Yunfei
    Shi, Guoyue
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (23) : 5929 - 5935
  • [9] In vivo monitoring of cerebral glucose with an updated on-line electroanalytical system
    Hui Gu
    Ping Xiong
    Huiling Tang
    Shu Chen
    Yunfei Long
    Guoyue Shi
    Analytical and Bioanalytical Chemistry, 2019, 411 : 5929 - 5935
  • [10] On-line monitoring of glucose in mammalian cell culture using a flow injection analysis (FIA) mediated biosensor
    Male, KB
    Gartu, PO
    Kamen, AA
    Luong, JHT
    BIOTECHNOLOGY AND BIOENGINEERING, 1997, 55 (03) : 497 - 504