HTS Vector Magnet for Magnetic Circular Dichroism Measurement

被引:8
|
作者
Furuse, Mitsuho [1 ]
Okano, Makoto [1 ]
Fuchino, Shuichiro [1 ]
Uchida, Akira [2 ]
Fujihira, Junichi [2 ,3 ]
Fujihira, Seiichi
Kadono, Toshiharu [4 ]
Fujimori, Atsushi [4 ]
Koide, Tsuneharu [5 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki 3058568, Japan
[2] Fujihira Co Ltd, Tsuchiura, Ibaraki 3000843, Japan
[3] FEDLIC Co Ltd, Tsukuba, Ibaraki 3050031, Japan
[4] Univ Tokyo, Bunkyo Ku, Tokyo 1130033, Japan
[5] High Energy Accelerator Res Org, Inst Mat Struct Sci, Tsukuba, Ibaraki 3050801, Japan
基金
日本学术振兴会;
关键词
Conduction-cooled magnet; vector magnet; X-ray magnetic circular dichroism; YBCO coil;
D O I
10.1109/TASC.2012.2236599
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this study, we developed a two-dimensional vector magnet with Y-Ba-Cu-O (YBCO) tape conductors for use in an X-ray magnetic circular dichroism measurement system. The characteristics of this high-T-c superconducting vector magnet are as follows: (a) it is a split magnet comprising four unit coils that enable an X-ray magnetic circular dichroism system to access the center sample space along three axes; (b) it is compact with a high magnetic field, and it exhibits a cone-shaped design to enable the split unit coils to come close together and to achieve a maximum magnetic field up to 1 T at the center; and (c) it is conduction-cooled by a cryocooler and a cryogen-free system. We fabricated four unit coils with YBCO tape conductors (manufactured by SuperPower Inc.). Two pairs of the unit coils were configured at right angles to each other in a cryostat and cooled below 30 K by the cryocooler. We confirmed that the magnetic field orientation can be controlled by adjusting the current of the unit coils. However, we observed hysteresis and drift of the magnetic field owing to the screening current induced in the YBCO tape conductors. Here, we design a high-T-c superconducting vector magnet and present its excitation test results.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] A Commercial HTS Dipole Magnet for X-Ray Magnetic Circular Dichroism (XMCD) Experiments
    Huang, T.
    Gao, X.
    Pooke, D.
    Chamritski, V.
    Briggs, N.
    Christian, M.
    Gibson, S.
    Mitchell, J.
    Miles, M.
    de Feijter, J.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2012, 22 (03)
  • [2] MEASUREMENT OF MAGNETIC CIRCULAR DICHROISM
    GILBY, AC
    TATHAM, PER
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1969, 2 (11): : 1004 - &
  • [3] Magnet-Free Time-Resolved Magnetic Circular Dichroism with Pulsed Vector Beams
    Cao, Jiaan
    Ye, Lyuzhou
    He, Dawei
    Zheng, Xiao
    Mukamel, Shaul
    JOURNAL OF PHYSICAL CHEMISTRY LETTERS, 2022, 13 (48): : 11300 - 11306
  • [4] MEASUREMENT OF MAGNETIC CIRCULAR DICHROISM ON A MODIFIED SPECTROPOLARIMETER
    ABUSHUMAYS, A
    DUFFIELD, JJ
    APPLIED SPECTROSCOPY, 1970, 24 (01) : 67 - +
  • [5] MEASUREMENT OF MAGNETIC CIRCULAR DICHROISM USING ALTERNATING MAGNETIC FIELDS
    ABUSHUMAYS, A
    HOOPER, GE
    DUFFIELD, JJ
    APPLIED SPECTROSCOPY, 1971, 25 (02) : 238 - +
  • [6] SPECTROMETER FOR MEASUREMENT OF MAGNETIC AND NATURAL CIRCULAR-DICHROISM
    SUTHERLAND, JC
    VICKERY, LE
    KLEIN, MP
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1974, 45 (09): : 1089 - 1094
  • [7] PERMANENT-MAGNET INTERFACE FOR MAGNETIC CIRCULAR-DICHROISM SPECTROSCOPY
    NORDEN, B
    HAKANSSON, R
    DANIELSSON, S
    CHEMICA SCRIPTA, 1977, 11 (02): : 52 - 56
  • [8] Eight Tesla superconducting solenoid magnet for the magnetic circular dichroism study
    Chang, C. H.
    Hwang, C. S.
    Lin, F. Y.
    Huang, M. H.
    Chen, H. H.
    Jan, J. C.
    Yang, C. S.
    Lin, H.-J.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2007, 17 (02) : 2204 - 2207
  • [10] Magnetic measurement by electron magnetic circular dichroism in the transmission electron microscope
    Song Dongsheng
    Wang Ziqiang
    Zhu Jing
    ULTRAMICROSCOPY, 2019, 201 : 1 - 17