GRIN lens rod based probe for endoscopic spectral domain optical coherence tomography with fast dynamic focus tracking

被引:79
|
作者
Xie, TQ [1 ]
Guo, SG
Chen, ZP
Mukai, D
Brenner, M
机构
[1] Univ Calif Irvine, Dept Biomed Engn, Beckman Laser Inst, Irvine, CA 92617 USA
[2] Univ Calif Irvine, Beckman Laser Inst, Div Pulm & Crit Care, Med Ctr, Orange, CA 92868 USA
来源
OPTICS EXPRESS | 2006年 / 14卷 / 08期
关键词
D O I
10.1364/OE.14.003238
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In this manuscript, a GRIN (gradient index) lens rod based probe for endoscopic spectral domain optical coherence tomography (OCT) with dynamic focus tracking is presented. Current endoscopic OCT systems have a fixed focal plane or working distance. In contrast, the focus of this endoscopic OCT probe can dynamically be adjusted at a high speed (500 mm/s) without changing reference arm length to obtain high quality OCT images for contact or non-contact tissue applications, or for areas of difficult access for probes. The dynamic focusing range of the probe can be from 0 to 7.5 mm without moving the probe itself. The imaging depth is 2.8 mm and the lateral scanning range is up to 2.7 mm or 4.5 mm (determined by the diameter of different GRIN lens rods). Three dimensional imaging can be performed using this system over an area of tissue corresponding to the GRIN lens surface. The experimental results demonstrate that this GRIN lens rod based OCT system can perform a high quality non-contact in vivo imaging. This rigid OCT probe is solid and can be adapted to safely access internal organs, to perform front or side view imaging with an imaging speed of 8 frames per second, with all moving parts proximal to the GRIN lens, and has great potential for use in extremely compact OCT endoscopes for in vivo imaging in both biological research and clinical applications. (c) 2006 Optical Society of America.
引用
收藏
页码:3238 / 3246
页数:9
相关论文
共 50 条
  • [1] In vitro 3D articular cartilage imaging using endoscopic spectral domain optical coherence tomography (SD-OCT) with a GRIN (gradient index) lens rod based probe
    Xie, Tuqiang
    Ahn, Yehchan
    Chen, Zhongping
    Peavy, George
    COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE XII, 2008, 6847
  • [2] A Miniature Endoscopic Optical Coherence Tomography Probe Based on C-Lens
    Luo, Site
    Zhou, Liang
    Wang, Donglin
    Duan, Can
    Liu, Hao
    Zhu, Yu
    Li, Guoxin
    Zhao, Hui
    Tang, Jianyu
    Wu, Yuqing
    An, Xin
    Li, Xinling
    Liu, Yabing
    Xie, Huikai
    Huo, Li
    IEEE PHOTONICS JOURNAL, 2018, 10 (05):
  • [3] Endoscopic optical coherence tomography with a focus-adjustable probe
    Liao, Wenchao
    Chen, Tianyuan
    Wang, Chengming
    Zhang, Wenxin
    Peng, Zhangkai
    Zhang, Xiao
    Ai, Shengnan
    Fu, Deyong
    Zhou, Tieying
    Xue, Ping
    OPTICS LETTERS, 2017, 42 (20) : 4040 - 4043
  • [4] Spectral-domain optical coherence tomography for endoscopic imaging
    Chen, Xiaodong
    Li, Qiao
    Li, Wanhui
    Wang, Yi
    Yu, Daoyin
    COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE XI, 2007, 6429
  • [5] Dynamic focus catheter design for endoscopic optical coherence tomography
    Meemon, Panomsak
    Murali, Supraja
    Lee, Kye-Sung
    Rolland, Jannick
    2007 IEEE LEOS ANNUAL MEETING CONFERENCE PROCEEDINGS, VOLS 1 AND 2, 2007, : 9 - 10
  • [6] Spectral-Domain Endoscopic Optical Coherence Tomography (EOCT) Using a Balloon Probe for Esophageal Imaging
    Isenberg, Gerard
    Kang, Wei
    Wang, Hui
    Pan, Yinsheng
    Jenkins, Michael
    Chak, Amitabh
    Agrawal, Deepak
    Rollins, Andrew
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2009, 104 : S29 - S30
  • [7] Optical fiber probe for optical coherence tomography with extended depth of field using a modified GRIN fiber lens
    Sun, Xiaoguang
    Bedard, Kyle
    Li, Jie
    OPTICAL FIBERS AND SENSORS FOR MEDICAL DIAGNOSTICS, TREATMENT AND ENVIRONMENTAL APPLICATIONS XXII, 2022, 11953
  • [8] The effect of contact lens wear on retinal spectral domain optical coherence tomography
    Aviram, Talia
    Beeri, Idan
    Berkow, David
    Zayit-Soudry, Shiri
    Blumenthal, Eytan Z.
    Shapira, Yinon
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2020, 103 (06) : 792 - 797
  • [9] Gradient-index lens rod based probe for office-based optical coherence tomography of the human larynx
    Guo, Shuguang
    Yu, Lingfeng
    Sepehr, Ali
    Perez, Jorge
    Su, Jianping
    Ridgway, James M.
    Vokes, David
    Wong, Brian J. F.
    Chen, Zhongping
    JOURNAL OF BIOMEDICAL OPTICS, 2009, 14 (01)
  • [10] Extended depth of focus adaptive optics spectral domain optical coherence tomography
    Sasaki, Kazuhiro
    Kurokawa, Kazuhiro
    Makita, Shuichi
    Yasuno, Yoshiaki
    BIOMEDICAL OPTICS EXPRESS, 2012, 3 (10): : 2353 - 2370