Characterization of a constant current charge detector

被引:5
|
作者
Mori, Masanobu [1 ]
Chen, Yongjing [1 ]
Ohira, Shin-Ichi [1 ]
Dasgupta, Purnendu K. [1 ]
机构
[1] Univ Texas Arlington, Dept Chem & Biochem, Arlington, TX 76019 USA
关键词
Constant current charge detector; Electric field induced dissociation; Membrane electro-transport; ION-EXCHANGE MEMBRANES; ELECTRODIALYTIC ELUENT GENERATOR; WATER DISSOCIATION; BIPOLAR MEMBRANES; PROTON-TRANSFER; OHMS LAW; BASE; ACID; CHROMATOGRAPHY; RECTIFICATION;
D O I
10.1016/j.talanta.2012.07.058
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ion exchangers are ionic equivalents of doped semiconductors, where cations and anions are equivalents of holes and electrons as charge carriers in solid state semiconductors. We have previously demonstrated an ion exchange membrane (IEM) based electrolyte generator which behaves similar to a light-emitting diode and a charge detector (ChD) which behaves analogous to a p-i-n photodiode. The previous work on the charge detector, operated at a constant voltage, established its unique ability to respond to the charge represented by the analyte ions regardless of their redox properties, rather than to their conductivities. It also suggested that electric field induced dissociation (EFID) of water occurs at one or both ion exchange membranes. A logical extension is to study the behavior of the same device, operated in a constant current mode (ChD(i)). The evidence indicates that in the present operational mode the device also responds to the charge represented by the analytes and not their conductivity. Injection of a base into a charge detector operated in the constant voltage mode was not previously examined; in the constant current mode, base injection appears to inhibit EFID. The effects of applied current, analyte residence time and outer channel fluid composition were individually examined; analyte ions of different mobilities as well as affinities for the respective IEMs were used. While the exact behavior is somewhat dependent on the applied current, strong electrolytes, both acids and salts, respond the highest and in a near-uniform fashion, weak acids and their salts respond in an intermediate fashion and bases produce the lowest responses. A fundamentally asymmetric behavior is observed. Injected bases but not injected acids produce a poor response; the effects of incorporating a strong base as the electrolyte in the anion exchange membrane (AEM) compartment is far greater than incorporating an acid in the cation exchange membrane (CEM) compartment. These observations suggest that EFID occurs primarily at the anion exchange membrane and is inhibited by the presence of OH-. Analyte peak widths and peak asymmetries were found to be governed by the mobility of the analyte ions and their affinities for the IEMs, respectively. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:44 / 52
页数:9
相关论文
共 50 条
  • [1] Transient-Current Technique Characterization of Diamond Detector and Proposal of a Diamond Detector with Charge Amplification
    Kholili, Muhammad Jauhar
    Shimaoka, Takehiro
    Hara, Asuka
    Tanaka, Manobu M.
    PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 2024, 221 (04):
  • [2] RELATIVE PERFORMANCE OF A PHOTOCONDUCTIVE DETECTOR IN CONSTANT CURRENT AND CONSTANT VOLTAGE BIAS MODES
    GOPAL, V
    INFRARED PHYSICS, 1981, 21 (01): : 31 - 36
  • [3] NEGATIVE CHARGE TRANSPORT IN FLUORETHYLENEPROPYLENE BY THE CONSTANT CURRENT METHOD
    GIACOMETTI, JA
    FERREIRA, GFL
    GROSS, B
    PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 1985, 88 (01): : 297 - 307
  • [4] CIRCUIT FOR CONSTANT CURRENT OPERATION OF AN ELECTRON-CAPTURE DETECTOR
    KNIGHTON, WB
    GRIMSRUD, EP
    ANALYTICAL CHEMISTRY, 1982, 54 (11) : 1892 - 1893
  • [5] CONSTANT-CURRENT AC ELECTRON-CAPTURE DETECTOR
    SINGH, H
    MILLIER, B
    AUE, WA
    JOURNAL OF CHROMATOGRAPHY A, 1995, 690 (01) : 103 - 108
  • [6] DIRECT CONSTANT-CURRENT ELECTRON-CAPTURE DETECTOR
    AUE, WA
    SIU, KWM
    ANALYTICAL CHEMISTRY, 1980, 52 (09) : 1544 - 1546
  • [7] Charge Collection Characterization with Semiconductor Pixel Detector Timepix
    Kroupa, M.
    Jakubek, J.
    Krejci, F.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 5066 - 5069
  • [8] The GOTTHARD charge integrating readout detector: design and characterization
    Mozzanica, A.
    Bergamaschi, A.
    Dinapoli, R.
    Graafsma, H.
    Greiffenberg, D.
    Henrich, B.
    Johnson, I.
    Lohmann, M.
    Valeria, R.
    Schmitt, B.
    Xintian, S.
    JOURNAL OF INSTRUMENTATION, 2012, 7
  • [9] A high-accuracy BiCMOS constant current charge circuit
    Chen, FJ
    Lai, XQ
    Li, YS
    Zhao, JH
    2005 6th International Conference on ASIC Proceedings, Books 1 and 2, 2005, : 560 - 563
  • [10] Equivalent charge current step model with constant break point
    Fan, Xiaoming
    Zhang, Li
    Wang, Guodong
    Han, Anrong
    Xu, Jianghua
    Dongbei Daxue Xuebao/Journal of Northeastern University, 1999, 20 (06): : 641 - 643