Brief report: validation of a system for automated measurement of knee laxity

被引:5
|
作者
Thompson, MT
Conditt, MA
Ismaily, SK
Agarwal, A
Noble, PC
机构
[1] Inst Orthoped Res & Educ, Houston, TX 77030 USA
[2] Baylor Coll Med, Houston, TX 77030 USA
关键词
knee; laxity; fluoroscope; AP drawer;
D O I
10.1016/j.clinbiomech.2003.12.004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. To determine the accuracy and repeatability of an automated quantitative fluoroscopic imaging system for measuring knee laxity. Design. Cadaveric validation study. Background. Current methods of measuring anterior-posterior laxity lack sufficient accuracy and repeatability. A commercially developed fluoroscopic software package, capable of measuring laxity, required validation. Methods. Five human cadaveric knees were used. A constant force of 130 N was applied anteriorly and posteriorly in turn to the tibia of each knee with the femur fixed in 30degrees and 90degrees of flexion. Quantitative fluoroscopic measurements of anterior-posterior laxity were determined using image analysis software. Fluoroscopic results were compared to the true anterior-posterior displacements of the tibia, which were simultaneously recorded using linear transducers directly attached to the cadaveric specimens. Results. The quantitative fluoroscopic method underestimated laxity by an average of 0.40 mm with a root mean square error of 0.49 mm. The 95% confidence intervals for anterior and posterior laxity error were calculated to be -0.99 to 0.25 mm and -0.89 to 0.03 mm, respectively, where a negative error represents an underestimation. Conclusions. The quantitative fluoroscopic method offers a dramatic improvement in accuracy over current laxity measurement techniques and acceptable repeatability for assessing ligament damage.
引用
收藏
页码:308 / 312
页数:5
相关论文
共 50 条
  • [1] MEASUREMENT OF KNEE LIGAMENT LAXITY
    HARDING, ML
    STEINGOLD, RF
    HOWARD, L
    AITKEN, W
    STEPHEN, RO
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01): : 159 - 159
  • [2] Dynamic knee laxity measurement devices
    Ahlden, Mattias
    Hoshino, Yuichi
    Samuelsson, Kristian
    Araujo, Paulo
    Musahl, Volker
    Karlsson, Jon
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (04) : 621 - 632
  • [3] Objective measurement of sagittal laxity of the knee
    Eriksson, E
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 1999, 7 (04) : 203 - 203
  • [4] Objective measurement of sagittal laxity of the knee
    Ejnar Eriksson
    Knee Surgery, Sports Traumatology, Arthroscopy, 1999, 7 : 203 - 203
  • [5] INSTRUMENTED MEASUREMENT OF ANTERIOR LAXITY OF THE KNEE
    DANIEL, DM
    MALCOM, LL
    LOSSE, G
    STONE, ML
    SACHS, R
    BURKS, R
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (05): : 720 - 726
  • [6] Dynamic knee laxity measurement devices
    Mattias Ahldén
    Yuichi Hoshino
    Kristian Samuelsson
    Paulo Araujo
    Volker Musahl
    Jón Karlsson
    Knee Surgery, Sports Traumatology, Arthroscopy, 2012, 20 : 621 - 632
  • [7] Validation of Quantitative Measures of Rotatory Knee Laxity
    Musahl, Volker
    Griffith, Chad
    Irrgang, James J.
    Hoshino, Yuichi
    Kuroda, Ryosuke
    Lopomo, Nicola
    Zaffagnini, Stefano
    Samuelsson, Kristian
    Karlsson, Jon
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2016, 44 (09): : 2393 - 2398
  • [8] Description and validation of a navigation system for intra-operative evaluation of knee laxity
    Martelli, Sandra
    Zaffagnini, Stefano
    Bignozzi, Simone
    Lopomo, Nicola
    Marcacci, Maurilio
    COMPUTER AIDED SURGERY, 2007, 12 (03) : 181 - 188
  • [9] EVALUATION OF AN ELECTROGONIOMETRIC INSTRUMENT FOR MEASUREMENT OF LAXITY OF THE KNEE
    GRANBERRY, WM
    NOBLE, PC
    WOODS, GW
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (09): : 1316 - 1322
  • [10] Current state of instrumental measurement of knee laxity
    Mayr, H. O.
    Stoehr, A.
    ARTHROSKOPIE, 2010, 23 (01) : 56 - 61