A haptic unit designed for magnetic-resonance-guided biopsy

被引:10
|
作者
Tse, Z. T. H. [1 ]
Elhawary, H. [1 ]
Rea, M. [1 ]
Young, I. [1 ]
Davis, B. L. [1 ]
Lamperth, M. [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, Mechatron Med Lab, Dept Mech Engn, London SW7 2AZ, England
关键词
magnetic-resonance-imaging-compatible image-guided interventions; haptic unit; master-slave; medical robotics; MRI; NEEDLE; SYSTEM; COMPATIBILITY; ELASTOGRAPHY; ROBOT;
D O I
10.1243/09544119JEIM468
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The magnetic fields present in the magnetic resonance (MR) environment impose severe constraints on any mechatronic device present in its midst, requiring alternative actuators, sensors, and materials to those conventionally used in traditional system engineering. In addition the spatial constraints of closed-bore scanners require a physical separation between the radiologist and the imaged region of the patient. This configuration produces a loss of the sense of touch from the target anatomy for the clinician, which often provides useful information. To recover the force feedback from the tissue, an MR-compatible haptic unit, designed to be integrated with a five-degrees-of-freedom mechatronic system for MR-guided prostate biopsy, has been developed which incorporates position control and force feedback to the operator. The haptic unit is designed to be located inside the scanner isocentre with the master console in the control room. MR compatibility of the device has been demonstrated, showing a negligible degradation of the signal-to-noise ratio and virtually no geometric distortion. By combining information from the position encoder and force sensor, tissue stiffness measurement along the needle trajectory is demonstrated in a lamb liver to aid diagnosis of suspected cancerous tissue.
引用
收藏
页码:159 / 172
页数:14
相关论文
共 50 条
  • [1] Magnetic-resonance-guided biopsy of focal liver lesions
    Smith, Ethan A.
    Grove, Jason J.
    Van der Spek, Abraham F. L.
    Jarboe, Marcus D.
    PEDIATRIC RADIOLOGY, 2017, 47 (06) : 750 - 754
  • [2] Magnetic-resonance-guided biopsy of focal liver lesions
    Ethan A. Smith
    Jason J. Grove
    Abraham F. L. Van Der Spek
    Marcus D. Jarboe
    Pediatric Radiology, 2017, 47 : 750 - 754
  • [3] Magnetic-resonance-guided percutaneous cryoablation of hepatic tumours
    Mala, T
    Edwin, B
    Samset, E
    Gladhaug, I
    Hol, PK
    Fosse, E
    Mathisen, O
    Bergan, A
    Soreide, O
    EUROPEAN JOURNAL OF SURGERY, 2001, 167 (08) : 610 - 617
  • [4] Design of a System for Magnetic-Resonance-Guided Irreversible Electroporation
    Hubmann, Joris
    Gerlach, Thomas
    Pannicke, Enrico
    Hensen, Bennet
    Wacker, Frank
    Speck, Oliver
    Vick, Ralf
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 1457 - 1461
  • [5] A Testbed for Haptic and Magnetic Resonance Imaging-Guided Percutaneous Needle Biopsy
    Mendoza, Evelyn
    Whitney, John Peter
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2019, 4 (04) : 3177 - 3183
  • [6] Magnetic-resonance-guided percutaneous laser ablation of uterine fibroids
    Law, P
    Gedroyc, WMW
    Regan, L
    LANCET, 1999, 354 (9195): : 2049 - 2050
  • [7] An evaluation of the accuracy of magnetic-resonance-guided gamma knife surgery
    Guo, WY
    Chu, WC
    Wu, MC
    Chung, WY
    Gwan, WP
    Lee, YL
    Pan, HC
    Chang, CY
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1996, 66 : 85 - 92
  • [8] Magnetic-resonance-guided radiotherapy The beginning of a new era in radiation oncology?
    Hoegen, P.
    Spindeldreier, C. K.
    Buchele, C.
    Rippke, C.
    Regnery, S.
    Weykamp, F.
    Klueter, S.
    Debus, J.
    Hoerner-Rieber, J.
    RADIOLOGE, 2021, 61 (01): : 13 - 20
  • [9] Application of diffusion tensor imaging to magnetic-resonance-guided brain tumor resection
    Tummala, RP
    Chu, RM
    Liu, HY
    Truwit, CL
    Hall, WA
    PEDIATRIC NEUROSURGERY, 2003, 39 (01) : 39 - 43
  • [10] Interventional magnetic-resonance-guided cryotherapy combined with microsurgery for recurrent glioblastoma: An innovative treatment?
    Cebula, H.
    Garnon, J.
    Todeschi, J.
    Noel, G.
    Lhermitte, B.
    Mallereau, C-H
    Chibbaro, S.
    Burckel, H.
    Schott, R.
    de Mathelin, M.
    Gangi, A.
    Proust, F.
    NEUROCHIRURGIE, 2022, 68 (03) : 267 - 272