Performance evaluation of a portable 3D vision coordinate measuring system

被引:1
|
作者
Icasio-Hernandez, Octavio [1 ,2 ]
Gonzalez-Barbosa, Jose-Joel [2 ]
Curiel-Razo, Yajaira-Ilse [2 ]
Hurtado-Ramos, Juan B. [2 ]
机构
[1] CENAM, Ctr Nacl Metrol, El Marques, Mexico
[2] CICATA Queretaro, Inst Politecn Nacl, Col Colinas Cimatario, Santiago De Queretaro, Mexico
关键词
Photogrammetry; LUT; portable CMM' sphere test; single point test; volumetric test; CLOSE-RANGE PHOTOGRAMMETRY; CAMERA CALIBRATION; SENSITIVITY-ANALYSIS; PNP PROBLEM; POSE;
D O I
10.1080/00051144.2017.1391616
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this work, we present a portable 3D vision coordinate measuring machine (PCMM) for short range-real time photogrammetry. The PCMM performs 3D measurements of points using a single camera in combination with a hand tool and a computer. The hand tool has infrared LEDs serving as photogrammetric targets. The positions of these targets were pre-calibrated with an optical coordinate-measuring machine defining a local coordinate system on the hand tool. The camera has an infrared filter to exclude all ambient light but infrared targets. Positions of the imaged infrared targets are converted to 3D coordinates using pixel positions and precalibrated positions of the targets. Also, we present a set of criteria for selecting the infrared LEDs and the camera filter, a camera calibration method, a tracking and POSE algorithms, and a 3D coordinate error correction for the PCMM. The correction is performed using the PCMM as a range meter, which implies comparing the 3D coordinate points of the PCMM with a coordinate measuring machine, and then generating a look up table (LUT) for correction. The global error of the PCMM was evaluated under ASME B89.4.22-2004. Sphere and single point errors were around 1 mm, volumetric error were under 3 mm.
引用
收藏
页码:253 / 265
页数:13
相关论文
共 50 条
  • [1] The modeling of portable 3D vision coordinate measuring system
    Liu, SG
    Huang, FS
    Peng, K
    Optical Design and Testing II, Pts 1 and 2, 2005, 5638 : 835 - 842
  • [2] Portable Light Pen 3D Vision Coordinate Measuring System-Probe Tip Center Calibration
    Liu, Shugui
    Zhang, Hongling
    Dong, Yinghua
    Tang, Shaliang
    Jiang, Zhenzhu
    MEASUREMENT SCIENCE REVIEW, 2013, 13 (04): : 194 - 199
  • [3] A portable 3D vision coordinate measurement system using a light pen
    Liu, S.G.
    Peng, K.
    Huang, F.S.
    Zhang, G.X.
    Li, P.
    Key Engineering Materials, 2005, 295-296 : 331 - 336
  • [4] A portable 3D vision coordinate measurement system using a light pen
    Liu, SG
    Peng, K
    Huang, FS
    Zhang, GX
    Li, P
    MEASUREMENT TECHNOLOGY AND INTELLIGENT INSTRUMENTS VI, 2005, 295-296 : 331 - 336
  • [5] Performance evaluation of a portable 3D imaging system
    Givi, Mehrdad
    Cournoyer, Luc
    Reain, Greg
    Eves, Brian J.
    PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2019, 59 : 156 - 165
  • [6] Performance evaluation for vision coordinate measuring machines
    Huang, CC
    Zhang, XD
    PROCEEDINGS OF THE FOURTEENTH ANNUAL MEETING OF THE AMERICAN SOCIETY FOR PRECISION ENGINEERING, 1999, : 529 - 532
  • [7] Image Processing in Vision 3D Coordinate Measurement System
    Zhou Xing-lin
    Peng Kai
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, VOLS 1-9, 2009, : 3350 - 3354
  • [8] A CAD CAM SYSTEM FOR 3D COORDINATE MEASURING DATA
    LEE, AC
    CHEN, DP
    LIN, CL
    INTERNATIONAL JOURNAL OF PRODUCTION RESEARCH, 1990, 28 (12) : 2353 - 2371
  • [9] Evaluation the Performance of Portable Coordinate-Measuring Systems with MSA Methodology
    Dukendjiev, Georgi
    Diakov, Dimitar
    Vassilev, Velizar
    Nikolova, Hristiana
    Nikolov, Rumen
    2022 XXXII INTERNATIONAL SCIENTIFIC SYMPOSIUM METROLOGY AND METROLOGY ASSURANCE (MMA), 2022, : 185 - 188
  • [10] Single camera 3D coordinate on-line measuring system
    Yu, ZJ
    Chen, G
    Liu, CY
    Che, RS
    Ye, D
    Wang, QZ
    Tian, W
    PROCEEDINGS OF THE SECOND INTERNATIONAL SYMPOSIUM ON INSTRUMENTATION SCIENCE AND TECHNOLOGY, VOL 2, 2002, : 307 - 313