Quasi-Static Calibration Method of a High-g Accelerometer

被引:10
|
作者
Wang, Yan [1 ,2 ]
Fan, Jinbiao [2 ]
Zu, Jing [1 ]
Xu, Peng [1 ]
机构
[1] North Univ China, Sci & Technol Elect Test & Measurement Lab, Taiyuan 030051, Peoples R China
[2] North Univ China, Sch Comp Sci & Control Engn, Taiyuan 030051, Peoples R China
基金
中国国家自然科学基金;
关键词
quasi-static calibration; high-g accelerometer; pulse width; gas gun; laser interferometer; PERFORMANCE EVALUATION; PENETRATION;
D O I
10.3390/s17020409
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
To solve the problem of resonance during quasi-static calibration of high-g accelerometers, we deduce the relationship between the minimum excitation pulse width and the resonant frequency of the calibrated accelerometer according to the second-order mathematical model of the accelerometer, and improve the quasi-static calibration theory. We establish a quasi-static calibration testing system, which uses a gas gun to generate high-g acceleration signals, and apply a laser interferometer to reproduce the impact acceleration. These signals are used to drive the calibrated accelerometer. By comparing the excitation acceleration signal and the output responses of the calibrated accelerometer to the excitation signals, the impact sensitivity of the calibrated accelerometer is obtained. As indicated by the calibration test results, this calibration system produces excitation acceleration signals with a pulse width of less than 1000 mu s, and realize the quasi-static calibration of high-g accelerometers with a resonant frequency above 20 kHz when the calibration error was 3%.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] High-G Calibration Denoising Method for High-G MEMS Accelerometer Based on EMD and Wavelet Threshold
    Lu, Qing
    Pang, Lixin
    Huang, Haoqian
    Shen, Chong
    Cao, Huiliang
    Shi, Yunbo
    Liu, Jun
    MICROMACHINES, 2019, 10 (02)
  • [2] Shock Calibration of the High-g Triaxial Accelerometer
    Wang, Yan
    Fan, Jinbiao
    Xu, Peng
    Zu, Jing
    Zhang, Zhijie
    2008 IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-5, 2008, : 741 - 745
  • [3] Design, fabrication and calibration of a high-G MEMS accelerometer
    Shi, Yunbo
    Zhao, Yongqi
    Feng, Hengzhen
    Cao, Huiliang
    Tang, Jun
    Li, Jie
    Zhao, Rui
    Liu, Jun
    SENSORS AND ACTUATORS A-PHYSICAL, 2018, 279 : 733 - 742
  • [4] High-G accelerometer dynamic calibration by a laser differential Doppler technique
    Wang, SY
    Cao, CZ
    Fei, PD
    Zhang, AH
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (05): : 2022 - 2025
  • [5] Static and quasi-static calibration of a bioMEMS device
    Slifka, AJ
    Panchawagh, H
    Mahajan, RL
    Finch, D
    Rorrer, RAL
    BIOMEDICAL SCIENCES INSTRUMENTATION, VOL 40, 2004, 449 : 429 - 434
  • [6] Fabrication of high-G piezoelectric accelerometer
    Wu, ZT
    Shao, XZ
    Ge, L
    Li, P
    Liu, GL
    Jiang, XD
    Tang, RD
    ISTM/2005: 6th International Symposium on Test and Measurement, Vols 1-9, Conference Proceedings, 2005, : 4601 - 4604
  • [7] Sensitivity Matrix Calibration of Triaxial High-G Accelerometer Based on Triaxial Shock Calibration Device
    Zhang, Wenyi
    Teng, Fei
    Zhang, Zhenhai
    IEEE SENSORS JOURNAL, 2024, 24 (07) : 10975 - 10982
  • [8] Research on a novel calibration method for high-g accelerometers
    Cao, Xuan
    Chen, Yufeng
    Xu, Kaisi
    Su, Weiguo
    Zhang, Wei
    2017 IEEE 12TH INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS (NEMS), 2017, : 591 - 594
  • [9] Measurement of exciting pulse for calibration of high-g accelerometer using lateral laser interferometer
    Fan, Jinbiao
    Li, Xi
    Xu, Peng
    Zu, Jing
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2015, 44 (02): : 497 - 502
  • [10] High-G MEMS Accelerometer Calibration Denoising Method Based on EMD and Time-Frequency Peak Filtering
    Wang, Chenguang
    Cui, Yuchen
    Liu, Yang
    Li, Ke
    Shen, Chong
    MICROMACHINES, 2023, 14 (05)