Variability in Delivered Dose from Pressurized Metered-Dose Inhaler Formulations Due to a Delay Between Shake and Fire

被引:28
|
作者
Hatley, Ross H. M. [1 ]
Parker, Jacob [1 ]
Pritchard, John N. [1 ]
von Hollen, Dirk [2 ]
机构
[1] Philips Respiron Resp Drug Delivery UK Ltd, Tangmere, Chichester, England
[2] Respironics Inc, Murrysville, PA USA
关键词
creaming; delivered drug; formulation; pressurized metered-dose inhaler (pMDI); sedimentation; shake-fire delay; IN-VITRO; INHALATION DEVICES; CORTICOSTEROIDS; PERFORMANCE; ASTHMA;
D O I
10.1089/jamp.2015.1284
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Pressurized metered-dose inhalers (pMDIs) should be shaken before use to prevent creaming or sedimentation of the drugs in solution; however, data published on this topic are limited, and it is rarely specified how soon after shaking the device should be actuated. Delays between shaking and firing the pMDI have previously been shown to cause significant inhomogeneity in delivered dose. We studied the effect of various shake-fire delays on the drug delivered from five commercially available pMDIs commonly prescribed for asthma and chronic obstructive pulmonary disease to assess the potential variability in delivered dose. Methods: The pMDI formulations tested were the Flovent HFA, Ventolin Evohaler, Airomir Inhaler, and Symbicort (suspension pMDIs), and the QVAR 100 Inhaler (solution pMDI). Each pMDI was shaken for 5 seconds before attachment to a dosage unit sampling apparatus collection tube and filter, and it was actuated once with shake-fire delays of 0, 5, 10, 20, 30, 40, 50, and 60 seconds. Analysis of the eluates from the collection tubes and filters was performed by using high-performance liquid chromatography. Three of each pMDI were tested twice with each time delay. Results: All of the suspension pMDIs produced variable amounts of drug over the shake-fire delays tested. A comparison of the delivered doses after the 0- and 60-second delays showed that the drug delivered increased for the Flovent HFA (320%), Ventolin Evohaler (346%), and Airomir Inhaler (230%) pMDIs; decreased for the Symbicort budesonide (75%) and formoterol fumarate (76%) pMDI; and remained consistent for the QVAR 100 Inhaler pMDI. Conclusions: The amount of drug delivered can vary widely over different shake-fire delays with suspension pMDIs. Therefore, guidance should be given to users/caregivers on the timing of firing after shaking their device, particularly with pediatrics, who may take time to become receptive to accepting their medication after pMDI shaking and before dose administration.
引用
收藏
页码:71 / 79
页数:9
相关论文
共 50 条
  • [1] Effects of Shake-Actuation Delay on Delivered Dose and Aerosol Characteristics of an Albuterol Suspension From a Pressurized Metered-Dose Inhaler
    Qaqish, Rasha
    Berlinski, Ariel
    RESPIRATORY CARE, 2021, 66 (08) : 1315 - 1322
  • [2] Delay Between Shaking and Actuation of a Hydrofluoroalkane Fluticasone Pressurized Metered-Dose Inhaler
    Berlinski, Ariel
    von Hollen, Dirk
    Pritchard, John N.
    Hatley, Ross H. M.
    RESPIRATORY CARE, 2018, 63 (03) : 289 - 293
  • [3] Budesonide/Formoterol Pressurized Metered-Dose Inhaler
    Katherine A. Lyseng-Williamson
    Dene Simpson
    Drugs, 2008, 68 : 1855 - 1864
  • [4] Budesonide/formoterol pressurized metered-dose inhaler
    Lyseng-Willianison, Katherine A.
    Simpson, Dene
    DRUGS, 2008, 68 (13) : 1855 - 1864
  • [5] Characterization of the Spray Velocities from a Pressurized Metered-Dose Inhaler
    Crosland, Brian Michael
    Johnson, Matthew Ronald
    Matida, Edgar Akio
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2009, 22 (02) : 85 - 97
  • [6] Theoretical investigation of the spray from a pressurized metered-dose inhaler
    Dunbar, CA
    Watkins, AP
    Miller, JF
    ATOMIZATION AND SPRAYS, 1997, 7 (04) : 417 - 436
  • [7] An investigation into the use of clathrates in metered-dose inhaler formulations
    Bouhoum, A.
    Champness, N.
    James, J.
    Roberts, C.
    Toon, R.
    JOURNAL OF PHARMACY AND PHARMACOLOGY, 2008, 60 : A21 - A21
  • [8] The bronchoprotective efficacy of salbutamol inhaled from a new metered-dose powder inhaler compared with a conventional pressurized metered-dose inhaler connected to a spacer
    Seppala, OP
    Herrala, J
    Hedman, J
    Alanko, K
    Liipo, K
    Terho, E
    Pietinalho, A
    Nyholm, JE
    Nieminen, MM
    RESPIRATORY MEDICINE, 1998, 92 (03) : 578 - 583
  • [9] FACTORS AFFECTING TOTAL AND RESPIRABLE DOSE DELIVERED BY A SALBUTAMOL METERED-DOSE INHALER
    EVERARD, ML
    DEVADASON, SG
    SUMMERS, QA
    LESOUEF, PN
    THORAX, 1995, 50 (07) : 746 - 749
  • [10] Study of a Pressurized Metered-dose Inhaler Spray Parameters in Fluent™
    Oliveira, Ricardo F.
    Teixeira, Senhorinha
    Silva, Luis F.
    Teixeira, Jose Carlos
    Antunes, Henedina
    WORLD CONGRESS ON ENGINEERING, WCE 2010, VOL II, 2010, : 1083 - 1087