Using overlap volume histogram and IMRT plan data to guide and automate VMAT planning: A head-and-neck case study

被引:78
|
作者
Wu, Binbin [1 ]
Pang, Dalong [1 ]
Simari, Patricio [2 ]
Taylor, Russell [3 ]
Sanguineti, Giuseppe [4 ]
McNutt, Todd [4 ]
机构
[1] Georgetown Univ Hosp, Dept Radiat Med, Washington, DC 20007 USA
[2] Univ Maryland, Dept Comp Sci, College Pk, MD 20740 USA
[3] Johns Hopkins Univ, Whiting Sch Engn, Dept Comp Sci, Baltimore, MD 21218 USA
[4] Johns Hopkins Univ, Dept Radiat Oncol & Mol Radiat Sci, Baltimore, MD 21231 USA
关键词
VMAT; OVH; IMRT database; DVH; head-and-neck; MODULATED ARC THERAPY; RADIATION-THERAPY; CANCER; QUALITY; RADIOTHERAPY; IMPACT;
D O I
10.1118/1.4788671
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate whether an overlap volume histogram (OVH)-driven planning application using an intensity-modulated radiation therapy (IMRT) database can guide and automate volumetric-modulated arc therapy (VMAT) planning for head-and-neck cancer. Methods: Based on comparable head-and-neck dosimetric results between planner-generated VMAT and IMRT plans, an inhouse developed, OVH-driven automated planning application containing a database of prior clinical head-and-neck IMRT plans is built into Pinnacle(3) SmartArc for VMAT planning. Double-arc VMAT plans of four oropharynx, four nasopharynx, and four larynx patients are generated and compared with corresponding clinical IMRT plans. Results: Each VMAT plan is automatically generated in two optimization rounds, while the average number of optimization rounds in generating a clinical IMRT plan is 43. In VMAT plans, statistical superiority (p < 0.01) in sparing of the cord+4 mm, brainstem, brachial plexus, larynx, and inner ear is observed with a slight degradation in low-dose-level planning target volume (PTV) coverage. On average, D-0.1 cc to the cord+4 mm, brainstem and brachial plexus is reduced by 3.7, 4.9, and 1.6 Gy, respectively; V(50 Gy) to the larynx is reduced by 5.3%; mean dose to the inner ear is reduced by 4.4 Gy; V-95 of low-dose-level PTV coverage is reduced by 0.3% with p = 0.25. Conclusions: IMRT-data-driven VMAT planning offers a potential method for generating VMAT plans that are comparable to IMRT plans in terms of dosimetric quality. (c) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4788671]
引用
收藏
页数:7
相关论文
共 45 条
  • [1] Using OVH and IMRT Plan Data to Automate VMAT Planning: A Head-And-Neck Study
    Wu, B.
    Pang, D.
    Sanguineti, G.
    Taylor, R.
    McNutt, T.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3597 - 3597
  • [2] Automate IMRT Planning in Pinnacle: A Study in Head-And-Neck Cancer
    Kusters, M.
    Bzdusek, K.
    Kumar, P.
    van Kollenburg, P.
    Kunze-Busch, M.
    Kaanders, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Fully Automated IMRT Planning is Feasible for Head and Neck Cancer: A Prospective Study using an Overlap Volume Histogram (OVH) Strategy
    Wu, B.
    Mcnutt, T.
    Zahurak, M.
    Kazhdan, M.
    Simari, P.
    Pang, D.
    Taylor, R.
    Sanguineti, G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S198 - S198
  • [4] Full Automation of Head-And-Neck Contouring, VMAT Planning, and Independent Plan Checks
    McCaroll, R.
    Balter, P.
    Beadle, B.
    Burger, H.
    Cardenas, C.
    Dalvic, S.
    Followill, D.
    Kisling, K.
    Mejia, M.
    Naidoo, K.
    Nelson, C.
    Peterson, C.
    Vorster, K.
    Wetter, J.
    Zhang, L.
    Yang, J.
    Court, L.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E626 - E627
  • [5] NTCP reduction in head-and-neck cancer VMAT using robust planning
    van Schie, M.
    Petoukhova, A.
    Rossi, L.
    Kwakkel-Huizenga, L.
    Hutschemaekers, S.
    Heijmen, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S355 - S356
  • [6] Dosimetric plan comparison of VMAT vs. IMRT in 257 patients with head-and-neck and prostate cancer
    Ostheimer, C.
    Ensminger, S.
    Janich, M.
    Sieker, F.
    Izaguirre, V.
    Gerlach, R.
    Vordermark, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1009 - S1010
  • [7] Correction to: Automated IMRT planning in Pinnacle—A study in head-and-neck cancer
    J. M. A. M. Kusters
    K. Bzdusek
    P. Kumar
    P. G. M van Kollenburg
    M. C. Kunze-Busch
    M. Wendling
    T. Dijkema
    J. H. A. M. Kaanders
    [J]. Strahlentherapie und Onkologie, 2017, 193 : 1077 - 1078
  • [8] Prediction of Dose Discrepancies Using Differences in Plan Normalization Values of Head-And-Neck VMAT Plans
    Sasaki, M.
    Ikushima, H.
    Kobuchi, S.
    Kuwahara, K.
    Kajino, A.
    Sato, D.
    Sugimoto, W.
    Oita, M.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [9] Plan N Check: An Open-Source Tool for Automatic Head-And-Neck VMAT Planning in Varian Eclipse™
    Sample, C.
    Clark, H.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [10] A Comparative Study of Standard IMRT and VMAT Planning Techniques for Unilateral and Bilateral Head & Neck Irradiation
    Pursley, J.
    Tishler, R. B.
    Margalit, D. N.
    Sher, D. J.
    Damato, A. L.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 361 - 361