Vacuum metrology based on refractive index of gas

被引:6
|
作者
Xu Yu-Rong [1 ]
Liu Yang-Yang [1 ]
Wang Jin [1 ,2 ]
Sun Yu [1 ,2 ]
Xi Zhen-Hua [3 ]
Li De-Tian [3 ]
Hu Shui-Ming [1 ,2 ]
机构
[1] Univ Sci & Technol China, Hefei Natl Lab Phys Sci Microscale, Hefei 230026, Peoples R China
[2] Univ Sci & Technol China, CAS Ctr Excellence Quantum Informat & Quantum Phy, Hefei 230026, Peoples R China
[3] Lanzhou Inst Phys, Sci & Technol Vacuum Technol & Phys Lab, Lanzhou 730000, Peoples R China
关键词
vacuum metrology; gas pressure; Virial state equation; Fabry-Perot cavity;
D O I
10.7498/aps.69.20200706
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
With the development of vacuum technology, subject to the influence of directional flow and uneven temperature, the thermodynamic equilibrium state is destroyed. In this case, the pressure reference is not suitable for characterizing the vacuum state. To ensure the long-term stability and reproducibility of the measurement system, vacuum metrology will be characterized by gas density. The precisive measurement of gas refractive index based on a Fabry-Perot cavity can be used to derive the gas density. This kind of an optical measurement of vacuum links macroscopic dielectric constants of gases with microscopic polarization parameters of atoms and molecules. It replaces the physical standard based on the mercury pressure gauge with the quantum standard. In this paper, we discuss the reverse process from refractive index to gas pressure, and use the laser-locked Fabry-Perot cavity method to measure the refractive index of argon gas. The contribution of related parameters to the uncertainty of determined gas pressure is analyzed. The influences of material parameters and experimental parameters such as gas molar susceptibility, molar susceptibility, dielectric second Virial coefficient and temperature on gas pressure accuracy are analyzed. The result shows that the uncertainty in our measurement of argon within 1 atm is u = root(6 mPa)(2) + (73 x 10(-6) p)(2). Currently, the uncertainty mainly comes from the measurement deviation of gas temperature inside the cavity. After repeating the measurement a few times, the results show that the statistical uncertainty of refractive index is within 100 ppm, which is limited by the accuracy of the pressure gauge used here. In addition, we compare the dipole calculated by the ab initio method with that by the DOSD method. The results show that the dynamic polarizability obtained by the ab initio method is consistent with our experimental results. In conclusion, these experimental results show that the measurement of gas pressure based on the gas refractive index has high repeatability and accuracy. If the temperature control and corresponding measurement accuracy of the gas are further improved, this method can also be used to obtain high-precision microscopic parameters such as the polarizabilities of atoms and molecules. In the future work, we will focus on improving the temperature control and the design of the cavity to reduce cavity leakage and deflation. It is possible that the measurement accuracy of the gas pressure will be increased to 10 ppm level, which is the same level as the current standard pressure gauge and will become a new standard for pressure measurement in the future.
引用
收藏
页数:7
相关论文
共 21 条
  • [1] Buckley T J, 2000, REV SCI
  • [2] Egan P, 2016, 2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016)
  • [3] Absolute refractometry of dry gas to ±3 parts in 109
    Egan, Patrick
    Stone, Jack A.
    [J]. APPLIED OPTICS, 2011, 50 (19) : 3076 - 3086
  • [4] Measured relationship between thermodynamic pressure and refractivity for six candidate gases in laser barometry
    Egan, Patrick F.
    Stone, Jack A.
    Scherschligt, Julia K.
    Harvey, Allan H.
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 2019, 37 (03):
  • [5] Cell-based refractometer for pascal realization
    Egan, Patrick F.
    Stone, Jack A.
    Ricker, Jacob E.
    Hendricks, Jay H.
    Strouse, Gregory F.
    [J]. OPTICS LETTERS, 2017, 42 (15) : 2944 - 2947
  • [6] Comparison measurements of low-pressure between a laser refractometer and ultrasonic manometer
    Egan, Patrick F.
    Stone, Jack A.
    Ricker, Jacob E.
    Hendricks, Jay H.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (05):
  • [7] Highly-accurate density-virial-coefficient values for helium, neon, and argon at 0.01 °C determined by dielectric-constant gas thermometry
    Gaiser, Christof
    Fellmuth, Bernd
    [J]. JOURNAL OF CHEMICAL PHYSICS, 2019, 150 (13):
  • [8] Polarizability of Helium, Neon, and Argon: New Perspectives for Gas Metrology
    Gaiser, Christof
    Fellmuth, Bernd
    [J]. PHYSICAL REVIEW LETTERS, 2018, 120 (12)
  • [9] Gibney E, 2017, NATURE, V551, P18
  • [10] ON DIAMAGNETIC SUSCEPTIBILITY OF GASES
    GLICK, RE