Comparison of the Accuracy of the Novel PrisMax Continuous Renal Replacement Therapy System to the Classic Prismaflex System

被引:5
|
作者
Bell, Max [1 ,2 ]
Broman, Marcus [3 ]
Joannes-Boyau, Olivier [4 ]
Ronco, Claudio [5 ,6 ]
机构
[1] Karolinska Univ Hosp, Dept Perioperat Med & Intens Care, SE-17176 Stockholm, Sweden
[2] Karolinska Inst, Dept Physiol & Pharmacol, Stockholm, Sweden
[3] Skane Univ Hosp, Dept Perioperat & Intens Care, Lund, Sweden
[4] CHU Bordeaux, Dept Anesthesie Reanimat Sud, Bordeaux, France
[5] St Bortolo Hosp, Dept Nephrol, Vicenza, Italy
[6] IRRIV, Vicenza, Italy
关键词
Acute kidney injury; Citrate; Continuous hemodiafiltration; Critical care; Dialysis efficiency; Dialysis machine; Hemodiafiltration; CRITICALLY-ILL PATIENTS; INTENSITY;
D O I
10.1159/000494764
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background/Aims: We assessed how the novel PrisMax continuous renal replacement therapy (CRRT) system performed in an international multicentre setting. The system has multiple novel tools aiming to increase accuracy and dose delivery. Methods: Data was prospectively collected from 7 intensive care units in 6 countries. The PrisMax device data logs constituted the raw material and last generation Prismaflex data was used as comparison. Clinical parameters like treatment time, filter life span, downtime as well as prescribed and delivered dose were recorded. Results: PrisMax delivered/prescribed effluent ratios (mean +/- SD) 0.92 +/- 0.15 vs. Prismaflex ratios 0.85 +/- 0.21, p < 0.001; delivered effluent dose (mL/kg/h) was 18.16 +/- 12.93 vs. 10.95 +/- 10.96, p < 0.0001; and (Kt/V) 0.76 +/- 0.52 vs. 0.44 +/- 0.44, p < 0.0001. Moreover, downtime was 27 minutes less for the newer device. Conclusion: The PrisMax CRRT device outperforms its predecessor with regard to dose delivery and accuracy. (c) 2018 S. Karger AG, Basel
引用
收藏
页码:166 / 170
页数:5
相关论文
共 50 条
  • [1] The Novel PrisMax Continuous Renal Replacement Therapy System in a Multinational, Multicentre Pilot Setting
    Broman, Marcus
    Bell, Max
    Joannes-Boyau, Olivier
    Ronco, Claudio
    [J]. BLOOD PURIFICATION, 2018, 46 (03) : 220 - 227
  • [2] Regional Citrate Anticoagulation for PrismaFlex Continuous Renal Replacement Therapy
    Burry, Lisa D.
    Tung, David D.
    Hallett, David
    Bailie, Toni
    Carvalhana, Virginia
    Lee, David
    Rarriganesh, Steve
    Richardson, Robert
    Mehta, Sangeeta
    Lapinsky, Stephen E.
    [J]. ANNALS OF PHARMACOTHERAPY, 2009, 43 (09) : 1419 - 1425
  • [3] The accuracy of a continuous volumetric balancing system in pediatric continuous renal replacement therapy
    Hanudel, Mark R.
    Salusky, Isidro B.
    Zaritsky, Joshua J.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2014, 37 (03): : 215 - 221
  • [4] Clinical evaluation of the Prismaflex™ HF 20 set and Prismaflex™ system 7.10 for acute continuous kidney replacement therapy (CKRT) in children
    Raj Munshi
    Kathy Lee-Son
    Richard M. Hackbarth
    Raymond Quigley
    Scott M. Sutherland
    Jorge Echeverri
    Stuart L. Goldstein
    [J]. Pediatric Nephrology, 2020, 35 : 2345 - 2352
  • [5] Clinical evaluation of the Prismaflex™ HF 20 set and Prismaflex™ system 7.10 for acute continuous kidney replacement therapy (CKRT) in children
    Munshi, Raj
    Lee-Son, Kathy
    Hackbarth, Richard M.
    Quigley, Raymond
    Sutherland, Scott M.
    Echeverri, Jorge
    Goldstein, Stuart L.
    [J]. PEDIATRIC NEPHROLOGY, 2020, 35 (12) : 2345 - 2352
  • [6] Prismaflex HF20 for Continuous Renal Replacement Therapy in Critically Ill Children
    Jose Santiago, Maria
    Lopez-Herce, Jesus
    [J]. ARTIFICIAL ORGANS, 2011, 35 (12) : 1194 - 1194
  • [7] Design and validation of a novel fluid management system for pediatric continuous renal replacement therapy
    Santhanakrishnan, Arvind
    Nestle, T.
    Moore, B. L.
    Yoganathan, A. P.
    Paden, M. L.
    [J]. PEDIATRIC NEPHROLOGY, 2013, 28 (01) : 183 - 183
  • [8] Outcomes of Continuous Renal Replacement Therapy in a Community Health System
    Schaffer, Patrick
    Chowdhury, Raad
    Jordan, Kim
    DeWitt, Jordan
    Elliott, John
    Schroeder, Kevin
    [J]. JOURNAL OF INTENSIVE CARE MEDICINE, 2022, 37 (08) : 1043 - 1048
  • [9] A COMPARISON OF MOLECULAR CLEARANCE RATES DURING CONTINUOUS HEMOFILTRATION AND HEMODIALYSIS WITH A NOVEL VOLUMETRIC CONTINUOUS RENAL REPLACEMENT SYSTEM
    JEFFREY, RF
    KHAN, AA
    PRABHU, P
    TODD, N
    GOUTCHER, E
    WILL, EJ
    DAVISON, AM
    [J]. ARTIFICIAL ORGANS, 1994, 18 (06) : 425 - 428
  • [10] Continuous renal replacement therapy for congestive heart failure: The wearable continuous ultrafiltration system
    Gura, V
    Beizai, M
    Ezon, C
    Rambod, E
    [J]. ASAIO JOURNAL, 2006, 52 (01) : 59 - 61