Percutaneous Stabilization System Osseofix® for Treatment of Osteoporotic Vertebral Compression Fractures - Clinical and Radiological Results after 12 Months

被引:22
|
作者
Ender, Stephan Albrecht [1 ]
Wetterau, Elmar [1 ]
Ender, Michaela [2 ]
Kuehn, Jens-Peter [2 ]
Merk, Harry Rudolf [1 ]
Kayser, Ralph [1 ]
机构
[1] Univ Med Greifswald, Dept Orthopaed & Orthopaed Surg, Greifswald, Germany
[2] Univ Med Greifswald, Dept Diagnost Radiol & Neuroradiol, Greifswald, Germany
来源
PLOS ONE | 2013年 / 8卷 / 06期
关键词
BALLOON KYPHOPLASTY; RISK-FACTORS; VERTEBROPLASTY; BODY; CEMENT; EXPERIENCE; EFFICACY;
D O I
10.1371/journal.pone.0065119
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Study Design: A prospective consecutive cohort study (follow-up study). Objective: Our study investigated whether implantation of an expandable titanium mesh cage (Osseofix (R)) is a successful and safe minimally invasive therapy for osteoporotic vertebral compression fractures (VCF). Our experiences, clinical and radiological findings after 12 months follow-up are presented. Kypho- and vertebroplasty are well-established minimally invasive procedures for the treatment of osteoporotic VCF. The main complications associated with both procedures are uncontrolled bone cement leakage. Therefore a suitable alternative has been investigated. Methods: During June 2010 to May 2011 24 patients were included with 32 osteoporotic VCF (T6 to L4). All of them were stabilized with the Osseofix (R) system. Preinterventionally we performed X-ray, MRI, and bone density measurements (DXA). Clinical and radiological results were evaluated preop., postop. and after 12 months postop. based on the Oswestry Disability Index (ODI) and the Visual Analogue Scale (VAS), X-ray (Beck Index, Cobb-angle) and CT. Results: There was a significant improvement in the mean ODI (70,6% to 30,1%) as well as a significant reduction in pain intensity (VAS) (7,7 to 1,4) after 12 month. The mean kyphotic angle according to Cobb showed significant improvements (11,7 degrees to 10,4 degrees) after 12 months. Postinterventional imaging showed only one case of loss of height in a stabilized vertebral body (3.1%). We saw no changes in posterior vertebral wall or adjacent fractures. Except for one pronounced postoperative hematoma we saw no surgical complications including no cement leakage. Conclusions: Stabilization of symptomatic osteoporotic VCF with Osseofix (R) system is a safe and effective procedure, even in fractures with posterior wall involvement. The clinical mid-term results are good at a very low complication rate. The Osseofix (R) system is an interesting alternative to the established procedures of cement augmentation.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Osseofix® System for Percutaneous Stabilization of Osteoporotic and Tumorous Vertebral Compression Fractures - Clinical and Radiological Results After 12 Months
    Ender, S. A.
    Gradl, G.
    Ender, M.
    Langner, S.
    Merk, H. R.
    Kayser, R.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2014, 186 (04): : 380 - 387
  • [2] Treatment of osteoporotic vertebral compression fractures with percutaneous vertebroplasty under local anesthesia: clinical and radiological results
    Balkarli, Huseyin
    Demirtas, Hakan
    Kilic, Mesut
    Ozturk, Ibrahim
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (09): : 16287 - 16293
  • [3] Percutaneous vertebroplasty for the treatment of osteoporotic vertebral compression fractures EVALUATION AFTER 36 MONTHS
    Muijs, S. P. J.
    Nieuwenhuijse, M. J.
    Van Erkel, A. R.
    Dijkstra, P. D. S.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (03): : 379 - 384
  • [4] Clinical analysis of percutaneous vertebroplasty in the treatment of osteoporotic vertebral compression fractures
    Shang, Q.
    Zhang, S.
    Tang, Y. D.
    JOURNAL OF BIOLOGICAL REGULATORS AND HOMEOSTATIC AGENTS, 2020, 34 (06): : 2277 - 2280
  • [5] Radiofrequency Kyphoplasty for the Treatment of Osteoporotic and Neoplastic Vertebral Compression Fractures - Initial Experience and Clinical Results After 6 Months
    Elgeti, F.
    Gebauer, B.
    JOURNAL FUR MINERALSTOFFWECHSEL, 2010, 17 (04): : 136 - 139
  • [6] The results of percutaneous balloon kyphoplasty technique for the treatment of osteoporotic vertebral compression fractures
    Akalin, S
    Çelik, B
    Erzen, F
    Benli, T
    Kiliç, M
    OSTEOPOROSIS INTERNATIONAL, 2004, 15 (07) : 587 - 588
  • [7] Risk factors for new vertebral fractures after percutaneous vertebroplasty or percutaneous kyphoplasty in the treatment of osteoporotic vertebral compression fractures
    Yang, Wencheng
    Zou, Kaiwei
    Lin, Xuping
    Yang, Yanfang
    Chen, Tianpei
    Wu, Xiuming
    Wang, Xiaomeng
    Liu, Qingjun
    Huang, Chunhui
    Su, Wanhan
    FRONTIERS IN MEDICINE, 2025, 12
  • [8] The clinical effect of percutaneous kyphoplasty for the treatment of multiple osteoporotic vertebral compression fractures and the prevention of new vertebral fractures
    Zhai, Weifeng
    Jia, Yongwei
    Wang, Jianjie
    Cheng, Liming
    Zeng, Zhili
    Yu, Yan
    Chen, Lei
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (08): : 13473 - 13481
  • [9] Clinical and radiological subsequent fractures after vertebral augmentation for treating osteoporotic vertebral compression fractures: a meta-analysis
    Hai-Bo Sun
    Xiao-Shan Jing
    Hai Tang
    Yong Hai
    Jin-Jun Li
    Jian-Lin Shan
    De-Cheng Wang
    European Spine Journal, 2020, 29 : 2576 - 2590
  • [10] Clinical and radiological subsequent fractures after vertebral augmentation for treating osteoporotic vertebral compression fractures: a meta-analysis
    Sun, Hai-Bo
    Jing, Xiao-Shan
    Tang, Hai
    Hai, Yong
    Li, Jin-Jun
    Shan, Jian-Lin
    Wang, De-Cheng
    EUROPEAN SPINE JOURNAL, 2020, 29 (10) : 2576 - 2590