Silicone oil adhesion to intraocular lenses: An experimental study comparing various biomaterials

被引:64
|
作者
Apple, DJ
Isaacs, RT
Kent, DG
Martinez, LM
Kim, S
Thomas, SG
Basti, S
Barker, D
Peng, Q
机构
[1] Department of Ophthalmology, Medical University of South Carolina, Charleston, SC 29425-2236
来源
关键词
D O I
10.1016/S0886-3350(97)80210-6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To perform an in vitro experimental study comparing the degree of adherence of silicone oil to various rigid and foldable intraocular lens (IOL) designs and to the human lens capsule. Setting: Center for Research on Ocular Therapeutics and Biodevices, Department of Ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA. Methods: Seven IOL styles comprising various biomaterials were studied: fluorine-treated (Fluorlens(TM)), heparin-surface-modified (HSM(TM)), hydrogel, Memory-Lens(TM), poly(methyl methacrylate) (PMMA), soft acrylic, and silicone lenses; the human crystalline lens was also studied. Each lens was immersed in silicone oil for 12 hours, then photographed, studied by scanning electron microscopy (except the crystalline lens), and subjected to computer-generated image analysis to determine the silicone oil coverage. Results: Silicone oil coverage of dry silicone lenses was 100% and of lenses immersed in normal saline, 82.5%. The least coverage was on the heparin-surface-modified lens (mean score 9.4%). Coverage of the other four lenses ranged from approximately 15.1% to 33.7%. Mean coverage of the human lens capsule was 10.9%. Conclusion: Although a silicone IOL shows maximal adherence to silicone oil, other lens biomaterials are not immune to this complication. Silicone oil coverage was related to the dispersive energy component of the surface charge of the IOL biomaterial. Low dispersive energy materials had less silicone oil coverage, while those with higher dispersive energy had more oil coverage.
引用
收藏
页码:536 / 544
页数:9
相关论文
共 50 条
  • [1] Solvent for removing silicone oil from intraocular lenses - Experimental study comparing various biomaterials
    Dick, HB
    Augustin, AJ
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (11): : 1667 - 1672
  • [2] Interaction of intraocular lenses with various concentrations of silicone oil:: An experimental study
    Öner, FH
    Saatcu, OA
    Sarioglu, S
    Durak, I
    Kaynak, S
    Çabuk, M
    OPHTHALMOLOGICA, 2003, 217 (02) : 124 - 128
  • [3] Intraocular silicone lenses in silicone oil: An experimental study
    Stolba, U
    Binder, S
    Velikay, M
    Wedrich, A
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1996, 234 (01) : 55 - 57
  • [4] Silicone oil adhesion characteristics of different intraocular lenses
    Potamitis, T
    Tsaloumas, M
    Stavrakas, Y
    Benson, MT
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1997, 38 (04) : 888 - 888
  • [5] Irreversible silicone oil adhesion to silicone intraocular lenses - A clinicopathologic analysis
    Apple, DJ
    Federman, JL
    Krolicki, TJ
    Sims, JCR
    Kent, DG
    Hamburger, HA
    Smiddy, WE
    Cox, MS
    Hassan, TS
    Compton, SM
    Thomas, SG
    OPHTHALMOLOGY, 1996, 103 (10) : 1555 - 1561
  • [6] SILICONE INTRAOCULAR LENSES AND SILICONE OIL
    BARTZSCHMIDT, KU
    KONEN, W
    ESSER, P
    WALTER, P
    HEIMANN, K
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1995, 207 (03) : 162 - 166
  • [8] Silicone oil adherence to silicone intraocular lenses
    Arthur, SN
    Peng, Q
    Escobar-Gomez, M
    Apple, DJ
    INTERNATIONAL OPHTHALMOLOGY CLINICS, 2001, 41 (03) : 33 - 45
  • [9] Cataract operation after silicone oil surgery - oil-adhesion to silicone intraocular lenses (IOLs)
    Weber, U
    Bullerkotte, J
    OPHTHALMOLOGE, 1998, 95 (04): : 219 - 224
  • [10] Interaction between silicone oil and silicone intraocular lenses: An in vitro study
    Senn, P
    Schmid, MK
    Schipper, I
    Hendrickson, P
    OPHTHALMIC SURGERY AND LASERS, 1997, 28 (09): : 776 - 779