Major FDA medical device recalls in ophthalmology from 2003 to 2015

被引:6
|
作者
Talati, Rushi K. [1 ]
Gupta, Ankur S. [2 ]
Xu, Shuai [3 ]
Ghobadi, Comeron W. [4 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Chicago, IL 60611 USA
[2] Univ Chicago, NorthShore Hosp, Chicago, IL 60637 USA
[3] Northwestern Univ, Dept Dermatol, Feinberg Sch Med, Chicago, IL 60611 USA
[4] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
关键词
PREMARKET APPROVAL PATHWAY; INNOVATION; IMPAIRMENT;
D O I
10.1016/j.jcjo.2017.08.001
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To assess recent high-risk ophthalmic medical device recalls. Methods: The publicly available Food and Drug Administration Center for Devices and Radiological Health database was mined for Class I (high-risk) ophthalmic device recalls from January 1, 2003 to December 31, 2015. The number of Class I ophthalmic device recalls was quantified. Additionally, recall characteristics and market entry data were determined for each device. Results: Twelve Class I ophthalmic device recall events were identified, collectively affecting over 68 million units in distribution. A median of 147,491 units (range 20 to 57,252,581) were recalled per event. 9 out of 12 recalls (75%) had at least one documented occurrence of an adverse event to a patient. Pre-market related issues accounted for one device recall (8%), post-market related issues accounted for nine device recalls (75%), and two device recalls (17%) were indeterminate. 510(k) clearance was the most common pathway to market, accounting for 50% of Class I recalls. Three devices were approved through pre-market approval (PMA) pathway, two devices were exempt from review, and one device failed to register with the FDA. Conclusion: Class I recalls surrounding ophthalmology are relatively infrequent compared to other medical specialties. However, given the impact of Class I recalls in the field, ophthalmologists have an impetus to advocate for stronger device regulation particularly in the context of post-marketing surveillance.
引用
收藏
页码:98 / 103
页数:6
相关论文
共 50 条
  • [1] Medical Device Safety in Gastroenterology: FDA Recalls of Duodenoscopes, 2015-2020
    Talati, Rushi K.
    Goodman, Adam
    Pochapin, Mark B.
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2021, 116 : S474 - S474
  • [2] FDA Class I recalls of approved ophthalmic drugs and medical devices from 2003-2015
    Gupta, Ankur
    Talati, Rushi K.
    Xu, Steve
    Ghobadi, Comeron
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [3] MEDICAL DEVICE RECALLS IN CANADA FROM 2005 TO 2015
    Gagliardi, Anna R.
    Takata, Julie
    Ducey, Ariel
    Lehoux, Pascale
    Ross, Sue
    Trbovich, Patricia L.
    Easty, Anthony
    Bell, Chaim M.
    Urbach, David R.
    INTERNATIONAL JOURNAL OF TECHNOLOGY ASSESSMENT IN HEALTH CARE, 2017, 33 (06) : 708 - 714
  • [4] Medical Device Recalls and the FDA Approval Process
    Zuckerman, Diana M.
    Brown, Paul
    Nissen, Steven E.
    ARCHIVES OF INTERNAL MEDICINE, 2011, 171 (11) : 1006 - 1011
  • [5] Study of Software-Related Causes in the FDA Medical Device Recalls
    Fu, Zhicheng
    Guo, Chunhui
    Zhang, Zhenyu
    Ren, Shangping
    Jiang, Yu
    Sha, Lui
    2017 22ND INTERNATIONAL CONFERENCE ON ENGINEERING OF COMPLEX COMPUTER SYSTEMS (ICECCS), 2017, : 60 - 69
  • [6] Preventing medical device recalls
    1600, Association for the Advancement of Medical Instrumentation (48):
  • [7] Medical device recalls from 2004 to 2006: A focus on class I recalls
    Villarraga, Marta L.
    Guerin, Heather L.
    Lam, Tack
    FOOD AND DRUG LAW JOURNAL, 2007, 62 (03) : 581 - 591
  • [8] Evaluation of orthopedic hip device recalls by the FDA from 2007 to 2017
    Peters W.
    Pellerin C.
    Janney C.
    Biomedical Instrumentation and Technology, 2020, 54 (06): : 418 - 426
  • [9] Medical device recalls and transparency in the UK
    Thompson, Matthew
    Heneghan, Carl
    Billingsley, Matthew
    Cohen, Deborah
    BRITISH MEDICAL JOURNAL, 2011, 342
  • [10] USER PROBLEMS AND MEDICAL DEVICE RECALLS
    HYMAN, WA
    SCHLAIN, LA
    MEDICAL INSTRUMENTATION, 1986, 20 (01): : 14 - 16