Calibration for isotope dilution mass spectrometry - Description of an alternative to the bracketing procedure

被引:0
|
作者
Thienpont, LM [1 ]
VanNieuwenhove, B [1 ]
Stockl, D [1 ]
DeLeenheer, AP [1 ]
机构
[1] STATE UNIV GHENT,LAB MED BIOCHEM & KLIN ANAL,B-9000 GHENT,BELGIUM
来源
JOURNAL OF MASS SPECTROMETRY | 1996年 / 31卷 / 10期
关键词
calibration; isotope dilution mass spectrometry; reference methods; bracketing; accuracy;
D O I
10.1002/(SICI)1096-9888(199610)31:10<1119::AID-JMS404>3.0.CO;2-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A calibration method for isotope dilution mass spectrometry is presented that fully accounts for non-linearity of calibration functions, caused by the interference of the analyte on the m/z used for measurement of the internal standard and vice versa. In this way, it is also possible to use incompletely labelled molecules or analogues with a mass increment of only 1 u for isotope dilution, on condition that certain restrictions are respected. In addition, the proposed method is less time consuming than bracketing. The method works with the computer-stored full calibration curve and a single calibration point that is measured daily. The calibration curve is constructed from the experimentally determined mass spectral overlap between the unlabelled analyte and the labelled internal standard at the mit values chosen for measurement. Measurement results for samples with unknown analyte concentration are calculated on the basis of a linear relationship between their ion abundance ratio and that of the daily single calibration point, but corrected by a factor derived from the theoretical calibration function. All calculations are performed with commonly available spreadsheet software. An application is presented far determining serum uric acid with a candidate gas chromatographic/isotope dilution mass spectrometric reference method.
引用
收藏
页码:1119 / 1125
页数:7
相关论文
共 50 条
  • [1] Calibration graphs in isotope dilution mass spectrometry
    Pagliano, Enea
    Mester, Zoltan
    Meija, Juris
    ANALYTICA CHIMICA ACTA, 2015, 896 : 63 - 67
  • [2] Evaluation of a bracketing calibration-based isotope dilution liquid chromatography–tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma
    Qiaoxuan Zhang
    Lu Zhang
    Haibiao Lin
    Zhiliang Cai
    Jun Yan
    Qiqin Wang
    Liqiao Han
    Jianbing Wang
    Peifeng Ke
    Junhua Zhuang
    Xianzhang Huang
    Analytical and Bioanalytical Chemistry, 2019, 411 : 7095 - 7104
  • [3] Evaluation and clinical application of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for serum theophylline
    Du, Yuanyuan
    Yan, Qiaofang
    Zhan, Min
    Zhang, Qiaoxuan
    Huang, Di
    Zhang, Pengwei
    Yan, Jun
    Wang, Jianbing
    Huang, Xianzhang
    Han, Liqiao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (22) : 4897 - 4906
  • [4] Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma
    Zhang, Qiaoxuan
    Zhang, Lu
    Lin, Haibiao
    Cai, Zhiliang
    Yan, Jun
    Wang, Qiqin
    Han, Liqiao
    Wang, Jianbing
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (27) : 7095 - 7104
  • [5] Comparison of bracketing calibration and classical calibration curve quantification methods in establishing a candidate reference measurement procedure for human serum 17β-estradiol by isotope dilution liquid chromatography tandem mass spectrometry
    Zhang, Qiaoxuan
    Cai, Zhiliang
    Liu, Yingyi
    Lin, Haibiao
    Wang, Qiqin
    Yan, Jun
    Han, Liqiao
    Wang, Jianbing
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    MICROCHEMICAL JOURNAL, 2020, 152
  • [6] EVALUATION OF LUTETIUM FOR VOLUMETRIC CALIBRATION BY ISOTOPE-DILUTION MASS-SPECTROMETRY
    SMITH, DH
    WALKER, RL
    PRITCHARD, CA
    CARTER, JA
    ANALYTICAL CHEMISTRY, 1983, 55 (03) : 578 - 580
  • [7] Measurement of urea in human serum by isotope dilution mass spectrometry: A reference procedure
    Kessler, A
    Siekmann, L
    CLINICAL CHEMISTRY, 1999, 45 (09) : 1523 - 1529
  • [8] Development of a high ratio isotope dilution mass spectrometry calibration technique for chemical metrology
    Hearn, R
    Evans, P
    Sargent, M
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2005, 20 (10) : 1019 - 1023
  • [10] A rapid isotope dilution inductively coupled plasma mass spectrometry procedure for uranium bioassay
    Kinman, William S.
    Lamont, Stephen P.
    Steiner, Robert E.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2009, 282 (03) : 1027 - 1030