cognitive functions;
magnetic resonance imaging;
safety;
static magnetic field;
STRAY FIELDS;
MRI SCANNERS;
TASK;
PERFORMANCE;
SAFETY;
INTERFERENCE;
D O I:
10.1002/jmri.23727
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Purpose: To test for potential changes in higher-order cognitive processes related to the exposure to a high static magnetic field. Materials and Methods: Twenty-four healthy volunteers participated in two experimental sessions inside a 3 Tesla (T) magnetic resonance imaging (MRI) magnet. During one session the magnetic field was ramped down. The tasks consisted of six well-established paradigms probing a variety of cognitive functions. Reaction times (RT) and accuracies (AC) were recorded for statistical analysis. Results: The overall performance was very similar in both sessions. Strong task-specific effects (all P < 0.006) were consistent with previously published results. Direct comparisons of task-specific effects between the two sessions (magnetic field on or off) remained insignificance for all paradigms (RT: all P > 0.196; AC: all P > 0.17; no corrections for multiple comparisons). Conclusion: The results did not indicate any apparent safety concerns with respect to cognitive performance in a static magnetic field of a typical whole-body magnet. In addition, comparisons of cognitive effects from testing situations with and without exposure to high static magnetic fields can be considered valid. J. Magn. Reson. Imaging 2012;36:835840. (c) 2012 Wiley Periodicals, Inc.