Research on improving the accuracy of near infrared non-invasive hemoglobin detection

被引:14
|
作者
Yuan, Jingze [1 ,2 ]
Ding, Haiquan [1 ]
Gao, Hongzhi [1 ]
Lu, Qipeng [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Opt Fine Mech & Phys, State Key Lab Appl Opt, Changchun 130033, Jilin, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
基金
国家高技术研究发展计划(863计划); 中国国家自然科学基金;
关键词
Near-infrared spectroscopy; Non-invasive detection; Directs orthogonal signal correction; Hemoglobin concentration;
D O I
10.1016/j.infrared.2015.06.009
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To optimize the accuracy of near-infrared non-invasive hemoglobin (Hb) clinical detection, high-performance instrument and prepossessing algorithm have been investigated. A near-infrared spectrophotometric system was constructed adopting InGaAs detector array with 16 pixels and plane grating spectrometer to obtain high signal noise ratio (SNR) spectral data. In our experiment, we applied the device independently to collect spectra data from 91 volunteers' fingertips non-invasively. Two prediction tests were conducted to verify the effects of preprocessing algorithms improving the accuracy of near-infrared Mb detection and exclude the occasionality of satisfactory results in a single trial. Our non-invasive Hb detection methods were based on partial least squares (PLS). In each test, PLS, MSC coupled with PLS, DOSC coupled with PLS, three methods for non-invasive Hb detection, were analyzed respectively. The results of two trials showed that only DOSC & PLS performed excellently in both predictive ability and stability, obviously better than other two methods. Relative RMSEP was 6.16% in predicting test 1, 6.08% in predicting test 2, almost reaching the requirements of clinical application. It indicates that our independent-developed high-performance instrument and the method DOSC coupled with PLS are promising in non-invasive Mb detection clinical application. (c) 2015 Published by Elsevier B.V.
引用
收藏
页码:117 / 121
页数:5
相关论文
共 50 条
  • [1] A feasibility study on improving the non-invasive detection accuracy of bottled Shuanghuanglian oral liquid using near infrared spectroscopy
    Cui, Pengdi
    Wang, Qiuyue
    Li, Zheng
    Wu, Chenlu
    Li, Gang
    Zhao, Jing
    Liu, Ming
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2022, 274
  • [2] Improving non-invasive hemoglobin measurement accuracy using nonparametric models
    Man, Jianing
    Zielinski, Martin D.
    Das, Devashish
    Wutthisirisart, Phichet
    Pasupathy, Kalyan S.
    JOURNAL OF BIOMEDICAL INFORMATICS, 2022, 126
  • [3] Accuracy of a continuous non-invasive hemoglobin monitor in the ICU
    D Frasca
    C Dahyot-Fizelier
    C Karen
    Q Levrat
    N Soumagne-Vialle
    M Boisson
    B Debaene
    O Mimoz
    Critical Care, 15 (Suppl 1):
  • [4] Non-invasive detection of brain microvascular dysfunction by near infrared spectroscopy
    Vayman, V.
    Martsiano, S.
    Weinstein, M.
    Nachum-Biala, Y.
    Troen, A.
    JOURNAL OF MOLECULAR NEUROSCIENCE, 2013, 51 : S126 - S127
  • [5] DETECTION OF ANEMIA BY NON-INVASIVE HEMOGLOBIN SENSOR
    Weinstein, A.
    Herzenstein, O.
    Kononenko, A.
    Gabis, E.
    Korenberg, A.
    Hod, M.
    VOX SANGUINIS, 2013, 105 : 107 - 108
  • [6] NON-INVASIVE DETERMINATION OF CEREBRAL VENOUS HEMOGLOBIN SATURATION IN THE DOG BY NEAR-INFRARED SPECTROSCOPY
    FERRARI, M
    HARTMANN, JF
    HANLEY, DF
    WILSON, DA
    TRAYSTMAN, RJ
    FASEB JOURNAL, 1988, 2 (04): : A508 - A508
  • [7] Feasibility research on non-invasive analysis of Tanreqing injection with near infrared spectroscopy
    Li, Wenlong
    Cheng, Zhiwei
    Wang, Yuefei
    Qu, Haibin
    JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2012, 20 (06) : 667 - 674
  • [8] Research of transmissive near infrared spectroscopy for non-invasive blood glucose measurement
    Yang, Wenming
    Liao, Ningfang
    Li, Yasheng
    Shao, Liwei
    Huang, Dehuang
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS VII, 2017, 0024
  • [9] Non-invasive detection of biological tissue oxygenation using near infrared spectroscopy
    Wang, Feng
    Li, Wei
    Lin, Fang
    Wang, Guangzhi
    Ding, Haishu
    Qinghua Daxue Xuebao/Journal of Tsinghua University, 1999, 39 (07): : 16 - 19
  • [10] Accuracy of non-invasive hemoglobin level measurement in the emergency department
    Aseri, Zohair Ahmed Ali Al
    Aldawood, Badr Abdulrahman
    Altamimi, Adel Abdullah
    Mosleh, Haytam Issa
    Qaw, Amnah Samir
    Albatran, Hawra Ibraheem
    AlMasri, Maram Mohammed
    AlRihan, Tasneem Muhammed
    AlAithan, Fatimah Bassem
    AlAssaf, Laura Suliaman
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2023, 64 : 200 - 203