Complementary Keratoconus Indices Based on Topographical Interpretation of Biomechanical Waveform Parameters: A Supplement to Established Keratoconus Indices

被引:11
|
作者
Goebels, Susanne [1 ]
Eppig, Timo [2 ]
Wagenpfeil, Stefan [3 ]
Cayless, Alan [4 ]
Seitz, Berthold [1 ]
Langenbucher, Achim [2 ]
机构
[1] Univ Saarland, Dept Ophthalmol, Med Ctr, Kirrberger Str 100,Bldg 22, D-66421 Homburg, Saar, Germany
[2] Univ Saarland, Dept Expt Ophthalmol, Kirrberger Str 100,Bldg 22, D-66421 Homburg, Saar, Germany
[3] Univ Saarland, Inst Med Biometry Epidemiol & Med Informat, Campus Homburg,Bldg 86, D-66421 Homburg, Saar, Germany
[4] Open Univ, Dept Phys Sci, Milton Keynes MK7 6AA, Bucks, England
关键词
MATCH INDEX;
D O I
10.1155/2017/5293573
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose. To build new models with the Ocular Response Analyzer (ORA) waveform parameters to create new indices analogous to established topographic keratoconus indices. Method. Biomechanical, tomographic, and topographic measurements of 505 eyes from the Homburger Keratoconus Centre were included. Thirty-seven waveform parameters (WF) were derived from the biomechanical measurement with the ORA. Area under curve (ROC, receiver operating characteristic) was used to quantify the screening performance. A logistic regression analysis was used to create two new keratoconus prediction models based on these waveform parameters to resample the clinically established keratoconus indices from Pentacam and TMS-5. Results. ROC curves show the best results for the waveform parameters p1area, p2area, h 1, h 2, dive1, mslew1, aspect1, aplhf, and dslope1. The new keratoconus prediction model to resample the Pentacam topographic keratoconus index (TKC) was WFTKC = -4.068 + 0.002 x p2area -0.005 x dive1 -0.01 x h1 -2.501 x aplhf, which achieves a sensitivity of 90.3% and specificity of 89.4%; to resample the TMS-5 keratoconus classification index (KCI) it was WFKCI = -3.606 + 0.002 x p2area, which achieves a sensitivity of 75.4% and a specificity of 81.8%. Conclusion. In addition to the biomechanically provided Keratoconus Index two new indices which were based on the topographic gold standards (either Pentacam or TMS-5) were created. Of course, these do not replace the original topographic measurement.
引用
收藏
页数:7
相关论文
共 22 条
  • [1] Keratoconus and Biomechanical Indices
    Peyman, Alireza
    Irajpour, Matin
    Pourazizi, Mohsen
    JOURNAL OF REFRACTIVE SURGERY, 2024, 40 (06)
  • [2] Interpretation of keratoconus indices
    Eppig, T.
    Spira-Eppig, C.
    Szentmary, N.
    Langenbucher, A.
    ACTA OPHTHALMOLOGICA, 2016, 94
  • [3] Correlation Between Corneal Biomechanical Indices and the Severity of Keratoconus
    Koh, Shizuka
    Inoue, Ryota
    Ambrosio Jr, Renato
    Maeda, Naoyuki
    Miki, Atsuya
    Nishida, Kohji
    CORNEA, 2020, 39 (02) : 215 - 221
  • [4] SENSITIVITY AND SPECIFICITY OF NOVEL BIOMECHANICAL WAVEFORM PARAMETERS IN DETECTING FORME FRUSTE KERATOCONUS AND KERATOCONUS
    Viswanathan, Deepa
    Males, John
    Kumar, Nikhil
    Graham, Stuart
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 42 : 87 - 87
  • [5] Staging of Keratoconus Indices Regarding Tomography, Topography, and Biomechanical Measurements
    Goebels, Susanne
    Eppig, Timo
    Wagenpfeil, Stefan
    Cayless, Alan
    Seitz, Berthold
    Langenbucher, Achim
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 159 (04) : 733 - 738
  • [6] Combined corneal biomechanical and tomographical indices in subclinical and forme fruste keratoconus
    Ganesh, Meghna
    Arora, Ritu
    Titiyal, Jeewan S.
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2024, 72 (09)
  • [7] Tomographic and Biomechanical Scheimpflug Imaging for Keratoconus Characterization: A Validation of Current Indices
    Steinberg, Johannes
    Siebert, Marlene
    Katz, Toam
    Frings, Andreas
    Mehlan, Juliane
    Druchkiv, Vasyl
    Buehren, Jens
    Linke, Stephan J.
    JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (12) : 840 - +
  • [8] Methodology for Keratoconus Risk Indices Using Ocular Response Analyzer (ORA) Waveform Analysis
    Luce, D.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [9] Screening Subclinical Keratoconus With Placido-Based Corneal Indices
    Ramos-Lopez, Dario
    Martinez-Finkelshtein, Andrei
    Castro-Luna, Gracia M.
    Burguera-Gimenez, Neus
    Vega-Estrada, Alfredo
    Pinero, David
    Alio, Jorge L.
    OPTOMETRY AND VISION SCIENCE, 2013, 90 (04) : 335 - 343
  • [10] Subclinical Keratoconus Detection Based On Pentacam Scheimpflug Tomography Indices
    Ruisenor Vazquez, Pablo
    Delrivo, Marianella
    Fuentes Bonthoux, Fernando
    Pfortner, Tomas
    Chiaradia, Pablo
    Galletti, Jeremias
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)