Performance validation of deformable image registration in the pelvic region

被引:25
|
作者
Zambrano, V. [1 ,2 ,3 ]
Furtado, H. [3 ,4 ]
Fabri, D. [3 ]
Luetgendorf-Caucig, C. [2 ]
Gora, J. [1 ,2 ]
Stock, M. [2 ,4 ]
Mayer, R. [1 ]
Birkfellner, W. [3 ,4 ]
Georg, D. [2 ,4 ]
机构
[1] EBG MedAustron GmbH, Wr Neustadt, Austria
[2] Med Univ Vienna AKH Wien, Dept Radiooncol, A-1090 Vienna, Austria
[3] Med Univ Vienna, Ctr Med Phys & Biomed Engn, A-1090 Vienna, Austria
[4] Med Univ Vienna, Christian Doppler Lab Med Radiat Res Radiat Oncol, A-1090 Vienna, Austria
关键词
radiotherapy; registration; deformable; organ; adaptive radiotherapy; PROSTATE-CANCER; DOSE ACCUMULATION; ADAPTIVE RADIOTHERAPY; MULTIMODALITY; THERAPY; ORGAN; MODEL; PROPAGATION; MOTION; TARGET;
D O I
10.1093/jrr/rrt045
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Patients undergoing radiotherapy will inevitably show anatomical changes during the course of treatment. These can be weight loss, tumour shrinkage, and organ motion or filling changes. For advanced and adaptive radiotherapy (ART) information about anatomical changes must be extracted from repeated images in order to be able to evaluate and manage these changes. Deformable image registration (DIR) is a tool that can be used to efficiently gather information about anatomical changes. The aim of the present study was to evaluate the performance of two DIR methods for automatic organ at risk (OAR) contour propagation. Datasets from ten gynaecological patients having repeated computed tomography (CT) and cone beam computed tomography (CBCT) scans were collected. Contours were delineated on the planning CT and on every repeated scan by an expert clinician. DIR using our in-house developed featurelet-based method and the iPlan((R)) BrainLab treatment planning system software was performed with the planning CT as reference and a selection of repeated scans as the target dataset. The planning CT contours were deformed using the resulting deformation fields and compared to the manually defined contours. Dice's similarity coefficients (DSCs) were calculated for each fractional patient scan structure, comparing the volume overlap using DIR with that using rigid registration only. No significant improvement in volume overlap was found after DIR as compared with rigid registration, independent of which image modality or DIR method was used. DIR needs to be further improved in order to facilitate contour propagation in the pelvic region in ART approaches.
引用
收藏
页码:120 / 128
页数:9
相关论文
共 50 条
  • [1] Validation of the deformable image registration system elastix in the head and neck region
    Zukauskaite, R.
    Brink, C.
    Hansen, C. R.
    Bertelsen, A.
    Grau, C.
    Johansen, J.
    Eriksen, J. G.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S499 - S500
  • [2] A validation framework to assess performance of commercial deformable image registration in lung radiotherapy
    Kumar, K.
    Gulal, O.
    Franich, R. D.
    Kron, T.
    Yeo, A. U.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 87 : 106 - 114
  • [3] Recommendations for validation and verification of deformable image registration in radiotherapy
    Bosma, L.
    Hussein, M.
    Jameson, M.
    Asghar, S.
    Brock, K.
    McClelland, J.
    Poeta, S.
    Yuen, J.
    Zachiu, C.
    Yeo, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S505 - S506
  • [4] Role and validation of deformable image registration in clinical practice
    van Herk, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S275 - S276
  • [5] Validation of three deformable image registration algorithms for the thorax
    Latifi, Kujtim
    Zhang, Geoffrey
    Stawicki, Marnix
    van Elmpt, Wouter
    Dekker, Andre
    Forster, Kenneth
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (01): : 19 - 30
  • [6] Characterization of Deformable Image Registration for the Pelvic Region Based On Prostate Shifting, Image Noise and the Existence of Implanted Fiducial Markers
    Ikeda, T.
    Takemura, A.
    Shinichi, U.
    Kojima, H.
    Noto, K.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3424 - 3424
  • [7] ASSESSING THE PERFORMANCE OF DEFORMABLE IMAGE REGISTRATION FOR PELVIS
    Kong, Vickie
    Moseley, Joanne
    Craig, Tim
    Chung, Peter
    Catton, Charles
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 120 : S26 - S26
  • [8] In-Vivo Correspondence Validation of a Deformable Image Registration Method
    Lian, J.
    Foskey, M.
    Wang, S.
    Shen, D.
    Chang, S.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [9] Physically-Based Validation of Deformable Medical Image Registration
    Lee, Huai-Ping
    Lin, Ming C.
    Foskey, Mark
    [J]. MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2008, PT II, PROCEEDINGS, 2008, 5242 : 830 - 838
  • [10] Validation of a deformable registration image of a commercial treated planning system
    Garcia-Molla, R.
    Marco-Blancas, N.
    Bonaque, J.
    Viduera, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S261 - S262