Multiple Mass Analysis Using an Ion Trap Array (ITA) Mass Analyzer

被引:9
|
作者
Xiao, Yu [1 ]
Chu, Yanqiu [1 ]
Ling, Xing [1 ]
Ding, Zhengzhi [1 ]
Xu, Chongsheng [1 ]
Ding, Li [2 ]
Ding, Chuan-Fan [1 ]
机构
[1] Fudan Univ, Dept Chem & Laser Chem Inst, Shanghai 200433, Peoples R China
[2] Shimadzu Res Lab Europe Ltd, Manchester M17 1GP, Lancs, England
关键词
Ion trap array; Mass analyzer; High-throughput; Simultaneous mass analyses; GEOMETRY OPTIMIZATION; ANALYTICAL PERFORMANCE; SPECTROMETER; IONIZATION; SYSTEM;
D O I
10.1007/s13361-013-0679-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel ion trap array (ITA) mass analyzer with six ion trapping and analyzing channels was investigated. It is capable of analyzing multiple samples simultaneously. The ITA was built with several planar electrodes made of stainless steel and 12 identical parallel zirconia ceramic substrates plated with conductive metal layers. Each two of the opposing ceramic electrode plates formed a boundary of an ion trap channel and six identical ion trapping and analyzing channels were placed in parallel without physical electrode between any two adjacent channels. The electric field distribution inside each channel was studied with simulation. The new design took the advantage of high precision machining attributable to the rigidity of ceramic, and the convenience of surface patterning technique. The ITA system was tested by using a two-channel electrospray ionization source, a multichannel simultaneous quadruple ion guide, and two detectors. The simultaneous analysis of two different samples with two adjacent ITA channels was achieved and independent mass spectra were obtained. For each channel, the mass resolution was tested. Additional ion trap functions such as mass-selected ion isolation and collision-induced dissociation (CID) were also tested. The results show that one ITA is well suited for multiple simultaneous mass analyses.
引用
收藏
页码:1420 / 1427
页数:8
相关论文
共 50 条
  • [1] Ion trap array mass analyzer: Structure and performance
    Laser Chemistry Institute, Department of Chemistry, Fudan University, Shanghai 200433, China
    不详
    不详
    Anal. Chem., 1600, 12 (4840-4846):
  • [2] Ion Trap Array Mass Analyzer: Structure and Performance
    Li, Xiaoxu
    Jiang, Gongyu
    Luo, Chan
    Xu, Fuxing
    Wang, Yuanyuan
    Ding, Li
    Ding, Chuan-Fan
    ANALYTICAL CHEMISTRY, 2009, 81 (12) : 4840 - 4846
  • [3] Optimization of Ion Trap Array Mass Analyzer by Computer Simulation
    Luo Chan
    Ding Chuan-Fan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2012, 40 (07) : 989 - 995
  • [4] A simplified toroidal ion trap mass analyzer
    Taylor, Nicholas
    Austin, Daniel E.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2012, 321 : 25 - 32
  • [5] Biomolecular dual-ion-trap mass analyzer
    Hsu, Yun-Fei
    Lin, Jung-Lee
    Chu, Ming-Lee
    Chen, Chung-Hsuan
    ANALYST, 2013, 138 (17) : 4823 - 4829
  • [6] Miniaturized Linear Wire Ion Trap Mass Analyzer
    Wu, Qinghao
    Li, Ailin
    Tian, Yuan
    Zare, Richard N.
    Austin, Daniel E.
    ANALYTICAL CHEMISTRY, 2016, 88 (15) : 7800 - 7806
  • [7] Design and performance of the halo ion trap mass analyzer
    Wang, Miao
    Austin, Daniel E.
    Tolley, Samuel E.
    Hansen, Brett J.
    Hawkins, Aaron R.
    Lee, Edgar D.
    Lee, Milton L.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2009, 237 : 275 - 275
  • [8] Analytical Performance of Printed Circuit Board Ion Trap Array Mass Analyzer with Electrospray Ionization
    Chu Yan-Qiu
    Xiao Yu
    Ling Xin
    Ding Chuan-Fan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2013, 41 (01) : 152 - 158
  • [9] Analysis of volatile organic compounds in air with a micro ion trap mass analyzer
    Moxom, J
    Reilly, PTA
    Whitten, WB
    Ramsey, M
    ANALYTICAL CHEMISTRY, 2003, 75 (15) : 3739 - 3743
  • [10] Ion-trap multiple mass spectrometry in pesticide analysis
    Larsen, BR
    ANALUSIS, 2000, 28 (10) : 941 - 946