Four-dimensional dose reconstruction through in vivo phase matching of cine images of electronic portal imaging device

被引:4
|
作者
Yoon, Jihyung [1 ]
Jung, Jae Won [1 ]
Kim, Jong Oh [2 ]
Yi, Byong Yong [3 ]
Yeo, Inhwan [4 ]
机构
[1] EAST Carolina Univ, Dept Phys, Greenville, NC 27858 USA
[2] Univ Pittsburgh, Inst Canc, Dept Radiat Oncol, Pittsburgh, PA 15232 USA
[3] Univ Maryland, Sch Med, Dept Radiat Oncol, Baltimore, MD 21201 USA
[4] Loma Linda Univ, Med Ctr, Dept Radiat Med, Loma Linda, CA 92354 USA
关键词
4D; inverse dose reconstruction; phase matching; EPID; Monte Carlo; VMAT QA; DOSIMETRY; MOTION; IMRT;
D O I
10.1118/1.4954317
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A method is proposed to reconstruct a four-dimensional (4D) dose distribution using phase matching of measured cine images to precalculated images of electronic portal imaging device (EPID). Methods: (1) A phantom, designed to simulate a tumor in lung (a polystyrene block with a 3 cm diameter embedded in cork), was placed on a sinusoidally moving platform with an amplitude of 1 cm and a period of 4 s. Ten-phase 4D computed tomography (CT) images of the phantom were acquired. A planning target volume (PTV) was created by adding a margin of 1 cm around the internal target volume of the tumor. (2) Three beams were designed, which included a static beam, a theoretical dynamic beam, and a planning-optimized dynamic beam (PODB). While the theoretical beam was made by manually programming a simplistic sliding leaf motion, the planning-optimized beam was obtained from treatment planning. From the three beams, three-dimensional (3D) doses on the phantom were calculated; 4D dose was calculated by means of the ten phase images (integrated over phases afterward); serving as "reference" images, phase-specific EPID dose images under the lung phantom were also calculated for each of the ten phases. (3) Cine EPID images were acquired while the beams were irradiated to the moving phantom. (4) Each cine image was phase-matched to a phase-specific CT image at which common irradiation occurred by intercomparing the cine image with the reference images. (5) Each cine image was used to reconstruct dose in the phase-matched CT image, and the reconstructed doses were summed over all phases. (6) The summation was compared with forwardly calculated 4D and 3D dose distributions. Accounting for realistic situations, intratreatment breathing irregularity was simulated by assuming an amplitude of 0.5 cm for the phantom during a portion of breathing trace in which the phase matching could not be performed. Intertreatment breathing irregularity between the time of treatment and the time of planning CT was considered by utilizing the same reduced amplitude when the phantom was irradiated. To examine the phase matching in a humanoid environment, the matching was also performed in a digital phantom (4D XCAT phantom). Results: For the static, the theoretical, and the planning-optimized dynamic beams, the 4D reconstructed doses showed agreement with the forwardly calculated 4D doses within the gamma pass rates of 92.7%, 100%, and 98.1%, respectively, at the isocenter plane given by 3%/3 mm criteria. Excellent agreement in dose volume histogram of PTV and lung-PTV was also found between the two 4D doses, while substantial differences were found between the 3D and the 4D doses. The significant breathing irregularities modeled in this study were found not to be noticeably affecting the reconstructed dose. The phase matching was performed equally well in a digital phantom. Conclusions: The method of retrospective phase determination of a moving object under irradiation provided successful 4D dose reconstruction. This method will provide accurate quality assurance and facilitate adaptive therapy when distinguishable objects such as well-defined tumors, diaphragm, and organs with markers (pancreas and liver) are covered by treatment beam apertures. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:4420 / 4430
页数:11
相关论文
共 25 条
  • [1] Four-Dimensional Dose Reconstruction Through Retrospective Phase Determination Using Cine Images of Electronic Portal Imaging Device
    Yoon, J.
    Jung, J.
    Yi, B.
    Kim, J.
    Yeo, I.
    MEDICAL PHYSICS, 2015, 42 (06) : 3667 - 3667
  • [2] Three-dimensional dose reconstruction using an electronic portal imaging device
    Hagness, C
    Herman, M
    MEDICAL PHYSICS, 2003, 30 (06) : 1345 - 1345
  • [3] Retrospective reconstruction of four-dimensional magnetic resonance from interleaved cine imaging - A comparative study with four-dimensional computed tomography in the lung
    Peteani, Giulia
    Paganelli, Chiara
    Giovannelli, Anna Chiara
    Bachtiary, Barbara
    Safai, Sairos
    Rogers, Susanne
    Pusterla, Orso
    Riesterer, Oliver
    Weber, Damien Charles
    Lomax, Antony John
    Baroni, Guido
    Fattori, Giovanni
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 29
  • [4] Reconstruction of four-dimensional computed tomography images during treatment time using electronic portal imaging device images based on a dynamic 2D/3D registration
    Nakamoto, T.
    Arimura, H.
    Hirose, T. A.
    Ohga, S.
    Umezu, Y.
    Nakamura, Y.
    Honda, H.
    Sasaki, T.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [5] Markerless Tumor Motion Tracking in Cine Images from Megavoltage Electronic Portal Imaging Device
    Kwang-Ho Cheong
    Jai-Woong Yoon
    Soah Park
    Sei-Kwon Kang
    Journal of the Korean Physical Society, 2019, 74 : 822 - 826
  • [6] Markerless Tumor Motion Tracking in Cine Images from Megavoltage Electronic Portal Imaging Device
    Cheong, Kwang-Ho
    Yoon, Jai-Woong
    Park, Soah
    Kang, Sei-Kwon
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2019, 74 (08) : 822 - 826
  • [7] Computer-Assisted Verification of Accumulated Dose Distribution During the Treatment Time Based On Estimation of Four-Dimensional Dose Distribution Using An Electronic Portal Imaging Device
    Arimura, H.
    Mizoguchi, A.
    Shioyama, Y.
    Nakamura, K.
    Yoshidome, S.
    Hirose, T.
    Anai, S.
    Umezu, Y.
    Honda, H.
    Ohki, M.
    Toyofuku, F.
    Hirata, H.
    MEDICAL PHYSICS, 2012, 39 (06) : 3679 - 3679
  • [8] Modeling Electronic Portal Imaging Device for Dose Reconstruction using Monte Carlo Method
    Jung, J. W.
    Kim, J. O.
    Yeo, I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S724 - S724
  • [9] A Real-Time Four-Dimensional Reconstruction Algorithm of Cine-Magnetic Resonance Imaging (Cine-MRI) Using Deep Learning
    Tamura, Yuto
    Demachi, Kazuyuki
    Igaki, Hiroshi
    Okamoto, Hiroyuki
    Nakano, Masahiro
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (03)
  • [10] Development of Artificial Intelligence as a Conversion Tool for Cine Electronic Portal Imaging Device Images to Radiotherapy Dosimetry: Preliminary Study
    Ramadhan M.M.
    Wibowo W.E.
    Pawiro S.A.
    Iranian Journal of Medical Physics, 2022, 19 (05): : 296 - 304