Quantitative assay of metronidazole by capillary zone electrophoresis with amperometric detection at a gold microelectrode

被引:2
|
作者
Jin, WR [1 ]
Li, W
Xu, Q
Dong, Q
机构
[1] Shandong Univ, Sch Chem, Analyt Sci Lab, Jinan 250100, Peoples R China
[2] Hlth & Dis Prevent Ctr Shandong Prov, Jinan, Peoples R China
[3] Anim Husb Bur Shandong Prov, Jinan, Peoples R China
关键词
capillary electrophoresis; electrochemical detection; metronidazole;
D O I
10.1002/(SICI)1522-2683(20000401)21:7<1409::AID-ELPS1409>3.0.CO;2-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Capillary zone electrophoresis was employed for the determination of metronidazole using end-column amperometric detection with a gold microelectrode at a constant potential of -0.52V vs, saturated calomel electrode. To overcome interference of oxygen in the solution, a deaeration injector and a deaeration protector at the detection cell were used. The optimum conditions of separation and detection are 1.0 x 10(-3) mol/L potassium dihydrogen citrate (KH2C6H5O7) for the buffer solution, 20 kV for the separation voltage, and 5 kV and 10 S for injection voltage and injection time, respectively. The limit of detection is 6.0 x 10(7) mol/L or 0.78 fmole (S/N = 3). The relative standard deviation is 3.9% for the electrophoretic peak current. The method was applied to the determination of metronidazole in human urine.
引用
收藏
页码:1409 / 1414