Determination of protein with TPPS by total internal reflection synchronous fluorescence spectroscopy

被引:0
|
作者
Chen, Y [1 ]
Yao, MN [1 ]
Tang, YJ [1 ]
Li, YQ [1 ]
机构
[1] Xiamen Univ, Dept Chem, MOE Key Lab Analyt Sci, Xiamen 361005, Peoples R China
关键词
total internal reflection; synchronous fluorescence; TPPS; protein;
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A new method of quantitative determination for serum albumin in aqueous solution has been developed by measuring total internal reflection synchronous fluorescence at the solid/liquid interface. The combination of bovine serum album in (BSA) and mesotetrakis (4-sulphonatophenyl) porphyrin (TPPS) adsorbed onto the glass surface produced a synchronous fluorescence signal at 421 nm. At pH 4.25, the signal intensity of BSA adsorbed on the interface was proportional to the BSA concentration in bulk solution. The linear range of 1.0-8.0 mu g center dot mL(-1) and the detection limit of 0.94 mu g center dot mL(-1) were obtained. The human serum samples were determined with satisfactory results.
引用
收藏
页码:2048 / 2051
页数:4
相关论文
共 8 条