Tandem mass spectrometry in quadrupole ion trap and ion cyclotron resonance mass spectrometers

被引:23
|
作者
Payne, AH [1 ]
Glish, GL [1 ]
机构
[1] Univ N Carolina, Dept Chem, Chapel Hill, NC 27599 USA
来源
关键词
D O I
10.1016/S0076-6879(05)02004-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Instruments that trap ions in a magnetic and/or electric field play a very important role in the analysis of biomolecules. The two predominant instruments in the category of trapping instrument are the quadrupole ion trap mass spectrometer (QIT-MS) and the ion cyclotron resonance (ICR) MS. The latter is also commonly called Fourier transform MS (FF-MS). The QIT is an inexpensive, simple, and rugged MS used for various routine applications. The ICR-MS is an expensive, high-performance instrument with figures of merit for resolution and mass accuracy surpassing all other mass spectrometers. This chapter covers the basic principles of operation of these instruments, including the trapping/manipulation/detection of ions and various approaches used to activate ions to perform tandem mass spectrometry experiments.