Strength retention of self-reinforced poly-L-lactide screws. A comparison of compression moulded and machine cut screws

被引:43
|
作者
Pohjonen, T
Helevirta, P
Tormala, P
Koskikare, K
Patiala, H
Rokkanen, P
机构
[1] TAMPERE UNIV TECHNOL,INST PLAST TECHNOL,BIOMAT LAB,FIN-33101 TAMPERE,FINLAND
[2] UNIV HELSINKI,CENT HOSP,DEPT ORTHOPAED & TRAUMATOL,FIN-00260 HELSINKI,FINLAND
基金
芬兰科学院;
关键词
D O I
10.1023/A:1018516529591
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effect of the manufacturing method on the strength retention of self-reinforced poly-L-lactide (SR-PLLA) screws was studied in vitro and in vivo from 3 up to at least 15 weeks. SR-PLLA screws were manufactured from axially oriented SR-PLLA billets by the conventional compression moulding process and an in-house developed machine cutting technique. New machined SR-PLLA screws (thread diameters 4.5 mm and 3.5 mm) were significantly stronger than older compression moulded SR-PLLA screws (4.5 mm and 3.5 mm) in bending and torque strength tests but significantly weaker in shear strength tests. In pull out tests there were not significant strength differences between the screws. Mechanical analysis and molecular weight measurements confirmed earlier observations that SR-PLLA degrades faster in vivo than in vitro. These results suggest that the new screws could be suitable for clinical use.
引用
收藏
页码:311 / 320
页数:10
相关论文
共 50 条
  • [21] Degradation of and intraosseous reactions to biodegradable poly-l-lactide screws: a study in minipigs
    M. Fuchs
    G. Köster
    T. Krause
    H.-A. Merten
    A. Schmid
    Archives of Orthopaedic and Trauma Surgery, 1998, 118 : 140 - 144
  • [22] Degradation of and intraosseous reactions to biodegradable poly-L-lactide screws:: a study in minipigs
    Fuchs, M
    Köster, G
    Krause, T
    Merten, HA
    Schmid, A
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1998, 118 (03) : 140 - 144
  • [23] Lumbar intervertebral disc replacement using bioabsorbable self-reinforced poly-L-lactide full-threaded screws, or cylindrical implants of polylactide polymers, bioactive glass and polyactive™
    Palmgren, T
    Ylinen, P
    Tulamo, RM
    Kellomäki, M
    Törmälä, P
    Rokkanen, P
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2003, 16 (03) : 138 - 144
  • [24] Tibial condylar fractures fixed with totally absorbable, self-reinforced polyglycolide screws. A preliminary report
    Kankare J.
    Archives of Orthopaedic and Trauma Surgery, 1997, 116 (3) : 133 - 136
  • [25] Comparison of poly-L-lactide and polylactide carbonate interference screws in an ovine anterior cruciate ligament reconstruction model
    Walsh, William R.
    Cotton, Nick J.
    Stephens, Paul
    Brunelle, John E.
    Langdown, Andy
    Auld, Jason
    Vizesi, Frank
    Bruce, Warwick
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2007, 23 (07): : 757 - 765
  • [26] Bioabsorbable self-reinforced poly-L-lactide, metallic, and silicone stents in the management of experimental tracheal stenosis
    Korpela, A
    Aarnio, P
    Sariola, H
    Törmälä, P
    Harjula, A
    CHEST, 1999, 115 (02) : 490 - 495
  • [27] Hydroxyapatite/poly-L-lactide acid screws have better biocompatibility and femoral burr hole closure than does poly-L-lactide acid alone
    Akagi, Hiroyuki
    Iwata, Munetaka
    Ichinohe, Tomu
    Amimoto, Hirokazu
    Hayashi, Yuuta
    Kannno, Nobuo
    Ochi, Hiroki
    Fujita, Yukihiro
    Harada, Yasuji
    Tagawa, Masahiro
    Hara, Yasushi
    JOURNAL OF BIOMATERIALS APPLICATIONS, 2014, 28 (06) : 954 - 962
  • [28] Do intramedullary rods of self-reinforced poly-L-lactide or poly-DL/L-lactide cause lactic acid acidosis in rabbits?
    Vasenius, J.
    Majola, A.
    Miettinen, E.L.
    Tormala, P.
    Rokkanen, P.
    Clinical Materials, 1992, 10 (04): : 213 - 218
  • [29] Comparison of intra- and extraosseally placed self-reinforced poly-L-lactide plates in the fixation of distal femoral osteotomies in rabbits
    Koskikare, K
    Pihlajamaki, H
    Patiala, H
    Rokkanen, P
    ANNALES CHIRURGIAE ET GYNAECOLOGIAE, 1997, 86 (03) : 261 - 268
  • [30] Fixation of osteotomies of the distal femur with absorbable, self-reinforced, poly-L-lactide plates - An experimental study on rabbits
    Koskikare, K
    Hirvensalo, E
    Patiala, H
    Rokkanen, P
    Pohjonen, T
    Tormala, P
    Lob, G
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1997, 116 (6-7) : 352 - 356