Patient body motion correction for dynamic cardiac PET-CT by attenuation-emission alignment according to projection consistency conditions

被引:7
|
作者
Hunter, Chad R. R. N. [1 ,2 ]
Klein, Ran [1 ,3 ]
Alessio, Adam M. [4 ]
deKemp, Robert A. [1 ,2 ]
机构
[1] Carleton Univ, 1125 Colonel By Dr, Ottawa, ON K1S 5B6, Canada
[2] Univ Ottawa Heart Inst, 40 Ruskin St, Ottawa, ON K1Y 4W7, Canada
[3] Ottawa Hosp, 1053 Carling Ave, Ottawa, ON K1Y 4E9, Canada
[4] Michigan State Univ, 775 Woodlot Dr, E Lansing, MI 48824 USA
基金
加拿大自然科学与工程研究理事会;
关键词
dynamic positron emission tomography; motion correction; patient body motion; MYOCARDIAL BLOOD-FLOW; RESPIRATORY MOTION; RB-82; QUANTIFICATION; STRESS; IMAGE; REST;
D O I
10.1002/mp.13419
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
IntroductionPatient body motion is known to cause large deviations in the determination of myocardial blood flow (MBF) with errors exceeding 300%. Accurate correction for patient whole-body motion is still a largely unsolved problem in cardiac positron emission tomography (PET) imaging. ObjectiveThis study evaluated the efficacy of using Natterer's formulation of the Helgason-Ludwig consistency conditions on the two-dimensional Radon transform to align computed tomography to PET projection data in multiple time frames of a dynamic sequence for the purpose of frame-by-frame correction of rigid whole-body motion. MethodsThe correction algorithm was evaluated with digital NCAT phantoms using realistic noise added by the analytical simulator. Count rates used in the simulation were derived from clinical patient data. In addition, a proof of concept test using measured data with a cardiac torso phantom was conducted. ResultsMotion correction resulted in significant improvement in the accuracy of MBF estimates, especially for high count-rate acquisitions. Maximum errors for 2cm of motion dropped from 325% to 25% and from 250% to 25% using global and regional partial-volume correction, respectively. Median MBF errors dropped from 33% to 4.5% and 27% to 3.8%, respectively. Importantly, the correction algorithm performed equally well to compensate for body motion in both early and late time frames. ConclusionCardiac PET-CT data used for attenuation correction (CTAC) alignment using projection consistency conditions was effective for reducing errors in MBF measurements due to simulated patient motion, and can be integrated into the image reconstruction workflow.
引用
收藏
页码:1697 / 1706
页数:10
相关论文
共 45 条
  • [1] Attenuation-Emission Alignment in Cardiac PET/CT with Consistency Conditions
    Alessio, Adam
    Caldwell, James
    Chen, Grace
    Branch, Kelley
    Kinahan, Paul
    [J]. 2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3288 - 3291
  • [2] Attenuation-emission alignment in cardiac PET/CT based on consistency conditions
    Alessio, Adam M.
    Kinahan, Paul E.
    Champley, Kyle M.
    Caldwell, James H.
    [J]. MEDICAL PHYSICS, 2010, 37 (03) : 1191 - 1200
  • [3] Automatic Attenuation and Emission Alignment in Cardiac PET/CT Based on Consistency Conditions
    Alessio, A. M.
    Caldwell, J. H.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (04) : 665 - 665
  • [4] Projection based patient body motion correction in Rb-82 PET-CT using consistency condition
    Hunter, Chad
    Alessio, Adam
    deKemp, Robert
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [5] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Osman, MM
    Cohade, C
    Nakamoto, Y
    Wahl, RL
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (04) : 603 - 606
  • [6] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Medhat M. Osman
    Christian Cohade
    Yuji Nakamoto
    Richard L. Wahl
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 603 - 606
  • [7] Alignment of dynamic cardiac PET images for correction of motion
    Turkington, TG
    DeGrado, TR
    Hanson, MW
    Coleman, RE
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (02) : 235 - 242
  • [8] A novel method for incorporating respiratory-matched attenuation correction in the motion correction of cardiac PET-CT studies
    McQuaid, Sarah J.
    Lambrou, Tryphon
    Hutton, Brian F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (10): : 2903 - 2915
  • [9] Effect of Non-Alignment/Alignment of Attenuation Map Without/With Emission Motion Correction in Cardiac SPECT/CT
    Dey, Joyoni
    Segars, W. Paul
    Pretorius, P. Hendrik
    King, Michael A.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (04) : 1813 - 1824
  • [10] Correction for patient body motion in Rb-82 PET-CT: An NCAT simulation study
    Hunter, Chad
    Klein, Ran
    deKemp, Robert
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55