Ceramic Debris in Hip Prosthesis: Correlation Between Synovial Fluid and Joint Capsule

被引:4
|
作者
De Pasquale, Dalila [1 ]
Stea, Susanna [1 ]
Beraudi, Alina [1 ]
Montesi, Monica [1 ]
Squarzoni, Stefano [2 ]
Toni, Aldo [1 ,3 ]
机构
[1] Ist Ortoped Rizzoli, Med Technol Lab, I-40136 Bologna, Italy
[2] CNR Natl Res Council Italy, Inst Mol Genet, Bologna, Ior, Italy
[3] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
来源
JOURNAL OF ARTHROPLASTY | 2013年 / 28卷 / 05期
关键词
synovial fluid; joint capsule; ceramic particles; hip prosthesis; WEAR DEBRIS; MOLECULAR-WEIGHT; POLYETHYLENE; PARTICLES; FAILURE; TISSUES; METAL;
D O I
10.1016/j.arth.2013.01.019
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Detection of ceramic particles in synovial fluids allows early diagnosis of ceramic damage, but there is no evidence of a relationship between ceramic debris in the articular space and in the joint capsule. The aim of the present study is to verify if the particles isolated in the synovial fluid are comparable with those stored in the capsular tissue. Twenty-one patients were enrolled. Both synovial fluid and capsular samples were collected during revision surgery and ceramic particles were isolated and analyzed by scanning electron microscopy and energy-dispersive X-ray microanalysis. It resulted a significant correlation between the samples couples (18 out of 21). This study confirms that the synovial fluid analysis can give a clear definition of the presence of particles in the joint capsule. (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:838 / 841
页数:4
相关论文
共 50 条
  • [1] Ceramic wear debris in hip prosthesis perimplant environment. Correlation between synovial fluid and neocapsule
    De Pasquale, D.
    Beraudi, A.
    Montesi, M.
    Squarzoni, S.
    Traina, F.
    Stea, S.
    Toni, A.
    [J]. BONE, 2012, 50 : S188 - S188
  • [2] Ceramic wear debris in synovial fluid allow early diagnosis of hip prosthesis failure
    Stea, S.
    Traina, F.
    Beraudi, A.
    Montesi, M.
    Squarzoni, S.
    Sudanese, A.
    Toni, A.
    [J]. BONE, 2009, 44 (02) : S289 - S289
  • [3] Synovial fluid microanalysis allows early diagnosis of ceramic hip prosthesis damage
    Stea, Susanna
    Traina, Francesco
    Beraudi, Alina
    Montesi, Monica
    Bordini, Barbara
    Squarzoni, Stefano
    Sudanese, Alessandra
    Toni, Aldo
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2012, 30 (08) : 1312 - 1320
  • [4] Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream
    De Pasquale, Dalila
    Stea, Susanna
    Squarzoni, Stefano
    Bordini, Barbara
    Amabile, Marilina
    Catalani, Simona
    Apostoli, Pietro
    Toni, Aldo
    [J]. INTERNATIONAL ORTHOPAEDICS, 2014, 38 (03) : 469 - 475
  • [5] Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream
    Dalila De Pasquale
    Susanna Stea
    Stefano Squarzoni
    Barbara Bordini
    Marilina Amabile
    Simona Catalani
    Pietro Apostoli
    Aldo Toni
    [J]. International Orthopaedics, 2014, 38 : 469 - 475
  • [6] Fraunhofer Develops a Ceramic Resurfacing Prosthesis for the Hip Joint
    不详
    [J]. PRAKTISCHE METALLOGRAPHIE-PRACTICAL METALLOGRAPHY, 2021, 58 (01): : 63 - 64
  • [7] SYNOVIAL-FLUID CIRCULATION IN THE HIP-JOINT
    AFOKE, A
    HUTTON, WC
    BYERS, PD
    [J]. MEDICAL HYPOTHESES, 1984, 15 (01) : 81 - 86
  • [8] Synovial joint fluid cytokine levels in hip disease
    Abe, Hirohito
    Sakai, Takashi
    Ando, Wataru
    Takao, Masaki
    Nishii, Takashi
    Nakamura, Nobuo
    Hamasaki, Toshimitsu
    Yoshikawa, Hideki
    Sugano, Nobuhiko
    [J]. RHEUMATOLOGY, 2014, 53 (01) : 165 - 172
  • [9] Comparison between synovial fluid cytology and joint capsule histopathology in horses with chronic osteochondritis dissecans
    Nocera, Irene
    Vitale, Valentina
    Citi, Simonetta
    Marianelli, Giulia
    Poli, Alessandro
    Bonelli, Francesca
    Melanie, Pierre
    Sgorbini, Micaela
    [J]. LARGE ANIMAL REVIEW, 2019, 25 (06): : 231 - 234
  • [10] Detection of cobalt in synovial fluid from metal-on-metal hip prosthesis: correlation with the ion haematic level
    Beraudi, Alina
    Catalani, Simona
    Montesi, Monica
    Stea, Susanna
    Sudanese, Alessandra
    Apostoli, Pietro
    Toni, Aldo
    [J]. BIOMARKERS, 2013, 18 (08) : 699 - 705