Comparison of different optimized irregular sparse 2D ultrasound arrays

被引:0
|
作者
Diarra, Bakary [1 ]
Roux, Emmanuel [2 ,3 ,4 ,5 ,6 ,7 ]
Liebgott, Herve [2 ,3 ,4 ,5 ,6 ]
Ravi, Samikannu [1 ]
Robini, Marc [2 ,3 ,4 ,5 ,6 ]
Tortoli, Piero [7 ]
Cachard, Christian [2 ,3 ,4 ,5 ,6 ]
机构
[1] Botswana Int Univ Sci & Technol, Elect Elect Comp & Telecommun Engn Dept, Palapye, Botswana
[2] Univ Lyon, CREATIS, Villeurbanne, France
[3] CNRS, UMR 5220, Villeurbanne, France
[4] INSERM, U1044, Villeurbanne, France
[5] Univ Lyon 1, Villeurbanne, France
[6] INSA Lyon, Villeurbanne, France
[7] Univ Florence, Dept Informat Engn, Florence, Italy
关键词
2D array; irregular sparse array; 3D ultrasound imaging; simulated annealing;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The role of 2D matrix arrays in the realization of real time 3D ultrasound imaging is crucial as the latter permit to acquire a complete volume. The lack of control systems for these arrays containing thousands of elements and the dimension of connection cables for such a number of elements are the main obstacles to the use of these technologies. The sparse array techniques present a solid candidate to solve these technological limitations but give rise to beam pattern deteriorations in terms of energy loss and unwanted lobes apparition. The irregular positioning of the array elements permits to drastically decrease the unwanted lobes while the energy loss may be compensated by increasing the element size [1][2]. However, as elements get larger, the directivity decreases. To solve this directivity reduction, we propose to combine elements of different sizes in such a manner that a broad directivity is maintained using small elements whereas the energy loss may be compensated by wide ones. Additionally, an innovative simulated annealing based algorithm [3] is used to refine the array beam profile when the active element number is set to 256. The results compared to those obtained by the basic non-grid array show significant improvement in the beam pattern.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] 1D and 2D algorithmically optimized sparse arrays
    Austeng, A
    Holm, S
    Weber, PK
    Aakvaag, N
    Iranpour, K
    [J]. 1997 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 & 2, 1997, : 1683 - 1686
  • [2] Improved beamforming using curved sparse 2D arrays in ultrasound
    Kirkebo, Jan Egil
    Austeng, Andreas
    [J]. ULTRASONICS, 2007, 46 (02) : 119 - 128
  • [3] Optimization of free-moving elements in 2D ultrasound sparse arrays
    Diarra, Bakary
    Robini, Marc
    Roux, Emmanuel
    Liebgott, Herve
    Cachard, Christian
    Roux, Emmanuel
    Tortoli, Piero
    [J]. 2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 2189 - 2192
  • [4] 2D BEAMFORMING ON SPARSE ARRAYS WITH SPARSE BAYESIAN LEARNING
    Nannuru, Santosh
    Gerstoft, Peter
    [J]. 2019 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING (ICASSP), 2019, : 4355 - 4359
  • [5] Variable-size elements in 2D sparse arrays for 3D medical ultrasound
    Diarra, Bakary
    Robini, Marc
    Liebgott, Herve
    Cachard, Christian
    Diarra, Bakary
    Tortoli, Piero
    [J]. 2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 508 - 511
  • [6] Ultrasound creates 2D arrays of droplets
    不详
    [J]. PHYSICS WORLD, 2016, 29 (11) : 4 - 4
  • [7] Toward Vibrotactile Rendering for Irregular 2D Tactor Arrays
    Lipari, Nicholas G.
    Borst, Christoph W.
    [J]. 2016 IEEE SYMPOSIUM ON 3D USER INTERFACES (3DUI), 2016, : 257 - 258
  • [8] OPTIMIZED 2D ARRAY DESIGN FOR ULTRASOUND IMAGING
    Diarra, Bakary
    Liebgott, Herve
    Robini, Marc
    Tortoli, Piero
    Cachard, Christian
    [J]. 2012 PROCEEDINGS OF THE 20TH EUROPEAN SIGNAL PROCESSING CONFERENCE (EUSIPCO), 2012, : 2718 - 2722
  • [9] Optimized arrays for 2D resistivity surveys with combined surface and buried arrays
    Loke, M. H.
    Kiflu, H.
    Wilkinson, P. B.
    Harro, D.
    Kruse, S.
    [J]. NEAR SURFACE GEOPHYSICS, 2015, 13 (05) : 505 - 517
  • [10] Advanced MEMS ferroelectric ultrasound 2D arrays
    Bernstein, JJ
    Bottari, J
    Houston, K
    Kirkos, G
    Miller, R
    Xu, B
    Ye, Y
    Cross, LE
    [J]. 1999 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 1999, : 1145 - 1153