A comparison of instrumentation using Naviflex and Profile nickel-titanium engine-driven rotary instruments

被引:6
|
作者
Ottosen, SR [1 ]
Nicholls, JI [1 ]
Steiner, JC [1 ]
机构
[1] Univ Washington, Sch Dent, Dept Endodont, Seattle, WA 98195 USA
关键词
D O I
10.1016/S0099-2399(99)80279-9
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This study was designed to compare the changes in canal configuration resulting from instrumentation by either Profile or Naviflex instruments. Forty mesial canals in extracted human molar teeth were embedded and sectioned at two root levels. Reassembled teeth were instrumented with a modified crown-down technique as described in the Profile training video for Profile files and in a similar manner for Naviflex instruments. Superimposed pre-and postinstrumented cross-sectional root images were projected, traced, and scanned into a computer for analysis. Canal movement, in relation to the furca, and canal area change were recorded. The results showed no significant difference in canal center movement or canal area change between the Profile or Naviflex groups. The degree of canal curvature had no effect on canal center movement or canal area change.
引用
收藏
页码:457 / 460
页数:4
相关论文
共 50 条
  • [1] A review of instrumentation kinematics of engine-driven nickel-titanium instruments
    Capar, I. D.
    Arslan, H.
    INTERNATIONAL ENDODONTIC JOURNAL, 2016, 49 (02) : 119 - 135
  • [2] Evolution and development: engine-driven endodontic rotary nickel-titanium instruments
    Yuhong Liang
    Lin Yue
    International Journal of Oral Science, 2022, 14
  • [3] Evolution and development:engine-driven endodontic rotary nickel-titanium instruments
    Yuhong Liang
    Lin Yue
    International Journal of Oral Science, 2022, (01) : 12 - 19
  • [4] Evolution and development:engine-driven endodontic rotary nickel-titanium instruments
    Yuhong Liang
    Lin Yue
    International Journal of Oral Science, 2022, 14 (01) : 12 - 19
  • [5] Evolution and development: engine-driven endodontic rotary nickel-titanium instruments
    Liang, Yuhong
    Yue, Lin
    INTERNATIONAL JOURNAL OF ORAL SCIENCE, 2022, 14 (01)
  • [6] Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments
    de Melo, MCC
    Bahia, MGD
    Buono, VTL
    JOURNAL OF ENDODONTICS, 2002, 28 (11) : 765 - 769
  • [7] Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments
    Haïkel, Y
    Serfaty, R
    Bateman, G
    Senger, B
    Allemann, C
    JOURNAL OF ENDODONTICS, 1999, 25 (06) : 434 - 440
  • [8] Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods
    Gambarini, Gianluca
    Grande, Nicola Maria
    Plotino, Gianluca
    Somma, Francesco
    Garala, Manish
    De Luca, Massimo
    Testarelli, Luca
    JOURNAL OF ENDODONTICS, 2008, 34 (08) : 1003 - 1005
  • [9] Canal wall planning by engine-driven nickel-titanium instruments, compared with stainless-steel hand instrumentation
    Tucker, DM
    Wenckus, CS
    Bentkover, SK
    JOURNAL OF ENDODONTICS, 1997, 23 (03) : 170 - 173
  • [10] Evaluation of the corrosion in engine-driven rotatory nickel-titanium endodontic instruments by Scanning Electron Microscopy
    Martins, RC
    Buono, VTL
    Bahia, MGA
    JOURNAL OF DENTAL RESEARCH, 2001, 80 (04) : 1029 - 1029