Evaluation of target coverage and margins adequacy during CyberKnife Lung Optimized Treatment

被引:20
|
作者
Ricotti, Rosalinda [1 ,8 ]
Seregni, Matteo [2 ]
Ciardo, Delia [1 ]
Vigorito, Sabrina [3 ]
Rondi, Elena [3 ]
Piperno, Gaia [1 ]
Ferrari, Annamaria [1 ]
Zerella, Maria Alessia [1 ,4 ]
Arculeo, Simona [1 ,4 ]
Francia, Claudia Maria [1 ,4 ]
Sibio, Daniela [1 ,4 ]
Cattani, Federica [3 ]
De Marinis, Filippo [5 ]
Spaggiari, Lorenzo [4 ,5 ]
Orecchia, Roberto [6 ,7 ]
Riboldi, Marco [2 ,9 ]
Baroni, Guido [2 ,8 ]
Jereczek-Fossa, Barbara Alicja [1 ,4 ]
机构
[1] European Inst Oncol, Div Radiat Oncol, Milan, Italy
[2] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, Milan, Italy
[3] European Inst Oncol, Unit Med Phys, Milan, Italy
[4] Univ Milan, Dept Oncol & Hematooncol, Milan, Italy
[5] European Inst Oncol, Div Thorac Oncol, Milan, Italy
[6] European Inst Oncol, Sci Directorate, Milan, Italy
[7] European Inst Oncol, Dept Med Imaging & Radiat Sci, Milan, Italy
[8] CNAO, Bioengn Unit, Pavia, Italy
[9] Ludwig Maximilians Univ Munchen, Dept Med Phys, Munich, Germany
关键词
CyberKnife; LOT; lung optimized treatment; lung stereotactic radiotherapy; margins; target coverage; STEREOTACTIC ABLATIVE RADIOTHERAPY; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; TIME TUMOR-TRACKING; BODY RADIATION-THERAPY; MOTION; CANCER; ACCURACY; SYSTEM; 4DCT; PATTERNS;
D O I
10.1002/mp.12804
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Evaluation of target coverage and verification of safety margins, in motion management strategies implemented by Lung Optimized Treatment (LOT) module in CyberKnife system. MethodsThree fiducial-less motion management strategies provided by LOT can be selected according to tumor visibility in the X ray images acquired during treatment. In 2-view modality the tumor is visible in both X ray images and full motion tracking is performed. In 1-view modality the tumor is visible in a single X ray image, therefore, motion tracking is combined with an internal target volume (ITV)-based margin expansion. In 0-view modality the lesion is not visible, consequently the treatment relies entirely on an ITV-based approach. Data from 30 patients treated in 2-view modality were selected providing information on the three-dimensional tumor motion in correspondence to each X ray image. Treatments in 1-view and 0-view modalities were simulated by processing log files and planning volumes. Planning target volume (PTV) margins were defined according to the tracking modality: end-exhale clinical target volume (CTV)+3mm in 2-view and ITV+5mm in 0-view. In the 1-view scenario, the ITV encompasses only tumor motion along the non-visible direction. Then, non-uniform ITV to PTV margins were applied: 3mm and 5mm in the visible and non-visible direction, respectively. We defined the coverage of each voxel of the CTV as the percentage of X ray images where such voxel was included in the PTV. In 2-view modality coverage was calculated as the intersection between the CTV centred on the imaged target position and the PTV centred on the predicted target position, as recorded in log files. In 1-view modality, coverage was calculated as the intersection between the CTV centred on the imaged target position and the PTV centred on the projected predictor data. In 0-view modality coverage was calculated as the intersection between the CTV centred on the imaged target position and the non-moving PTV. Similar to dose-volume histogram, CTV coverage-volume histograms (defined as CVH) were derived for each patient and treatment modality. The geometric coverages of the 90% and 95% of CTV volume (C90, C95, respectively) were evaluated. Patient-specific optimal margins (ensuring C9595%) were computed retrospectively. ResultsThe medianinterquartile-rage of C90 and C95 for upper lobe lesions was 99.10.6% and 99.0 +/- 3.1%, whereas they were 98.9 +/- 4.2% and 97.8 +/- 7.5% for lower and middle lobe tumors. In 2-view, 1-view and 0-view modality, adopted margins ensured C9595% in 70%, 85% and 63% of cases and C9590% in 90%, 88% and 83% of cases, respectively. In 2-view, 1-view and 0-view a reduction in margins still ensured C9595% in 33%, 78% and 59% of cases, respectively. ConclusionsCTV coverage analysis provided an a-posteriori evaluation of the treatment geometric accuracy and allowed a quantitative verification of the adequacy of the PTV margins applied in CyberKnife LOT treatments offering guidance in the selection of CTV margins.
引用
收藏
页码:1360 / 1368
页数:9
相关论文
共 50 条
  • [1] Evaluation of target coverage in lung stereotactic radiotherapy with Cyberknife system
    Ricotti, R.
    Seregni, M.
    Ciardo, D.
    Vigorito, S.
    Rondi, E.
    Piperno, G.
    Zerella, M. A.
    Arculeo, S.
    Francia, C.
    Sibio, D.
    Cattani, F.
    Orecchia, R.
    Riboldi, M.
    Baroni, G.
    Jereczek-Fossa, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1107 - S1108
  • [2] Evaluation of PTV Dose Margins for Lung SBRT On Cyberknife Treatment
    Fan, J.
    Wang, L.
    Jin, L.
    Xu, Q.
    Chen, X.
    Li, J.
    Price, R.
    Ma, C.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Optimized CyberKnife Lung Treatment: Effect of Fractionated Tracking Volume Change on Tracking Results
    Li, Guo-quan
    Wang, Ye
    Qiu, Meng-jun
    Yang, Jing
    Peng, Zhen-jun
    Zhang, Sheng
    Fang, Xiefan
    Yang, Sheng-li
    [J]. DISEASE MARKERS, 2020, 2020
  • [4] Analysis of normal lung irradiation in radiosurgery treatments: a comparison of lung optimized treatment (LOT) on cyberknife, 4D target volume on helical tomotherapy, and DIBH on linear accelerator
    Raghavendra Holla
    D. Khanna
    V. K. Sathiya Narayanan
    Deb Narayan Dutta
    [J]. Physical and Engineering Sciences in Medicine, 2021, 44 : 1321 - 1329
  • [5] Analysis of normal lung irradiation in radiosurgery treatments: a comparison of lung optimized treatment (LOT) on cyberknife, 4D target volume on helical tomotherapy, and DIBH on linear accelerator
    Holla, Raghavendra
    Khanna, D.
    Narayanan, V. K. Sathiya
    Dutta, Deb Narayan
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2021, 44 (04) : 1321 - 1329
  • [6] Evaluation of Lung and Liver Tumor Dose Coverage Treated With the CyberKnife Synchrony System With Consideration of Measured Tracking Errors
    Akino, Yuichi
    Shiomi, Hiroya
    Higashinaka, Naoichi
    Kouno, Tomohiro
    Mabuchi, Nobuhisa
    Isohashi, Fumiaki
    Seo, Yuji
    Fujiwara, Kei
    Tamenaga, Setsuo
    Ogawa, Kazuhiko
    [J]. ANTICANCER RESEARCH, 2023, 43 (01) : 231 - 238
  • [7] Lung optimized treatment with CyberKnife® in inoperable lung cancer patients: feasibility analysis of a mono-institutional 115 patient series
    Francia, C. M.
    Marvaso, G.
    Piperno, G.
    Gandini, S.
    Ferrari, A.
    Zerella, M. A.
    Arculeo, S.
    Sibio, D.
    Fodor, C.
    Pepa, M.
    Trivellato, S.
    Rondi, E.
    Vigorito, S.
    Cattani, F.
    Spaggiari, L.
    de Marinis, F.
    Orecchia, R.
    Ciardo, D.
    Jereczek-Fossa, B. A.
    [J]. NEOPLASMA, 2020, 67 (03) : 684 - 691
  • [8] The impacts of mid-treatment CBCT-guided patient repositioning on target coverage during lung VMAT
    Mathieu, Dominique
    Campeau, Marie-Pierre
    Bedwani, Stephane
    Roberge, David
    Doucet, Robert
    Zerouali, Karim
    Bahig, Houda
    Vu, Toni
    Lambert, Louise
    Masucci, Laura
    Filion, Edith
    [J]. JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2017, 61 (04) : 543 - 549
  • [9] Evaluation of a New Method for CyberKnife Treatment for Central Lung and Mediastinal Tumors by Tracheobronchial Tracking
    Suzuki, Toshihiro
    Saito, Masahide
    Takahashi, Hiroshi
    Suzuki, Hidekazu
    Makino, Koji
    Ueda, Koji
    Mochizuki, Koji
    Mochizuki, Zennosuke
    Nemoto, Hikaru
    Sano, Naoki
    Onishi, Hiroshi
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2024, 23
  • [10] Target coverage: VMAT vs 3D in the treatment of lung cancer
    Rincon Perez, M.
    Luna Tirado, J.
    Gomez-Tejedor Alonso, S.
    Garcia Castejon, M. A.
    Penedo Cobos, J. M.
    Marin Arango, J. P.
    Perez Casas, A. M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S770 - S771