A study of matching fluid loss in a biomedical microwave tomography system

被引:23
|
作者
Gilmore, Colin [1 ]
Zakaria, Amer [2 ]
LoVetri, Joe [2 ]
Pistorius, Stephen [3 ,4 ]
机构
[1] TRLabs, Winnipeg, MB R3T 6A8, Canada
[2] Univ Manitoba, Dept Elect & Comp Engn, Winnipeg, MB R3T 5V6, Canada
[3] Univ Manitoba, CancerCare Manitoba, Winnipeg, MB R3T 2N2, Canada
[4] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3T 2N2, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
microwave imaging; electromagnetic tomography; BREAST-CANCER DETECTION; IMAGE-RECONSTRUCTION; INVERSION; RADAR; WOMEN;
D O I
10.1118/1.4788640
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Effective imaging of human tissue with microwave tomography systems requires a matching fluid to reduce the wave reflections at the tissue boundary. Further, in order to match the idealized mathematical model used for imaging with the complicated physical measurement environment, loss must be added to the matching fluid. Both too little and too much loss result in low-quality images, but due to the nonlinear nature of the imaging problem, the exact nature of loss-to-image quality cannot be predicted a priori. Possible optimal loss levels include a single, highly sensitive value, or a broad range of acceptable losses. Herein, the authors outline a process of determining an appropriate level of loss inside the matching fluid and attempt to determine the bounds for which the images are the highest quality. Methods: Our biomedical microwave tomography system is designed for 2D limb imaging, operating from 0.8 to 1.2 GHz. Our matching fluid consists of deionized water with various levels of loss introduced by the addition of table salt. Using two homogeneous tissue-mimicking phantoms, and eight different matching fluids of varying salt concentrations, the authors introduce quantitative image quality metrics based on L-norms, mean values, and standard deviations to test the tomography system and assess image quality. Images are generated with a balanced multiplicative regularized contrast source inversion algorithm. The authors further generate images of a human forearm which may be analyzed qualitatively. Results: The image metrics for the phantoms support the claim that the worst images occur at the extremes of high and low salt concentrations. Importantly, the image metrics show that there exists a broad range of salt concentrations that result in high-quality images, not a single optimal value. In particular, 2.5-4.5 g of table salt per liter of deionized water provide the best reconstruction quality for simple phantoms. The authors argue that qualitatively, the human forearm data provide the best images at approximately the same salt concentrations. Conclusions: There exists a relatively large-range of matching fluid losses (i.e., salt concentrations) that provide similar image quality. In particular, it is not necessary to spend time highly optimizing the level of loss in the matching fluid. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4788640]
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Matching Medium for Biomedical Microwave Imaging
    Lui, Hoi-Shun
    Fhager, Andreas
    Persson, Mikael
    [J]. 2015 INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION (ISAP), 2015,
  • [2] TEMPERATURE INFLUENCE OF MATCHING LIQUID IN A MICROWAVE TOMOGRAPHY PLATFORM SYSTEM
    Kim, J. Y.
    Lee, K. J.
    Son, S. H.
    Jeon, S. I.
    Kim, N.
    [J]. MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2014, 56 (12) : 2978 - 2981
  • [3] MICROWAVE DIFFRACTION TOMOGRAPHY FOR BIOMEDICAL APPLICATIONS
    PERONNET, G
    GUERQUINKERN, JL
    GAUTHERIE, M
    SZELES, C
    MICHEL, Y
    [J]. MICROWAVES & RF, 1983, 22 (08) : 146 - 146
  • [4] MICROWAVE DIFFRACTION TOMOGRAPHY FOR BIOMEDICAL APPLICATIONS
    BOLOMEY, JC
    IZADNEGAHDAR, A
    JOFRE, L
    PICHOT, C
    PERONNET, G
    SOLAIMANI, M
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1982, 30 (11) : 1998 - 2000
  • [5] MICROWAVE DIFFRACTION TOMOGRAPHY FOR BIOMEDICAL APPLICATIONS - COMMENT
    GUO, TC
    GUO, WW
    LARSEN, LE
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1984, 32 (04) : 473 - 474
  • [6] An Electromagnetic Metasurface for Impedance Matching in Microwave Biomedical Applications
    Dellabate, Alessandro
    Brizi, Danilo
    Shaw, Tarakeswar
    Perez, Mauricio
    Augustine, Robin
    Monorchio, Agostino
    [J]. 2024 18TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION, EUCAP, 2024,
  • [7] Microwave Tomography as a New Modality for Biomedical and Biopharmaceutical Imaging
    Wu, Zhi-Peng
    [J]. PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON MEDICINE AND BIOPHARMACEUTICALS, 2016, : 687 - 692
  • [8] An attempt of time domain microwave computed tomography for biomedical use
    Miyakawa, M
    Eiyama, M
    Ishii, N
    [J]. PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 3300 - 3303
  • [9] System and Formulation Options for Biomedical Microwave Imaging
    LoVetri, Joe
    Mojabi, Puyan
    Zakaria, Amer
    Ostadrahimi, Majid
    Jeffrey, Ian
    [J]. 2014 XXXITH URSI GENERAL ASSEMBLY AND SCIENTIFIC SYMPOSIUM (URSI GASS), 2014,
  • [10] A Tomographic Multistatic System for Biomedical Microwave Sensing
    Bisio, Igor
    Estatico, Claudio
    Fedeli, Alessandro
    Lavagetto, Fabio
    Pasiorino, Matteo
    Randazzo, Andrea
    Sciarrone, Andrea
    [J]. 2020 14TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP 2020), 2020,