Characterizations of BC501A and BC537 liquid scintillator detectors

被引:16
|
作者
Qin, Jianguo [1 ,2 ,3 ]
Lai, Caifeng [2 ]
Ye, Bangjiao [1 ]
Liu, Rong [2 ]
Zhang, Xinwei [4 ]
Jiang, Li [2 ]
机构
[1] Univ Sci & Technol China, Dept Modern Phys, Hefei 230026, Peoples R China
[2] China Acad Engn Phys, Inst Nucl Phys & Chem, Mianyang 621900, Peoples R China
[3] China Acad Engn Phys, Grad Sch, Mianyang 621900, Peoples R China
[4] Inst Appl Phys & Computat Math, Beijing 100088, Peoples R China
基金
中国国家自然科学基金;
关键词
Liquid scintillator detector; Light output function; Energy resolution function; gamma-Ray response matrix; Detection efficiency; NEUTRON; DISCRIMINATION; CALIBRATION; SPECTRA;
D O I
10.1016/j.apradiso.2015.06.008
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Two 2 '' x 2 '' liquid scintillator detectors BC537 and BC501A have been characterized for their responses and efficiencies to gamma-ray detection. Light output resolution and response functions were derived by least-squares minimization of a simulated response function, fitted to experimental data. The gamma-ray response matrix and detection efficiency were simulated with Monte Carlo (MC) methods and validated. For photon energies below 2.4 MeVee, the resolution, as well as the efficiency, of BC501A is better than BC537 scintillator. The situation is reversed when the energy is higher than 2.4 MeVee. BC537 has higher gamma-ray detection efficiency than BC501A if the impinging photon energy is more than 2 MeV due to different ratios of C to H/D atoms. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:15 / 24
页数:10
相关论文
共 48 条
  • [1] Calibration of BC501A liquid scintillator cells with monochromatic neutron beams
    Arneodo, F
    Benetti, P
    Bettini, A
    di Tigliole, AB
    Calligarich, E
    Carpanese, C
    Casagrande, F
    Cavalli, D
    Cavanna, F
    Cennini, P
    Centro, S
    Cesana, A
    Chen, C
    Chen, YB
    Cline, D
    Consorte, O
    De Mitri, I
    Dolfini, R
    Ferrari, A
    Berzolari, AG
    He, KL
    Huang, XP
    Li, ZH
    Lu, F
    Ma, JM
    Mannocchi, S
    Matthey, C
    Mauri, F
    Mazzone, L
    Montanari, C
    Nardo, R
    Otwinowski, S
    Parlati, S
    Pascoli, D
    Pepato, A
    Periale, L
    Piano Mortari, G
    Piazzoli, A
    Picchi, P
    Pietropaolo, F
    Rappoldi, A
    Raselli, GL
    Resconi, S
    Revol, JP
    Rossella, M
    Rossi, C
    Rubbia, C
    Sala, P
    Scannicchio, D
    Sergiampietri, F
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 418 (2-3): : 285 - 299
  • [2] Calibration of BC501A liquid scintillator cells with monochromatic neutron beams
    Arneodo, F.
    Benetti, P.
    Bettini, A.
    di Tigliole, A.Borio
    Calligarich, E.
    Carpanese, C.
    Casagrande, F.
    Cavalli, D.
    Cavanna, F.
    Cennini, P.
    Centro, S.
    Cesana, A.
    Chen, C.
    Chen, Y.B.
    Cline, D.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1998, 418 (2-3): : 285 - 299
  • [3] Energy calibration of a BC501A liquid scintillator using a γ-γ coincidence technique
    Yan Jie
    Liu Rong
    Li Cheng
    Jiang Li
    Lu Xin-Xin
    Zhu Tong-Hua
    CHINESE PHYSICS C, 2010, 34 (07) : 993 - 997
  • [4] Energy calibration of a BC501A liquid scintillator using a γ-γ coincidence technique
    言杰
    刘荣
    李澄
    蒋励
    鹿心鑫
    朱通华
    Chinese Physics C, 2010, 34 (07) : 993 - 997
  • [5] γ-ray response function measurements and Monte-Carlo simulation of BC537 liquid scintillator
    Feng, Song
    Chen, Xiaohui
    Liu, Rong
    Yang, Yiwei
    Lu, Xinxin
    Zhu, Tonghua
    Chen, X. (chenxh1988@126.com), 1600, Editorial Office of High Power Laser and Particle Beams (26):
  • [6] Application research on neutron-gamma discrimination based on BC501A liquid scintillator
    Luo, J.
    Hou, S.
    KERNTECHNIK, 2021, 86 (06) : 437 - 444
  • [7] Neutron/gamma discrimination employing the power spectrum analysis of the signal from the liquid scintillator BC501A
    Luo, X. L.
    Wang, Y. K.
    Yang, J.
    Liu, G.
    Lin, C. B.
    Hu, Q. Q.
    Peng, J. X.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 717 : 44 - 50
  • [8] Application of a BC501A Liquid Scintillation Detector with a Gain Stabilization System on the EAST Tokamak
    彭星宇
    陈忠靖
    杜腾飞
    胡志猛
    葛理健
    陈金象
    李湘庆
    樊铁栓
    Plasma Science and Technology, 2016, 18 (01) : 23 - 29
  • [9] Application of a BC501A Liquid Scintillation Detector with a Gain Stabilization System on the EAST Tokamak
    彭星宇
    陈忠靖
    杜腾飞
    胡志猛
    葛理健
    陈金象
    李湘庆
    樊铁栓
    Plasma Science and Technology, 2016, (01) : 23 - 29
  • [10] Application of a BC501A Liquid Scintillation Detector with a Gain Stabilization System on the EAST Tokamak
    Peng Xingyu
    Chen Zhongjing
    Du Tengfei
    Hu Zhimeng
    Ge Lijian
    Chen Jinxiang
    Li Xiangqing
    Fan Tieshuan
    PLASMA SCIENCE & TECHNOLOGY, 2016, 18 (01) : 23 - 29