Development of an automated turbidimetric immunoassay for quantification of bovine serum immunoglobulin G

被引:0
|
作者
Etzel, LR [1 ]
Strohbehn, RE [1 ]
McVicker, JK [1 ]
机构
[1] MIDLAND BIOPROD,BOONE,IA 50036
关键词
D O I
暂无
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective-To develop an automated turbidimetric immunoassay (TIA) for measurement of bovine IgG. Sample Population-24 bovine serum samples. Procedure-IgG concentration was determined by use of the TIA, and results were compared with those of the radial immunodiffusion (RID) method. Variables were determined, using commercially available reagents and a clinical biochemical analyzer. For the TIA, polyclonal goat anti-bovine IgG (Fc specific) serum, bovine IgG calibrator serum, and polyethylene glycol reaction buffer were used. Sample concentrations were determined by the instrument, using the linear regression method of least squares. The accuracy of this assay was validated by referencing to a purified bovine IgG standard and by recovery of control standards. Parallelism was documented by assay linearity and serial sample dilution linearity. Interference resulting from hemolyzed samples was examined. Results-The TIA method correlated positively (r = 0.9957) and significantly (P < 0.05) with the RID method, yielding a regression equation with slope of 0.78708 and y-intercept of 1.02102. Bias attributable to hemolysis was not observed. Conclusions-The TIA method is automated, accurate, and precise for bovine serum IgG quantification. Clinical Relevance-This assay provides sample results in approximately 10 minutes and may be used as an alternative to the manual RID method.
引用
收藏
页码:1201 / 1205
页数:5
相关论文
共 50 条
  • [1] Short communication: Evaluation of serum immunoglobulin G concentrations using an automated turbidimetric immunoassay in dairy calves
    Alley, M. L.
    Haines, D. M.
    Smith, G. W.
    [J]. JOURNAL OF DAIRY SCIENCE, 2012, 95 (08) : 4596 - 4599
  • [2] Quantitative determination of human serum gelsolin:: Development and validation of an automated turbidimetric immunoassay
    Christensen, PH
    Gronkjær, K
    [J]. CLINICA CHIMICA ACTA, 2005, 355 : S221 - S221
  • [3] Comparison of enzyme-linked immunosorbent assay and Fassisi® bovine immunoglobulin G (IgG) immunoassay for quantification of bovine IgG in neonatal calf serum
    Hampe, Marian
    Soellner-Donat, Stefanie
    Failing, Klaus
    Wehrend, Axel
    [J]. VETERINARY WORLD, 2021, 14 (12) : 3211 - 3215
  • [4] Measurement of serum IgG in foals by radial immunodiffusion and automated turbidimetric immunoassay
    Davis, DG
    Schaefer, DMW
    Hinchcliff, KW
    Wellman, ML
    Willet, VE
    Fletcher, JM
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 2005, 19 (01): : 93 - 96
  • [5] Quantification of human complement C2 protein using an automated turbidimetric immunoassay
    Tange, Clare Elizabeth
    Johnson-Brett, Bridget
    Cook, Alex
    Stordeur, Patrick
    Brohet, Fabian
    Jolles, Stephen
    Steven, Rachel
    Ponsford, Mark
    Roberts, Andrew
    El-Shanawany, Tariq
    Harding, Stephen
    Wallis, Gregg
    Parker, Antony Richard
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (09) : 1498 - 1506
  • [6] A NEW AUTOMATED TURBIDIMETRIC IMMUNOASSAY FOR QUANTIFYING ALPHA-1-ANTITRYPSIN IN SERUM
    VIEDMA, JA
    DELAIGLESIA, A
    PARERA, M
    LOPEZ, MT
    [J]. CLINICAL CHEMISTRY, 1986, 32 (06) : 1020 - 1022
  • [7] DEVELOPMENT OF AN AUTOMATED FLOW-INJECTION CHEMILUMINESCENCE IMMUNOASSAY FOR HUMAN-IMMUNOGLOBULIN-G
    HACKER, A
    HINTERLEITNER, M
    SHELLUM, C
    GUBITZ, G
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 352 (7-8): : 793 - 796
  • [8] TURBIDIMETRIC LATEX IMMUNOASSAY FOR SERUM FERRITIN
    BERNARD, A
    LAUWERYS, R
    [J]. JOURNAL OF IMMUNOLOGICAL METHODS, 1984, 71 (02) : 141 - 147
  • [9] Quantification of human C1 esterase inhibitor protein using an automated turbidimetric immunoassay
    Tange, Clare E.
    Kaur, Amrit
    Verma, Nisha
    Hickey, Alaco
    Grigoriadou, Sofia
    Scott, Chris
    Kiani, Sorena
    Steven, Rachael
    Ponsford, Mark
    El-Shanawany, Tariq
    Jolles, Stephen
    Harding, Stephen
    Parker, Antony R.
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2019, 33 (01)
  • [10] TURBIDIMETRIC ASSAY FOR QUANTIFICATION OF IMMUNOGLOBULIN-A IN SALIVA
    KAGE, A
    HAMPEL, P
    WEITZEL, D
    KOTTGEN, E
    [J]. JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1988, 26 (11): : 752 - 753